Stray radiation dose reduction using a backscatter shield and a rectangular collimator for a handheld intraoral X-ray unit

被引:0
|
作者
Otaka, Yusei [1 ,2 ]
Harata, Yasuo [1 ]
Izawa, Maki [1 ]
Shiba, Noriyoshi [1 ]
Iwawaki, Atsushi [3 ]
Ishii, Takeru [3 ]
Ochiai, Akihiro [3 ]
Takahashi, Nobutoshi [1 ]
Saka, Hideki [2 ,3 ]
Kito, Shinji [1 ]
机构
[1] Meikai Univ, Dept Diagnost & Therapeut Sci, Div Dent Radiol, Sch Dent, 1-1 Keyakidai, Sakado, Saitama 3500283, Japan
[2] Meikai Univ, Forens Odontol Ctr, Sch Dent, 1-1 Keyakidai, Sakado, Saitama 3500283, Japan
[3] Meikai Univ, Dept Diagnost & Therapeut Sci, Div Forens Odontol, Sch Dent, 1-1 Keyakidai, Sakado, Saitama 3500283, Japan
基金
日本学术振兴会;
关键词
D O I
10.1093/rpd/ncad239
中图分类号
X [环境科学、安全科学];
学科分类号
08 ; 0830 ;
摘要
The objective of this study was to study the dose reduction effect on occupational exposure of operators and public exposure other than operators when a backscatter shield and a rectangular collimator are used in conjunction with a handheld intraoral X-ray unit. The occupational exposure was reduced to 40% when the backscatter shield was attached to the cone-tip, to 13% when the rectangular collimator was attached and to 7.7% when the backscatter shield and rectangular collimator were used together. On the other hand, the public exposure was reduced to 20% when the rectangular collimator was attached, but the backscatter shield was not effective in reducing the public exposure. Attaching a backscatter shield is effective in reducing the occupational exposure, and a rectangular aperture is effective in reducing the occupational exposure, as well as the public exposure.
引用
收藏
页码:2349 / 2355
页数:7
相关论文
共 50 条
  • [1] ASSESSMENT OF OCCUPATIONAL RADIATION DOSE FROM CAMERA MODEL INTRAORAL HANDHELD X-RAY DEVICE WITHOUT STRAY RADIATION PROTECTION SHIELD
    Moshfeghi, Mahkameh
    Safi, Yaser
    Afzalan, Alireza
    Ahsaie, Mitra Ghazizadeh
    RADIATION PROTECTION DOSIMETRY, 2022, 198 (1-2) : 1 - 7
  • [2] Experimental Study of the Reduction of the Dose to the Radiographic Operator Using an Additional Rectangular Collimator and Detector Holder During Handheld Exposure with a Portable Intraoral X-Ray Device
    Otaka, Yusei
    Harata, Yasuo
    Izawa, Maki
    Shiba, Noriyoshi
    Takahashi, Nobutoshi
    Kito, Shinji
    JOURNAL OF INDIAN ACADEMY OF ORAL MEDICINE AND RADIOLOGY, 2024, 36 (04) : 381 - 386
  • [3] Development of an open project rectangular collimator for use with intraoral dental X-ray unit
    Poletto, Marina C.
    Thomazi, Eduardo
    Zorzi, Janete E.
    Gamba, Thiago O.
    Perottoni, Claudio A.
    RADIOLOGICAL PHYSICS AND TECHNOLOGY, 2024, 17 (01) : 315 - 321
  • [4] Development of an open project rectangular collimator for use with intraoral dental X-ray unit
    Marina C. Poletto
    Eduardo Thomazi
    Janete E. Zorzi
    Thiago O. Gamba
    Cláudio A. Perottoni
    Radiological Physics and Technology, 2024, 17 : 315 - 321
  • [5] Dose reduction from loss of backscatter in superficial x-ray radiation therapy with the Pantak SXT 150 unit
    B. J. Healy
    S. Sylvander
    K. N. Nitschke
    Australasian Physics & Engineering Sciences in Medicine, 2008, 31 : 49 - 55
  • [6] Dose reduction from loss of backscatter in superficial x-ray radiation therapy with the Pantak SXT 150 unit
    Healy, B. J.
    Sylvander, S.
    Nitschke, K. N.
    AUSTRALASIAN PHYSICAL & ENGINEERING SCIENCES IN MEDICINE, 2008, 31 (01) : 49 - 55
  • [7] SPECTROMETER SHIELD FOR STRAY IRON X-RAY RADIATION IN SCANNING ELECTRON MICROSCOPE
    BALLARD, DB
    YAKOWITZ, H
    REVIEW OF SCIENTIFIC INSTRUMENTS, 1971, 42 (01): : 14 - &
  • [8] Significant Reduction in X-ray Dose with a Simple Tube Housing Shield
    Steege, P. C.
    Gainor, J. P.
    Wilson, R. F.
    HEALTH PHYSICS, 2023, 125 (01): : 36 - 36
  • [9] Detection of explosive devices using X-ray backscatter radiation
    Faust, AA
    PENETRATING RADIATION SYSTEMS AND APPLICATIONS IV, 2002, 4786 : 17 - 28
  • [10] Assessment of the Occupational Radiation Dose from a Handheld Portable X-ray Unit During Full-mouth Intraoral Dental Radiographs in the Dog and the Cat - A Pilot Study
    Villamizar-Martinez, Lenin A. A.
    Losey, Jeannie