Early clinical results of St. Jude Medical Symmetry aortic connector

被引:3
|
作者
Toyama M. [1 ]
Usui A. [1 ]
Abe T. [1 ]
Yoshikawa M. [1 ]
Akita T. [1 ]
Ueda Y. [1 ]
机构
[1] Department of Cardiothoracic Surgery, Nagoya University, Graduate School of Medicine, Showa-ku, Nagoya 466-8550
关键词
Aortic connector; Graft flow; Hyperlipidemia; Patency; Saphenous vein graft;
D O I
10.1007/s10047-005-0293-0
中图分类号
学科分类号
摘要
An automated anastomosis device named St. Jude Medical symmetry aortic connector has been used worldwide for off-pump coronary artery bypass grafting. However, early graft obstruction was recently reported, and its predictors should be clarified. From April 2002 to March 2004, 38 patients in our institution underwent off-pump coronary artery bypass grafting using the St. Jude Medical Symmetry aortic connector for saphenous vein graft (SVG) procedures; measurement of intraoperative graft flow and postoperative coronary angiography was performed. Early SVG events occurred in 9 (24%) patients: 8 occlusion cases and 1 case of stenosis. Predictors of early SVG events were assessed from a comparative study. Univariate logistic regression identified hyperlipidemia as the only significant predictor of early SVG events (P = 0.02, odds ratio 7.78). Lower SVG flow rate and poor ejection fraction did not show statistical significance (P = 0.09, odds ratio 1.09 and P = 0.09, odds ratio 0.96). The SVG event rate was much higher for the left circumflex branch compared with other locations (31% vs. 9%, P = 0.03) and decreased with increasing aortic connector size (small, 32%; median, 14%; large, 0%). Multivariate analysis did not identify a predictor of SVG events. The aortic connector is associated with a high incidence of early SVG events. Patients should be cautiously selected and the procedure should not be used for left anterior descending coronary artery or culprit lesions. © The Japanese Society for Artificial Organs 2005.
引用
收藏
页码:95 / 99
页数:4
相关论文
共 50 条
  • [31] St. Jude Medical Trifecta™ aortic valve perioperative performance in 200 patients
    Permanyer, Eduard
    Estigarribia, Arnaldo-Javier
    Ysasi, Alejandro
    Herrero, Enrique
    Semper, Omar
    Llorens, Rafael
    INTERACTIVE CARDIOVASCULAR AND THORACIC SURGERY, 2013, 17 (04) : 669 - 672
  • [32] Aortic valve replacement with the St. Jude Medical prosthesis and fixed dose anticoagulation
    Katircioglu, SF
    Yamak, B
    Ulus, AT
    Iscan, HZ
    Mavitas, B
    Tasdemir, O
    JOURNAL OF CARDIAC SURGERY, 1997, 12 (06) : 363 - 370
  • [33] Hemodynamic Performance of the St. Jude Medical Epic™ Supra Aortic Stented Valve
    Maitland, Andrew
    Hirsch, Gregory M.
    Pascoe, Edward A.
    JOURNAL OF HEART VALVE DISEASE, 2011, 20 (03): : 327 - 331
  • [34] THE ST-JUDE MEDICAL VALVE - EARLY CLINICAL-RESULTS IN 253 PATIENTS
    DUNCAN, JM
    COOLEY, DA
    LIVESAY, JJ
    OTT, DA
    REUL, GJ
    WALKER, WE
    FRAZIER, OH
    TEXAS HEART INSTITUTE JOURNAL, 1983, 10 (01) : 11 - 16
  • [35] THE SMALL AORTIC ROOT - EARLY CLINICAL-EVALUATION OF THE ST-JUDE MEDICAL VALVE
    GILL, CC
    KING, H
    LYTLE, BW
    COSGROVE, DM
    GOLDING, LAR
    LOOP, FD
    CIRCULATION, 1981, 64 (04) : 257 - 257
  • [36] Long-term results of aortic valve replacement with a small St. Jude Medical Valve in Japanese patients
    Yoshikawa, Kazuhiro
    Fukunaga, Shuji
    Arinaga, Koichi
    Hori, Hidetsugu
    Nakamura, Eiji
    Ueda, Tomohiro
    Tayama, Eiki
    Aoyagi, Shigeaki
    ANNALS OF THORACIC SURGERY, 2008, 85 (04): : 1303 - 1309
  • [37] St. jude medical and carboMedics mechanical heart valves in the aortic position comparison of long-term results
    Kandemir, Ozer
    Tokmakoglu, Hilmi
    Yildiz, Ulku
    Tezcaner, Tevfik
    Yorgancioglu, A. Cem
    Gunay, Iihan
    Suzer, Kaya
    Zorlutuna, Yaman
    TEXAS HEART INSTITUTE JOURNAL, 2006, 33 (02) : 154 - 159
  • [38] Early results with the Carbo-seal composite valve conduit for aortic root replacement: Comparison with the St. Jude Medical/Hemashield composite graft
    Tayama E.
    Akashi H.
    Ishihara K.
    Hayashida N.
    Akasu K.
    Kosuga T.
    Fukunaga S.
    Kawara T.
    Aoyagi S.
    Journal of Artificial Organs, 2000, 3 (2) : 120 - 125
  • [39] Early and mid-term haemodynamic performance and clinical outcomes of St. Jude Medical Trifecta™ valve
    Raimundo, Renata
    Moreira, Soraia
    Saraiva, Francisca
    Cerqueira, Rui J.
    Teixeira, Pedro
    Salgueiro, Elson
    Lourenco, Andre
    Amorim, Mario J.
    Almeida, Jorge
    Pinho, Paulo
    Leite-Moreira, Adelino F.
    JOURNAL OF THORACIC DISEASE, 2018, 10 (02) : 889 - 898
  • [40] St. Jude Toronto biologic aortic root prosthesis: Early FDA phase II IDE study results
    Bavaria, J
    Gleason, TG
    David, T
    Cosselli, J
    CIRCULATION, 2003, 108 (17) : 658 - 658