Effect of surface coating on the screw loosening of dental abutment screws

被引:0
|
作者
Chan-Ik Park
Han-Cheol Choe
Chae-Heon Chung
机构
[1] Chosun University,Department of Prosthodontics, College of Dentistry
[2] Chosun University,Department of Dental Materials & Research Center of Nano
来源
关键词
screw loosening; screw surface treatment; coated screw; scratch test;
D O I
暂无
中图分类号
学科分类号
摘要
Regardless of the type of performed restoration, in most cases, a screw connection is employed between the abutment and implant. For this reason, implant screw loosening has remained a problem in restorative practices. The purpose of this study was to compare the surface of coated/plated screws with titanium and gold alloy screws and to evaluate the physical properties of coated/plated material after scratch tests via FE-SEM (field emission scanning electron microscopy) investigation. GoldTite, titanium screws provided by 3i (Implant Innovation, USA) and TorqTite, titanium screws by Steri-Oss (Nobel Biocare, USA) and gold screws and titanium screws by AVANA (Osstem Implant, Korea) were selected for this study. The surface, crest, and root of the abutment screws were observed by FE-SEM. A micro-diamond needle was also prepared for the scratch test. Each abutment screw was fixed, and a scratch on the surface of the head region was made at constant load and thereafter the fine trace was observed with FE-SEM. The surface of GoldTite was smoother than that of other screws and it also had abundant ductility and malleability compared with titanium and gold screws. The scratch tests also revealed that teflon particles were exfoliated easily in the screw coated with teflon. The titanium screw had rough surface and low ductility. The clinical use of gold-plated screws is recommended as a means of preventing screw loosening.
引用
收藏
页码:549 / 553
页数:4
相关论文
共 50 条
  • [41] Effect of surface modification on the long-term stability of dental implant abutment screws by plasma nitriding treatment
    Sun, Fei
    Wang, Lei
    Li, Xin-Chang
    Cheng, Wei
    Lin, Zeng
    Ba, De-Chun
    Song, Gui-Qiu
    Sun, Chuan-Sheng
    SURFACE & COATINGS TECHNOLOGY, 2020, 399
  • [42] Coating Dental Implant Abutment Screws with Diamondlike Carbon Doped with Diamond Nanoparticles: The Effect on Maintaining Torque After Mechanical Cycling
    Souto Lepesqueur, Laura Soares
    Goncalves de Figueiredo, Viviane Maria
    Ferreira, Leandro Lameirao
    da Silva Sobrinho, Argemiro Soares
    Massi, Marcos
    Bottino, Marco Antonio
    Nogueira Junior, Lafayette
    INTERNATIONAL JOURNAL OF ORAL & MAXILLOFACIAL IMPLANTS, 2015, 30 (06) : 1310 - 1316
  • [43] Can gold coating on abutment screws increase preload?
    Liu, YH
    Lin, LD
    Wang, JS
    JOURNAL OF DENTAL RESEARCH, 2000, 79 (05) : 1329 - 1329
  • [44] The dental abutment screw and new methods to remove fractured abutment screw from jawbone: A review
    Shedmake, Swapnil
    Awate, Nilesh
    Walke, Pramod
    MATERIALS TODAY-PROCEEDINGS, 2022, 60 : 2254 - 2256
  • [45] Effects of Abutment Screw Coating on Implant Preload
    Park, Jae-Kyoung
    Choi, Jin-Uk
    Jeon, Young-Chan
    Choi, Kyung-Soo
    Jeong, Chang-Mo
    JOURNAL OF PROSTHODONTICS-IMPLANT ESTHETIC AND RECONSTRUCTIVE DENTISTRY, 2010, 19 (06): : 458 - 464
  • [46] Abutment screw loosening in angulation-correcting implants: An in vitro study
    Hotinski, Elena
    Dudley, James
    JOURNAL OF PROSTHETIC DENTISTRY, 2019, 121 (01): : 151 - 155
  • [47] Correlation between microleakage and screw loosening at implant-abutment connection
    Sahin, Cem
    Ayyildiz, Simel
    JOURNAL OF ADVANCED PROSTHODONTICS, 2014, 6 (01): : 35 - 38
  • [48] Influence of the implant abutment types and the dynamic loading on initial screw loosening
    Kim, Eun-Sook
    Shin, Soo-Yeon
    JOURNAL OF ADVANCED PROSTHODONTICS, 2013, 5 (01): : 21 - 28
  • [49] Abutment screw loosening in single-implant restorations: A systematic review
    Theoharidou, Anna
    Petridis, Haralampos P.
    Tzannas, Konstantinos
    Garefis, Pavlos
    INTERNATIONAL JOURNAL OF ORAL & MAXILLOFACIAL IMPLANTS, 2008, 23 (04) : 681 - 690
  • [50] Analysis of a device for assessment of abutment and prosthesis screw loosening in oral implants
    Hanses, G
    Smedberg, FI
    Nilner, K
    CLINICAL ORAL IMPLANTS RESEARCH, 2002, 13 (06) : 666 - 670