Reducing absorbed dose to eye lenses in head CT examinations: the effect of bismuth shielding

被引:0
|
作者
Alberto Ciarmatori
L. Nocetti
G. Mistretta
G. Zambelli
T. Costi
机构
[1] Az. Ospedaliero – Universitaria di Modena- Policlinico,Medical Physics Department
[2] Alma Mater Studiorum University,Post Graduate School in Medical Physics
[3] Lavoro e Ambiente Srl,undefined
关键词
Computed tomography; Eye lens; Bismuth shielding; Dose reduction; Image quality;
D O I
暂无
中图分类号
学科分类号
摘要
The eye lens is considered to be among the most radiosensitive human tissues. Brain CT scans may unnecessarily expose it to radiation even if the area of clinical interest is far from the eyes. The aim of this study is to implement a bismuth eye lens shielding system for Head-CT acquisitions in these cases. The study is focused on the assessment of the dosimetric characteristics of the shielding system as well as on its effect on image quality. The shielding system was tested in two set-ups which differ for distance (“contact” and “4 cm” Set up respectively). Scans were performed on a CTDI phantom and an anthropomorphic phantom. A reference set up without shielding system was acquired to establish a baseline. Image quality was assessed by signal (not HU converted), noise and contrast-to-noise ratio (CNR) evaluation. The overall dose reduction was evaluated by measuring the CTDIvol while the eye lens dose reduction was assessed by placing thermoluminescent dosimeters (TLDs) on an anthropomorphic phantom. The image quality analysis exhibits the presence of an artefact that mildly increases the CT number up to 3 cm below the shielding system. Below the artefact, the difference of the Signal and the CNR are negligible between the three different set-ups. Regarding the CTDI, the analysis demonstrates a decrease by almost 12 % (in the “contact” set-up) and 9 % (in the “4 cm” set-up). TLD measurements exhibit an eye lens dose reduction by 28.5 ± 5 and 21.1 ± 5 % respectively at the “contact” and the “4 cm” distance. No relevant artefact was found and image quality was not affected by the shielding system. Significant dose reductions were measured. These features make the shielding set-up useful for clinical implementation in both studied positions.
引用
收藏
页码:583 / 589
页数:6
相关论文
共 50 条
  • [1] Reducing absorbed dose to eye lenses in head CT examinations: the effect of bismuth shielding
    Ciarmatori, Alberto
    Nocetti, L.
    Mistretta, G.
    Zambelli, G.
    Costi, T.
    AUSTRALASIAN PHYSICAL & ENGINEERING SCIENCES IN MEDICINE, 2016, 39 (02) : 583 - 589
  • [2] REDUCING ABSORBED DOSE TO THYROID IN NECK CT EXAMINATIONS: THE EFFECTS OF SABA SHIELDING
    Saba, Valiallah
    Shuraki, Jalal Kargar
    Valizadeh, Abdollah
    Zahedinia, Mohsen
    Barkhordari, Maziar
    RADIATION PROTECTION DOSIMETRY, 2020, 191 (03) : 349 - 360
  • [3] Optimization of head CT protocol to reduce the absorbed dose in eye lenses and thyroid: A phantom study
    Kalhor P.
    Changizi V.
    Hosseini A.
    Jazayeri E.
    Iranian Journal of Medical Physics, 2019, 16 (01) : 64 - 74
  • [4] Eye radiation dose saving in head CT examinations using copper-bismuth radiation shield
    Keshtkar, Mohammad
    Blouri, Basir
    Mahmoudabadi, Alireza
    Alami, Ali
    RADIATION PROTECTION DOSIMETRY, 2023, 199 (02) : 146 - 151
  • [5] Would lead shielding the eye lenses during a head CT be beneficial?
    Alqahtani, F. F.
    Saeed, M. K.
    Abdallah, Y.
    Alshamrani, K.
    RADIOPROTECTION, 2023, 58 (03) : 221 - 228
  • [6] DOSE REDUCTION USING BISMUTH SHIELDING DURING PAEDIATRIC CT EXAMINATIONS IN SLOVAKIA
    Gbelcova, L'.
    Nikodemova, D.
    Horvathova, M.
    RADIATION PROTECTION DOSIMETRY, 2011, 147 (1-2) : 160 - 163
  • [7] Comparison of Bismuth Shielding and Globally Reducing Tube Current in Head and Thoracic CT
    Wang, J.
    Duan, X.
    Christner, J.
    Leng, S.
    Yu, L.
    McCollough, C.
    MEDICAL PHYSICS, 2011, 38 (06)
  • [8] Adaptive statistical iterative reconstruction and bismuth shielding for evaluation of dose reduction to the eye and image quality during head CT
    Kim, Myeong Seong
    Choi, Jiwon
    Kim, Sun Young
    Kweon, Dae Cheol
    JOURNAL OF THE KOREAN PHYSICAL SOCIETY, 2014, 64 (06) : 923 - 928
  • [9] Adaptive statistical iterative reconstruction and bismuth shielding for evaluation of dose reduction to the eye and image quality during head CT
    Myeong Seong Kim
    Jiwon Choi
    Sun Young Kim
    Dae Cheol Kweon
    Journal of the Korean Physical Society, 2014, 64 : 923 - 928
  • [10] Eye-lens bismuth shielding in paediatric head CT: artefact evaluation and reduction
    Raissaki, Maria
    Perisinakis, Kostas
    Damilakis, John
    Gourtsoyiannis, Nicholas
    PEDIATRIC RADIOLOGY, 2010, 40 (11) : 1748 - 1754