Would lead shielding the eye lenses during a head CT be beneficial?

被引:0
|
作者
Alqahtani, F. F. [1 ,2 ]
Saeed, M. K. [1 ]
Abdallah, Y. [3 ]
Alshamrani, K. [1 ]
机构
[1] Najran Univ, Appl Med Sci Coll, Dept Radiol Sci, Najran, Saudi Arabia
[2] Najran Univ, Hlth Res Ctr, Najran, Saudi Arabia
[3] Majmaah Univ, Coll Appl Med Sci, Dept Radiol Sci & Med Imaging, Al Majmaah 11952, Saudi Arabia
关键词
eye lenses; computed tomography; lead shielding; phantom; RADIATION-DOSE REDUCTION; SUPERFICIAL ORGANS; CURRENT MODULATION; IMAGE QUALITY; CATARACT; PROTECTION; RISK;
D O I
10.1051/radiopro/2023014
中图分类号
X [环境科学、安全科学];
学科分类号
08 ; 0830 ;
摘要
The lens of the eye is one of the radiosensitive organs. Patients receiving computed tomography (CT) scanning of the head, as well as patients undergoing repeated CT scanning of the head, are at risk for this consequence. The goal of this study was to see if a lead equivalent might reduce radiation reaching the eye's lens during regular cranial CT. The goal was accomplished by determining the ideal shielding thickness for patient applications by first evaluating the image quality using phantom measurements taken with and without lead shielding. LiF-thermoluminescent dosimeters were used to quantify the entrance surface doses (ESDs) to the eye lenses of the ATOM phantom without, with one, two, and three layers of lead shielding. The use of customized lead shields up to 0.1 mm thick might be used, according to ACR CT Phantom studies, without having a noticeable impact on the image quality for the central and posterior areas. In these experiments, 0.1 mm thick lead shielding reduces ESDs to the eye's lens by 31%. Eye shields significantly reduced image quality in the anterior region, but just slightly in the centre and posterior regions. Given the foregoing, using modified lead shields could drastically lower dose to the lens without severely affecting image quality.
引用
收藏
页码:221 / 228
页数:8
相关论文
共 50 条
  • [1] Would thyroid and breast shielding be beneficial in CT of the head?
    T. Beaconsfield
    R. Nicholson
    A. Thornton
    A. Al-Kutoubi
    European Radiology, 1998, 8 : 664 - 667
  • [2] Would thyroid and breast shielding be beneficial in CT of the head?
    Beaconsfield, T
    Nicholson, R
    Thornton, A
    Al-Kutoubi, A
    EUROPEAN RADIOLOGY, 1998, 8 (04) : 664 - 667
  • [3] Reducing absorbed dose to eye lenses in head CT examinations: the effect of bismuth shielding
    Ciarmatori, Alberto
    Nocetti, L.
    Mistretta, G.
    Zambelli, G.
    Costi, T.
    AUSTRALASIAN PHYSICAL & ENGINEERING SCIENCES IN MEDICINE, 2016, 39 (02) : 583 - 589
  • [4] Reducing absorbed dose to eye lenses in head CT examinations: the effect of bismuth shielding
    Alberto Ciarmatori
    L. Nocetti
    G. Mistretta
    G. Zambelli
    T. Costi
    Australasian Physical & Engineering Sciences in Medicine, 2016, 39 : 583 - 589
  • [5] Eye Shielding During Head CT Scans: Dose Reduction and Image Quality Evaluation
    Lee, Yuan-Hao
    Yang, Shih-Hung
    Lin, Yen-Kuang
    Glickman, Randolph D.
    Chen, Chia-Yuen
    Chan, Wing P.
    ACADEMIC RADIOLOGY, 2020, 27 (11) : 1523 - 1530
  • [6] Efficacy of breast shielding during CT of the head
    Z. Brnić
    B. Vekić
    A. Hebrang
    P. Anić
    European Radiology, 2003, 13 : 2436 - 2440
  • [7] Efficacy of breast shielding during CT of the head
    Brnic, Z
    Vekic, B
    Hebrang, A
    Anic, P
    EUROPEAN RADIOLOGY, 2003, 13 (11) : 2436 - 2440
  • [8] The efficiency of lead and non-lead shielding on breast dose in head CT
    Trozic, Sejla
    Mekis, Nejc
    Zalokar, Nika
    JOURNAL OF RADIOLOGICAL PROTECTION, 2020, 40 (03) : 816 - 826
  • [9] IRRADIATION DOSE TO THE LENS OF THE EYE DURING CT OF THE HEAD
    LUND, E
    HALABURT, H
    NEURORADIOLOGY, 1982, 22 (04) : 181 - 184
  • [10] Adaptive statistical iterative reconstruction and bismuth shielding for evaluation of dose reduction to the eye and image quality during head CT
    Kim, Myeong Seong
    Choi, Jiwon
    Kim, Sun Young
    Kweon, Dae Cheol
    JOURNAL OF THE KOREAN PHYSICAL SOCIETY, 2014, 64 (06) : 923 - 928