Eye Shielding During Head CT Scans: Dose Reduction and Image Quality Evaluation

被引:6
|
作者
Lee, Yuan-Hao [1 ]
Yang, Shih-Hung [1 ,2 ]
Lin, Yen-Kuang [3 ,4 ]
Glickman, Randolph D. [5 ]
Chen, Chia-Yuen [1 ,2 ]
Chan, Wing P. [1 ,2 ]
机构
[1] Taipei Med Univ, Wan Fang Hosp, Dept Radiol, 111 Hsing Long Rd,Sect 3, Taipei, Taiwan
[2] Taipei Med Univ, Coll Med, Sch Med, Dept Radiol, Taipei, Taiwan
[3] Taipei Med Univ, Biostat Ctr, Taipei, Taiwan
[4] Taipei Med Univ, Grad Inst Nursing, Taipei, Taiwan
[5] Univ Texas Hlth Sci Ctr San Antonio, Dept Ophthalmol, San Antonio, TX 78229 USA
关键词
Eye shield; Head computed tomography scan; Image quality; Radiation dose; Safety; TUBE CURRENT MODULATION;
D O I
10.1016/j.acra.2019.12.011
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Rationale and Objectives: In this study, we assessed the radiation dose to the lens and the impacts of various eye shields using either a fixed or modulated tube current. Materials and Methods: Patients undergoing head computed tomography (CT) examinations were recruited, and each was randomly assigned to one of five imaging groups, either without a CT eye shield or with one of two types of shielding and topogram-based tube current modulation (TCM). The radiation dose at the eye lens was estimated using Gafchromic films. All CT images were analyzed for quality in the orbit and brain areas. Two radiologists also qualitatively assessed image artifacts and their impacts on image quality using three-point Likert scales. Results: Both barium sulfate and bismuth-antimony shields significantly reduced radiation dose to the lens (by 28.60% -31.92% and 43.87% -47.00%, respectively) while significantly inducing image artifacts. The image quality of the intraocular structure, but not the intracranial structure, was significantly degraded by shielding. In addition, discriminating the periocular tissues was improved using a bismuthantimony shield and topogram-based TCM. Compared to fixed tube current, topogram-based TCM provided better signal-to-noise and contrast-to-noise ratios in the intracranial structures when the bismuth-antimony and barium sulfate shields were applied, respectively. Conclusion: Artifacts resulting from the application of eye shields during head CT examinations can be reduced by using topogram-based TCM instead of a fixed tube current. This could be an alternative approach for maintaining image quality in CT scans that do not encompass organ-based TCM.
引用
收藏
页码:1523 / 1530
页数:8
相关论文
共 50 条
  • [1] Adaptive statistical iterative reconstruction and bismuth shielding for evaluation of dose reduction to the eye and image quality during head CT
    Myeong Seong Kim
    Jiwon Choi
    Sun Young Kim
    Dae Cheol Kweon
    [J]. Journal of the Korean Physical Society, 2014, 64 : 923 - 928
  • [2] Adaptive statistical iterative reconstruction and bismuth shielding for evaluation of dose reduction to the eye and image quality during head CT
    Kim, Myeong Seong
    Choi, Jiwon
    Kim, Sun Young
    Kweon, Dae Cheol
    [J]. JOURNAL OF THE KOREAN PHYSICAL SOCIETY, 2014, 64 (06) : 923 - 928
  • [3] Assessment of dose reduction and image quality by barium composite shielding in head and chest CT
    Lee, Jong-Woong
    Kweon, Dae Cheol
    [J]. RADIATION EFFECTS AND DEFECTS IN SOLIDS, 2020, 175 (9-10): : 809 - 826
  • [4] COMPARISON OF RADIATION DOSE AND IMAGE QUALITY IN HEAD CT SCANS AMONG MULTIDETECTOR CT SCANNERS
    Abuzaid, Mohamed M.
    Elshami, Wiam
    Tekin, H. O.
    Sulieman, Abdelmoneim
    Bradley, D. A.
    [J]. RADIATION PROTECTION DOSIMETRY, 2021, 196 (1-2) : 10 - 16
  • [5] National Survey of Radiation Dose and Image Quality in Adult CT Head Scans in Taiwan
    Lin, Chung-Jung
    Mok, Greta S. P.
    Tsai, Mang-Fen
    Tsai, Wei-Ta
    Yang, Bang-Hung
    Tu, Chun-Yuan
    Wu, Tung-Hsin
    [J]. PLOS ONE, 2015, 10 (06):
  • [6] Eye-lens bismuth shielding in paediatric head CT: artefact evaluation and reduction
    Maria Raissaki
    Kostas Perisinakis
    John Damilakis
    Nicholas Gourtsoyiannis
    [J]. Pediatric Radiology, 2010, 40 : 1748 - 1754
  • [7] Eye-lens bismuth shielding in paediatric head CT: artefact evaluation and reduction
    Raissaki, Maria
    Perisinakis, Kostas
    Damilakis, John
    Gourtsoyiannis, Nicholas
    [J]. PEDIATRIC RADIOLOGY, 2010, 40 (11) : 1748 - 1754
  • [8] SU-F-I-46: Optimizing Dose Reduction in Adult Head CT Protocols While Maintaining Image Quality in Postmortem Head Scans
    Lipnharski, I.
    Carranza, C.
    Quails, N.
    Correa, N.
    Rajderkar, D.
    Bennett, J.
    Rill, L.
    Arreola, M.
    [J]. MEDICAL PHYSICS, 2016, 43 (06) : 3397 - 3397
  • [9] Effect of spineboard and headblocks on the image quality of head CT scans
    Hemmes B.
    Jeukens C.R.L.P.N.
    Al-Haidari A.
    Hofman P.A.M.
    vd Linden E.S.
    Brink P.R.G.
    Poeze M.
    [J]. Emergency Radiology, 2016, 23 (3) : 263 - 268
  • [10] Image quality with reduced mA head CT scans: Comparison to standard head CT scanning
    Bove, P
    Lev, MH
    O'Reilly, CE
    Hunter, GJ
    Hamberg, LM
    Gonzalez, RG
    [J]. RADIOLOGY, 2001, 221 : 346 - 346