Application of a variable filter for presampled modulation transfer function analysis with the edge method

被引:3
|
作者
Higashide R. [1 ,3 ]
Ichikawa K. [2 ]
Kunitomo H. [3 ]
Ohashi K. [3 ]
机构
[1] Graduate School of Medical Science, Kanazawa University, 5-11-80 Kodatsuno, Kanazawa, 920-0942, Ishikawa
[2] Institute of Medical, Pharmaceutical and Health Sciences, Kanazawa University, 5-11-80 Kodatsuno, Kanazawa, 920-0942, Ishikawa
[3] Department of Radiology, Nagoya City University Hospital, 1 Kawasumi, Mizuho-cho, Mizuho-ku, Nagoya, 467-0001, Aichi
关键词
Digital radiography; Edge method; Noise-reduction technique; Presampled modulation transfer function (MTF); Variable filter;
D O I
10.1007/s12194-015-0325-2
中图分类号
学科分类号
摘要
We devised a new noise filtering method to reduce the noise in the line spread function (LSF) for presampled modulation transfer function (MTF) analysis with the edge method. A filter was designed to reduce noise effectively using a position-dependent filter controlled by the boundary frequency b for low-pass filtering, which is calculated by 1/2d (d: distance from the LSF center). In this filtering process, strong filters with very low b can be applied to regions distant from the LSF center, and the region near the LSF center can be maintained simultaneously by a correspondingly high b. Presampled MTF accuracies derived by use of the proposed method and an edge spread function (ESF)-fitting method were compared by use of simulated ESFs with and without noise, resembling a computed radiography (CR) and an indirect-type flat panel detector (FPD), respectively. In addition, the edge images of clinical CR, indirect-type FPD, and direct-type FPD systems were examined. For a simulated ESF without noise, the calculated MTFs of the variable filtering method agreed precisely with the true MTFs. The excellent noise-reduction ability of the variable filter was demonstrated for all simulated noisy ESFs and those of three clinical systems. Although the ESF-fitting method provided excellent noise reduction only for the CR-like simulated ESF with noise, its noise elimination performance could not be demonstrated due to the lesser robustness of the fitting. © 2015, Japanese Society of Radiological Technology and Japan Society of Medical Physics.
引用
收藏
页码:320 / 330
页数:10
相关论文
共 50 条
  • [41] MODULATION TRANSFER FUNCTION AND ITS APPLICATION TO RADIOISOTOPE SCANNING
    CRADDUCK, TD
    PHYSICS IN MEDICINE AND BIOLOGY, 1970, 15 (01): : 149 - &
  • [42] A new analytical edge spread function fitting model for modulation transfer function measurement
    Li, Tiecheng
    Feng, Huajun
    Xu, Zhihai
    CHINESE OPTICS LETTERS, 2011, 9 (03)
  • [43] APPLICATION OF MODULATION TRANSFER FUNCTION TO REDUCE UNSHARPNESS IN SCINTIGRAMS
    OKUMURA, Y
    INTERNATIONAL JOURNAL OF APPLIED RADIATION AND ISOTOPES, 1971, 22 (01): : 49 - &
  • [44] Effect of the Edge Tool on Modulation Transfer Function Measurement in Digital Mammography
    Karaaslan, Melike Kaya
    Gundogdu, Ozcan
    Muzoglu, Nedim
    TIP TEKNOLOJILERI KONGRESI (TIPTEKNO'21), 2021,
  • [45] A new analytical edge spread function fitting model for modulation transfer function measurement
    李铁成
    冯华君
    徐之海
    Chinese Optics Letters, 2011, 9 (03) : 37 - 40
  • [46] Application of transfer function method to dynamic analysis of a nonlocal elastic beam
    School of Aerospace and Material Engineering, National University of Defense Technology, Changsha 410073, China
    J Vib Shock, 2007, 5 (4-7):
  • [47] Modal analysis of cantilever beams with an edge notch using asymptotic transfer function method
    Yu, Zhigang
    Feng, Xiang
    Wu, Xun
    Chu, Fulei
    DYNAMICS OF CONTINUOUS DISCRETE AND IMPULSIVE SYSTEMS-SERIES B-APPLICATIONS & ALGORITHMS, 2007, 14 : 279 - 283
  • [48] A METHOD FOR MODULATION TRANSFER-FUNCTION DETERMINATION FROM EDGE PROFILES WITH CORRECTION FOR FINITE-ELEMENT DIFFERENTIATION
    CUNNINGHAM, IA
    FENSTER, A
    MEDICAL PHYSICS, 1987, 14 (04) : 533 - 537
  • [49] Improved Modulation Transfer Function evaluation method of a camera at any field point with a scanning tilted knife edge
    Homassel, Etienne
    Barrat, Catherine
    Alves, Frederic
    Aubry, Gilles
    Arquetoux, Guillaume
    ELECTRO-OPTICAL AND INFRARED SYSTEMS: TECHNOLOGY AND APPLICATIONS XVII, 2020, 11537
  • [50] A simple edge device method for determining the presampling modulation transfer function (MTF) of flat field digital detectors
    Wang, BL
    Barner, K
    Lee, D
    Rodricks, B
    Liu, CD
    Medical Imaging 2005: Physics of Medical Imaging, Pts 1 and 2, 2005, 5745 : 1191 - 1198