Application of a variable filter for presampled modulation transfer function analysis with the edge method

被引:3
|
作者
Higashide R. [1 ,3 ]
Ichikawa K. [2 ]
Kunitomo H. [3 ]
Ohashi K. [3 ]
机构
[1] Graduate School of Medical Science, Kanazawa University, 5-11-80 Kodatsuno, Kanazawa, 920-0942, Ishikawa
[2] Institute of Medical, Pharmaceutical and Health Sciences, Kanazawa University, 5-11-80 Kodatsuno, Kanazawa, 920-0942, Ishikawa
[3] Department of Radiology, Nagoya City University Hospital, 1 Kawasumi, Mizuho-cho, Mizuho-ku, Nagoya, 467-0001, Aichi
关键词
Digital radiography; Edge method; Noise-reduction technique; Presampled modulation transfer function (MTF); Variable filter;
D O I
10.1007/s12194-015-0325-2
中图分类号
学科分类号
摘要
We devised a new noise filtering method to reduce the noise in the line spread function (LSF) for presampled modulation transfer function (MTF) analysis with the edge method. A filter was designed to reduce noise effectively using a position-dependent filter controlled by the boundary frequency b for low-pass filtering, which is calculated by 1/2d (d: distance from the LSF center). In this filtering process, strong filters with very low b can be applied to regions distant from the LSF center, and the region near the LSF center can be maintained simultaneously by a correspondingly high b. Presampled MTF accuracies derived by use of the proposed method and an edge spread function (ESF)-fitting method were compared by use of simulated ESFs with and without noise, resembling a computed radiography (CR) and an indirect-type flat panel detector (FPD), respectively. In addition, the edge images of clinical CR, indirect-type FPD, and direct-type FPD systems were examined. For a simulated ESF without noise, the calculated MTFs of the variable filtering method agreed precisely with the true MTFs. The excellent noise-reduction ability of the variable filter was demonstrated for all simulated noisy ESFs and those of three clinical systems. Although the ESF-fitting method provided excellent noise reduction only for the CR-like simulated ESF with noise, its noise elimination performance could not be demonstrated due to the lesser robustness of the fitting. © 2015, Japanese Society of Radiological Technology and Japan Society of Medical Physics.
引用
收藏
页码:320 / 330
页数:10
相关论文
共 50 条
  • [21] Edge-based modulation transfer function measurement method using a variable oversampling ratio (vol 29, 37628, 2021)
    Masaoka, Kenichiro
    OPTICS EXPRESS, 2022, 30 (01): : 179 - 179
  • [22] Research on the Modulation Transfer Function Detection Method of a Bayer Filter Color Camera
    Fan, Yuan-Peng
    Wei, Lei
    Li, Lin
    Yang, Lin
    Hu, Zi-Qiang
    Zheng, Yuan-Hao
    Wang, Yu-Hao
    SENSORS, 2023, 23 (09)
  • [23] Modulation Transfer Function Detection Method of Bayer Filter Color Space Camera
    Wang, Yuhao
    Fan, Yuanpeng
    Wei, Lei
    ACTA OPTICA SINICA, 2024, 44 (19)
  • [25] Regularized slanted-edge method for measuring the modulation transfer function of imaging systems
    Xie, Xufen
    Fan, Hongda
    Wang, Anding
    Zou, Nianyu
    Zhang, Yuncui
    APPLIED OPTICS, 2018, 57 (22) : 6552 - 6558
  • [26] Measurement of the Modulation Transfer Function of Infrared Imaging System by Modified Slant Edge Method
    Li, Hang
    Yang, Changxiang
    Shao, Jianbing
    JOURNAL OF THE OPTICAL SOCIETY OF KOREA, 2016, 20 (03) : 381 - 388
  • [27] Effect of readout direction in the edge profile on the modulation transfer function of computed radiographic systems by use of the edge method
    Tanaka N.
    Morishita J.
    Tsuda N.
    Ohki M.
    Radiological Physics and Technology, 2013, 6 (2) : 474 - 479
  • [28] Simulation analysis on the performance of a circular-edge technique in measurements of the modulation transfer function
    Maruyama, Sho
    JOURNAL OF MEDICAL PHYSICS, 2023, 48 (01) : 90 - 97
  • [29] Analysis of the precision of modulation transfer function measurements with a circular edge: influence of object circularity
    Maruyama, Sho
    JOURNAL OF MEDICAL IMAGING, 2023, 10 (04)
  • [30] A moving slanted-edge method to measure the temporal modulation transfer function of fluoroscopic systems
    Friedman, S. N.
    Cunningham, I. A.
    MEDICAL PHYSICS, 2008, 35 (06) : 2473 - 2484