Application of a variable filter for presampled modulation transfer function analysis with the edge method

被引:3
|
作者
Higashide R. [1 ,3 ]
Ichikawa K. [2 ]
Kunitomo H. [3 ]
Ohashi K. [3 ]
机构
[1] Graduate School of Medical Science, Kanazawa University, 5-11-80 Kodatsuno, Kanazawa, 920-0942, Ishikawa
[2] Institute of Medical, Pharmaceutical and Health Sciences, Kanazawa University, 5-11-80 Kodatsuno, Kanazawa, 920-0942, Ishikawa
[3] Department of Radiology, Nagoya City University Hospital, 1 Kawasumi, Mizuho-cho, Mizuho-ku, Nagoya, 467-0001, Aichi
关键词
Digital radiography; Edge method; Noise-reduction technique; Presampled modulation transfer function (MTF); Variable filter;
D O I
10.1007/s12194-015-0325-2
中图分类号
学科分类号
摘要
We devised a new noise filtering method to reduce the noise in the line spread function (LSF) for presampled modulation transfer function (MTF) analysis with the edge method. A filter was designed to reduce noise effectively using a position-dependent filter controlled by the boundary frequency b for low-pass filtering, which is calculated by 1/2d (d: distance from the LSF center). In this filtering process, strong filters with very low b can be applied to regions distant from the LSF center, and the region near the LSF center can be maintained simultaneously by a correspondingly high b. Presampled MTF accuracies derived by use of the proposed method and an edge spread function (ESF)-fitting method were compared by use of simulated ESFs with and without noise, resembling a computed radiography (CR) and an indirect-type flat panel detector (FPD), respectively. In addition, the edge images of clinical CR, indirect-type FPD, and direct-type FPD systems were examined. For a simulated ESF without noise, the calculated MTFs of the variable filtering method agreed precisely with the true MTFs. The excellent noise-reduction ability of the variable filter was demonstrated for all simulated noisy ESFs and those of three clinical systems. Although the ESF-fitting method provided excellent noise reduction only for the CR-like simulated ESF with noise, its noise elimination performance could not be demonstrated due to the lesser robustness of the fitting. © 2015, Japanese Society of Radiological Technology and Japan Society of Medical Physics.
引用
收藏
页码:320 / 330
页数:10
相关论文
共 50 条
  • [11] Measurement of the presampled two-dimensional modulation transfer function of digital imaging systems
    Fetterly, KA
    Hangiandreou, NJ
    Schueler, BA
    Ritenour, ER
    MEDICAL PHYSICS, 2002, 29 (05) : 913 - 921
  • [12] Analysis of the Slanted-Edge Measurement Method for the Modulation Transfer Function of Remote Sensing Cameras
    Yu, Jian
    Zhang, Yu
    Qi, Biao
    Bai, Xiaotian
    Wu, Wei
    Liu, Hongxing
    APPLIED SCIENCES-BASEL, 2023, 13 (24):
  • [13] A method to measure the modulation transfer function of Bayer filter color camera
    Duan Ya-Xuan
    Liu Shang-Kuo
    Chen Yong-Quan
    Xue Xun
    Zhao Jian-Ke
    Gao Li-Min
    ACTA PHYSICA SINICA, 2017, 66 (07)
  • [14] Determination of the modulation transfer function using the edge method: Influence of scattered radiation
    Neitzel, U
    Buhr, E
    Hilgers, G
    Granfors, PR
    MEDICAL PHYSICS, 2004, 31 (12) : 3485 - 3491
  • [15] Modified slanted-edge method and multidirectional modulation transfer function estimation
    Masaoka, Kenichiro
    Yamashita, Takayuki
    Nishida, Yukihiro
    Sugawara, Masayuki
    OPTICS EXPRESS, 2014, 22 (05): : 6040 - 6046
  • [16] The research on Modulation Transfer Function of underwater imaging system based on Edge Method
    Li, Dailin
    Li, Guilei
    Yang, Dan
    Xing, Yi
    Wang, Ning
    Zhu, Huafeng
    Jiao, Zhiyong
    OCEAN OPTICS AND INFORMATION TECHNOLOGY, 2018, 10850
  • [17] A NEW METHOD FOR DETERMINING THE MODULATION-TRANSFER FUNCTION FROM EDGE TRACES
    FANG, L
    TIZIANI, HJ
    OPTIK, 1986, 74 (01): : 17 - 21
  • [18] Modified slanted-edge method to measure the modulation transfer function of camera
    Zhang, Haosu
    Li, Chuang
    Duan, Yaxuan
    OPTIK, 2018, 157 : 635 - 643
  • [19] Error of the slanted edge method for measuring the modulation transfer function of imaging systems
    Xie, Xufen
    Fan, Hongda
    Wang, Hongyuan
    Wang, Zebin
    Zou, Nianyu
    APPLIED OPTICS, 2018, 57 (07) : B83 - B91
  • [20] An experimental study of the accuracy in measurement of modulation transfer function using an edge method
    Lee, Dong-Hoon
    Kim, Ye-seul
    Park, Hye-Suk
    Lee, Young-Jin
    Kim, Hee-Joung
    MEDICAL IMAGING 2015: PHYSICS OF MEDICAL IMAGING, 2015, 9412