Laser Resonance Frequency Analysis: A Novel Measurement Approach to Evaluate Acetabular Cup Stability During Surgery

被引:13
|
作者
Kikuchi, Shunsuke [1 ]
Mikami, Katsuhiro [2 ]
Nakashima, Daisuke [1 ]
Kitamura, Toshiyuki [3 ]
Hasegawa, Noboru [3 ]
Nishikino, Masaharu [3 ]
Kanaji, Arihiko [1 ]
Nakamura, Masaya [1 ]
Nagura, Takeo [1 ]
机构
[1] Keio Univ, Dept Orthopaed Surg, Sch Med, Shinjuku Ku, 35 Shinanomachi, Tokyo 1608582, Japan
[2] Kindai Univ, Dept Biol Oriented Sci & Technol, 930 Nishi Mitani, Wakayama 6496493, Japan
[3] Natl Inst Quantum & Radiol Sci & Technol, Kansai Photon Sci Inst, 817 Umemidai, Kizugawa City, Kyoto 6190215, Japan
基金
日本学术振兴会;
关键词
total hip arthroplasty; implant stability; acetabular cup; laser; resonance frequency analysis; finite element method; TOTAL HIP-ARTHROPLASTY; ON-METAL BEARINGS; IMPLANT STABILITY; INITIAL STABILITY; FIXATION; FIT; VIVO;
D O I
10.3390/s19224876
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
Artificial joint acetabular cup stability is essential for successful total hip arthroplasty. However, a quantitative evaluation approach for clinical use is lacking. We developed a resonance frequency analysis (RFA) system involving a laser system that is fully contactless. This study aimed to investigate the usefulness of laser RFA for evaluating acetabular cup stability. First, the finite element method was performed to determine the vibration mode for analysis. Second, the acetabular cup was press-fitted into a reamed polyurethane cavity that replicated the human acetabular roof. The implanted acetabular cup was vibrated with pulse laser irradiation and the induced vibration was detected with a laser Doppler vibrometer. The time domain signal from the vibrometer was analyzed by fast Fourier transform to obtain the vibration frequency spectrum. After laser RFA, the pull-down force of the acetabular cup was measured as conventional implant fixation strength. The frequency of the first highest amplitude between 2 kHz and 6 kHz was considered as the resonance peak frequency, and its relationship with the pull-down force was assessed. The peak frequency could predict the pull-down force (R-2 = 0.859, p < 0.000). Our findings suggest that laser RFA might be useful to measure acetabular cup stability during surgery.
引用
收藏
页数:12
相关论文
共 50 条
  • [1] Evaluation of bone integrity around the acetabular cup using noninvasive laser resonance frequency analysis
    Hatakeyama, Takuto
    Nakashima, Daisuke
    Mikami, Katsuhiro
    Oya, Akihito
    Fujie, Atsuhiro
    Sujino, Asahi
    Nakamura, Masaya
    Nagura, Takeo
    JOURNAL OF ORTHOPAEDIC RESEARCH, 2024, 42 (11) : 2552 - 2561
  • [2] Influence of the orientation of the Osstell® transducer during measurement of dental implant stability using resonance frequency analysis: A numerical approach
    Capek, Lukas
    Simunek, Antonin
    Slezak, Radovan
    Dzan, Ladislav
    MEDICAL ENGINEERING & PHYSICS, 2009, 31 (07) : 764 - 769
  • [3] Experimental analysis of the deformation behaviour and fixation stability of a novel monolithic acetabular cup made of composite ceramics
    Schulze, C.
    Dammer, R.
    Begand, S.
    Oberbach, T.
    Bader, R.
    BIOMEDICAL ENGINEERING-BIOMEDIZINISCHE TECHNIK, 2014, 59 : S1140 - S1143
  • [4] Reliability of the Osstell Implant Stability Quotient and Penguin Resonance Frequency Analysis to Evaluate Implant Stability
    Buyukguclu, Gul
    Ozkurt-Kayahan, Zeynep
    Kazazoglu, Ender
    IMPLANT DENTISTRY, 2018, 27 (04) : 429 - 433
  • [5] Laser frequency stability: a simple approach for a quantitative analysis
    Cabral, Alexandre
    Abreu, Manuel
    Rebordao, Jose M.
    INTERNATIONAL CONFERENCE ON APPLICATIONS OF OPTICS AND PHOTONICS, 2011, 8001
  • [6] Jumbo Bionic Trabecular Metal Acetabular Cups Improve Cup Stability During Acetabular Bone Defect Reconstruction: A Finite Element Analysis Study
    Xiao, Jianlin
    Xue, Haowen
    Qian, Zhihui
    Xu, Shenghao
    Shen, Xianyue
    Zhang, Zhuo
    Ren, Lei
    Ren, Luquan
    JOURNAL OF BIONIC ENGINEERING, 2023, 20 (06) : 2814 - 2825
  • [7] Jumbo Bionic Trabecular Metal Acetabular Cups Improve Cup Stability During Acetabular Bone Defect Reconstruction: A Finite Element Analysis Study
    Jianlin Xiao
    Haowen Xue
    Zhihui Qian
    Shenghao Xu
    Xianyue Shen
    Zhuo Zhang
    Lei Ren
    Luquan Ren
    Journal of Bionic Engineering, 2023, 20 : 2814 - 2825
  • [8] Measurement of mini-implant stability using resonance frequency analysis
    Nienkemper, Manuel
    Wilmes, Benedict
    Panayotidis, Agamemnon
    Pauls, Alexander
    Golubovic, Vladimir
    Schwarz, Frank
    Drescher, Dieter
    ANGLE ORTHODONTIST, 2013, 83 (02) : 230 - 238
  • [9] Measurement of dental implant stability by resonance frequency analysis: A review of the literature
    Pilar Quesada-Garcia, Ma
    Prados-Sanchez, Estrella
    Victoria Olmedo-Gaya, Ma
    Munoz-Soto, Esther
    Paloma Gonzalez-Rodriguez, Ma
    Vallecillo-Capilla, Manuel
    MEDICINA ORAL PATOLOGIA ORAL Y CIRUGIA BUCAL, 2009, 14 (10): : E538 - E546
  • [10] Measurement of Primary Stability of Mini Implants Using Resonance Frequency Analysis
    Singh, Ashish Kumar
    Kannan, Sridhar
    Arora, Nitin
    Bajaj, Yatin
    Revankar, Ameet V.
    APOS TRENDS IN ORTHODONTICS, 2018, 8 (03) : 139 - 145