The efficacy of the arrow staple device for securing central venous catheters to human skin

被引:7
|
作者
Motonaga, GK [1 ]
Lee, KK [1 ]
Kirsch, JR [1 ]
机构
[1] Johns Hopkins Med Inst, Dept Anesthesiol & Crit Care Med, Baltimore, MD 21287 USA
来源
ANESTHESIA AND ANALGESIA | 2004年 / 99卷 / 05期
关键词
D O I
10.1213/01.ANE.0000132544.15467.10
中图分类号
R614 [麻醉学];
学科分类号
100217 ;
摘要
We tested the hypothesis that the force necessary to dislodge a central venous catheter from human skin is related to its method of attachment. Specifically, we compared the peak axial force and torque required to remove a catheter hub attached to human skin with 2 3-0 silk sutures, 4 0.022-in. staples, or 4 0.025-in. staples. We used the tissue from the proximal end of discarded extremities after below-knee amputations to serve as a surrogate for the skin of living humans that would normally be the site for catheter stabilization. Central venous catheter hubs were secured to the area of grossly normal appearing tissue with 2 3-0 silk sutures or 4 staples. For each attachment method, 16 dislodgement trials were conducted. The mean peak force required for axial dislodgement with 3-0 silk sutures (40.9 +/- 10.7 N; n = 16) was more than that for 40.022-in. staples (34.0 +/- 7.2 N; n = 16; P = 0.04) but was not different from that for 4 0.025-in. staples (40.4 +/- 5.8 N; n = 16). The major finding of this study is that securing a central venous hub with 4 0.025-in. staples is as effective as 2 3-0 silk sutures in preventing axial dislodgement.
引用
收藏
页码:1436 / 1439
页数:4
相关论文
共 50 条
  • [31] Power injection of contrast media using central venous catheters: Feasibility, safety, and efficacy
    Herts, BR
    O'Malley, CM
    Wirth, SL
    Lieber, ML
    Pohlman, B
    AMERICAN JOURNAL OF ROENTGENOLOGY, 2001, 176 (02) : 447 - 453
  • [32] Antimicrobial Central Venous Catheters in Oncology: Efficacy of a Rifampicin-Miconazole-releasing Catheter
    Schierholz, Joerg Michael
    Nagelschmidt, Klaus
    Nagelschmidt, Manfred
    Lefering, Rolf
    Yuecel, Nedim
    Beuth, Josef
    ANTICANCER RESEARCH, 2010, 30 (04) : 1353 - 1358
  • [33] Efficacy and Safety of a New Technique of Conversion from Temporary to Tunneled Central Venous Catheters
    Lomonte, Carlo
    Libutti, Pasquale
    Casucci, Francesco
    Lisi, Piero
    Basile, Carlo
    SEMINARS IN DIALYSIS, 2015, 28 (04) : 435 - 438
  • [34] Efficacy and safety of peripherally inserted central venous catheters in acute cardiac care management
    Poletti, Fabrizio
    Coccino, Claudio
    Monolo, Davide
    Crespi, Paolo
    Ciccioli, Giorgio
    Cordio, Giuseppe
    Seveso, Giovanni
    De Servi, Stefano
    JOURNAL OF VASCULAR ACCESS, 2018, 19 (05): : 455 - 460
  • [35] Occlusion Reduction and Heparin Elimination Trial Using an Antireflux Device on Peripheral and Central Venous Catheters
    Jasinsky, Lisa M.
    Wurster, Julie
    JOURNAL OF INFUSION NURSING, 2009, 32 (01) : 33 - 39
  • [36] Histologic changes in the human vein wall adjacent to indwelling central venous catheters
    Forauer, AR
    Theoharis, C
    JOURNAL OF VASCULAR AND INTERVENTIONAL RADIOLOGY, 2003, 14 (09) : 1163 - 1168
  • [38] POSITIONING THE TIP OF CENTRAL VENOUS CATHETERS BY THE INTRACAVITARY ECG METHOD, USING A NEW SPECIFIC DEVICE
    Pittiruti, M.
    La Greca, A.
    Emoli, A.
    Scoppettuolo, G.
    INTENSIVE CARE MEDICINE, 2010, 36 : S145 - S145
  • [39] Misplacement of Central Venous Catheters in the Subclavian Artery: Safe Retrieval with a Percutaneous Closure Device (AngioSeal)
    Bangard, C.
    Chang, D-H.
    Libicher, M.
    Bovenschulte, H.
    Kochanek, M.
    Reuter, H.
    Maintz, D.
    ROFO-FORTSCHRITTE AUF DEM GEBIET DER RONTGENSTRAHLEN UND DER BILDGEBENDEN VERFAHREN, 2013, 185 (06): : 546 - 549
  • [40] Evaluation of the efficacy and safety of a novel dermatotomy device for central venous cannulation
    Keeyapaj, Worasak
    Cheung, Albert T.
    ANESTHESIA AND ANALGESIA, 2021, 132 (5S_SUPPL): : 994 - 994