The efficacy of the arrow staple device for securing central venous catheters to human skin

被引:7
|
作者
Motonaga, GK [1 ]
Lee, KK [1 ]
Kirsch, JR [1 ]
机构
[1] Johns Hopkins Med Inst, Dept Anesthesiol & Crit Care Med, Baltimore, MD 21287 USA
来源
ANESTHESIA AND ANALGESIA | 2004年 / 99卷 / 05期
关键词
D O I
10.1213/01.ANE.0000132544.15467.10
中图分类号
R614 [麻醉学];
学科分类号
100217 ;
摘要
We tested the hypothesis that the force necessary to dislodge a central venous catheter from human skin is related to its method of attachment. Specifically, we compared the peak axial force and torque required to remove a catheter hub attached to human skin with 2 3-0 silk sutures, 4 0.022-in. staples, or 4 0.025-in. staples. We used the tissue from the proximal end of discarded extremities after below-knee amputations to serve as a surrogate for the skin of living humans that would normally be the site for catheter stabilization. Central venous catheter hubs were secured to the area of grossly normal appearing tissue with 2 3-0 silk sutures or 4 staples. For each attachment method, 16 dislodgement trials were conducted. The mean peak force required for axial dislodgement with 3-0 silk sutures (40.9 +/- 10.7 N; n = 16) was more than that for 40.022-in. staples (34.0 +/- 7.2 N; n = 16; P = 0.04) but was not different from that for 4 0.025-in. staples (40.4 +/- 5.8 N; n = 16). The major finding of this study is that securing a central venous hub with 4 0.025-in. staples is as effective as 2 3-0 silk sutures in preventing axial dislodgement.
引用
收藏
页码:1436 / 1439
页数:4
相关论文
共 50 条
  • [11] THE CHEST TUBE - A SIMPLE DEVICE FOR THE TUNNELING OF CENTRAL VENOUS CATHETERS
    HOELZER, DJ
    TUGGLE, DW
    JUDD, AT
    SMITH, EI
    JOURNAL OF PARENTERAL AND ENTERAL NUTRITION, 1987, 11 (03) : 326 - 327
  • [12] A prospective trial on a new sutureless securement device for central venous catheters
    Daniel Cordovani
    Richard M. Cooper
    Canadian Journal of Anesthesia/Journal canadien d'anesthésie, 2013, 60 : 504 - 505
  • [13] HAEMOCATCH®: A NEW DEVICE FOR THE MANAGEMENT OF CVC (CENTRAL VENOUS CATHETERS) FOR HEMODIALYSIS
    Mambelli, Emanuele
    Mancini, Elena
    Elia, Cinzia
    Guadagno, Vincenza
    Facchini, Maria Grazia
    Zucchelli, Annalisa
    Grazia, Michele
    Patregnani, Laura
    Santoro, Antonio
    NEPHROLOGY DIALYSIS TRANSPLANTATION, 2012, 27 : 253 - 253
  • [14] A prospective trial on a new sutureless securement device for central venous catheters
    Cordovani, Daniel
    Cooper, Richard M.
    CANADIAN JOURNAL OF ANESTHESIA-JOURNAL CANADIEN D ANESTHESIE, 2013, 60 (05): : 504 - 505
  • [15] THE EFFECT OF MEDICINAL HONEY ON SKIN COLONIZATION AROUND CENTRAL VENOUS CATHETERS
    Mueller, M. C. A.
    Kwakman, P. H. S.
    de Borgie, C. A. J. M.
    Binnekade, J. M.
    Schultz, M. J.
    Zaat, S. A. J.
    INTENSIVE CARE MEDICINE, 2011, 37 : S80 - S80
  • [16] Efficacy of peripherally inserted central venous catheters placed in noncentral veins
    Thiagarajan, RR
    Bratton, SL
    Gettmann, T
    Ramamoorthy, C
    ARCHIVES OF PEDIATRICS & ADOLESCENT MEDICINE, 1998, 152 (05): : 436 - 439
  • [17] Efficacy of antibiotic-coated central venous catheters - The author replies
    Raad, I
    CRITICAL CARE MEDICINE, 1999, 27 (06) : 1217 - 1217
  • [18] Sutureless securement device reduces complications of peripherally inserted central venous catheters
    Yamamoto, AJ
    Solomon, JA
    Soulen, MC
    Tang, J
    Parkinson, K
    Lin, R
    Schears, GJ
    JOURNAL OF VASCULAR AND INTERVENTIONAL RADIOLOGY, 2002, 13 (01) : 77 - 81
  • [19] Are skin-tunnelled central venous catheters essential for myeloma autograft transplant?
    Bishop, L.
    Collosio, M.
    BONE MARROW TRANSPLANTATION, 2007, 39 : S264 - S265
  • [20] 5% NaHCO3 Is Appropriate for Skin Cleaning With Central Venous Catheters
    Wu, Hong-lei
    Xu, Yang-hui
    Shi, Jia-hai
    AMERICAN JOURNAL OF THE MEDICAL SCIENCES, 2017, 353 (01): : 12 - 16