Noninvasive Blood Pressure Measurement Using the NICCI Monitor in Adolescents During Intraoperative Anesthetic Care

被引:3
|
作者
Owusu-Bediako, Kwaku [1 ]
Bekiroglu, Ismail [1 ]
Rice-Weimer, Julie [1 ]
Murillo-Deluquez, Marcelino [1 ]
Tobias, Joseph D. [1 ,2 ]
机构
[1] Nationwide Childrens Hosp, Dept Anesthesiol & Pain Med, Columbus, OH 43205 USA
[2] Ohio State Univ, Dept Anesthesiol & Pain Med, Columbus, OH 43210 USA
关键词
Continuous blood pressure monitoring; Invasive arterial pressure; Intraoperative monitoring; Pediatrics anesthesiology; DEVICE; ACCURACY; CNAP(TM); SURGERY;
D O I
10.14740/cr1378
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Background: The intermittent measurement of blood pressure (BP) remains the standard of care during anesthesia or procedural sedation. To improve the early identification of hemodynamic compromise, various noninvasive BP devices have been developed which provide a continuous BP reading. The current study evaluates the accuracy of a novel continuous BP device, the NICCI system, in adolescents weighing 40 - 80 kg. Methods: During intraoperative anesthetic care, BP readings (systolic, diastolic, and mean) were captured from the arterial cannula and the NICC1 device every second. Results: The study cohort included 44 pediatric patients undergoing major orthopedic, cardiac, and neurosurgical procedures. A total of 383,126 pairs of systolic blood pressure (SBP), diastolic blood pressure (DBP), and mean arterial pressure (MAP) values from the arterial cannula and the MCCI device were analyzed. The absolute difference for SBP, DBP, and MAP values from the MCCI monitor and the arterial cannula were 10 +/- 8, 9 +/- 7, and 9 +/- 7 mm Hg, respectively. The difference between the BP values from the MCCI and the arterial cannula was <= 10 mm Hg for 60% of the SBP readings, 67% of the DBP readings, and 56% of the MAP readings. Using Bland-Altman analysis, the bias was 2, 3, and 4 mm Hg for the SBP, DBP, and MAP. Conclusions: Although there were technical limitations related to patient size that affected its ability to meet the strict accuracy criteria set by the American National Standards Institute/Association for the Advancement of Medical Instrumentation standards for noninvasive BP measurement (ANSI/AAMI SP10), the NICCI system provided a continuous noninvasive beat-to-beat BP measurement which was clinically relevant during a significant portion of intraoperative care.
引用
收藏
页码:154 / 161
页数:8
相关论文
共 50 条
  • [42] Intraoperative Blood Pressure Misclassification Due to Inaccuracy of Noninvasive Oscillometric Cuff-Measured Blood Pressure
    Kim, Danny Sungsoo
    Kwon, Sohee
    Bovik, Alan C.
    Markey, Mia K.
    Giovannucci, Edward L.
    Cannesson, Maxime
    HYPERTENSION, 2025, 82 (02) : e1 - e3
  • [43] NONINVASIVE CONTINUOUS FINGER BLOOD-PRESSURE MEASUREMENT DURING ORTHOSTATIC STRESS COMPARED TO INTRAARTERIAL PRESSURE
    IMHOLZ, BPM
    SETTELS, JJ
    VANDERMEIRACKER, AH
    WESSELING, KH
    WIELING, W
    CARDIOVASCULAR RESEARCH, 1990, 24 (03) : 214 - 221
  • [44] NONINVASIVE, CONTINUOUS BLOOD-PRESSURE MEASUREMENT BY ARTERIAL TONOMETRY DURING ANESTHESIA IN CHILDREN
    KEMMOTSU, O
    OHNO, M
    TAKITA, K
    SUGIMOTO, H
    OTSUKA, H
    MORIMOTO, Y
    MAYUMI, T
    ANESTHESIOLOGY, 1994, 81 (05) : 1162 - 1168
  • [45] NONINVASIVE, CONTINUOUS BLOOD-PRESSURE MEASUREMENT BY ARTERIAL TONOMETRY DURING PEDIATRIC ANESTHESIA
    OHNO, M
    OHTSUKA, K
    UEDA, M
    KEMMOTSU, O
    ANESTHESIA AND ANALGESIA, 1993, 76 (02): : S308 - S308
  • [46] A novel art of continuous noninvasive blood pressure measurement
    Jürgen Fortin
    Dorothea E. Rogge
    Christian Fellner
    Doris Flotzinger
    Julian Grond
    Katja Lerche
    Bernd Saugel
    Nature Communications, 12
  • [47] NONINVASIVE, CONTINUOUS BLOOD-PRESSURE MEASUREMENT IN PSYCHOPHYSIOLOGY
    GODAERT, G
    KRUITWAGEN, J
    JASPERS, J
    PSYCHOPHYSIOLOGY, 1985, 22 (05) : 591 - 592
  • [48] Validity and application of noninvasive measurement of blood pressure in hamsters
    Kato, Y
    Iwase, M
    Kanazawa, H
    Nishizawa, T
    Zhao, YL
    Takagi, K
    Nagata, K
    Noda, A
    Koike, Y
    Yokota, M
    EXPERIMENTAL ANIMALS, 2003, 52 (04) : 359 - 363
  • [50] Accuracy improvement of noninvasive finger blood pressure measurement
    Lackovic, I
    Santic, A
    MEDICON 2001: PROCEEDINGS OF THE INTERNATIONAL FEDERATION FOR MEDICAL & BIOLOGICAL ENGINEERING, PTS 1 AND 2, 2001, : 281 - 284