Noninvasive Blood Pressure Measurement Using the NICCI Monitor in Adolescents During Intraoperative Anesthetic Care

被引:3
|
作者
Owusu-Bediako, Kwaku [1 ]
Bekiroglu, Ismail [1 ]
Rice-Weimer, Julie [1 ]
Murillo-Deluquez, Marcelino [1 ]
Tobias, Joseph D. [1 ,2 ]
机构
[1] Nationwide Childrens Hosp, Dept Anesthesiol & Pain Med, Columbus, OH 43205 USA
[2] Ohio State Univ, Dept Anesthesiol & Pain Med, Columbus, OH 43210 USA
关键词
Continuous blood pressure monitoring; Invasive arterial pressure; Intraoperative monitoring; Pediatrics anesthesiology; DEVICE; ACCURACY; CNAP(TM); SURGERY;
D O I
10.14740/cr1378
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Background: The intermittent measurement of blood pressure (BP) remains the standard of care during anesthesia or procedural sedation. To improve the early identification of hemodynamic compromise, various noninvasive BP devices have been developed which provide a continuous BP reading. The current study evaluates the accuracy of a novel continuous BP device, the NICCI system, in adolescents weighing 40 - 80 kg. Methods: During intraoperative anesthetic care, BP readings (systolic, diastolic, and mean) were captured from the arterial cannula and the NICC1 device every second. Results: The study cohort included 44 pediatric patients undergoing major orthopedic, cardiac, and neurosurgical procedures. A total of 383,126 pairs of systolic blood pressure (SBP), diastolic blood pressure (DBP), and mean arterial pressure (MAP) values from the arterial cannula and the MCCI device were analyzed. The absolute difference for SBP, DBP, and MAP values from the MCCI monitor and the arterial cannula were 10 +/- 8, 9 +/- 7, and 9 +/- 7 mm Hg, respectively. The difference between the BP values from the MCCI and the arterial cannula was <= 10 mm Hg for 60% of the SBP readings, 67% of the DBP readings, and 56% of the MAP readings. Using Bland-Altman analysis, the bias was 2, 3, and 4 mm Hg for the SBP, DBP, and MAP. Conclusions: Although there were technical limitations related to patient size that affected its ability to meet the strict accuracy criteria set by the American National Standards Institute/Association for the Advancement of Medical Instrumentation standards for noninvasive BP measurement (ANSI/AAMI SP10), the NICCI system provided a continuous noninvasive beat-to-beat BP measurement which was clinically relevant during a significant portion of intraoperative care.
引用
收藏
页码:154 / 161
页数:8
相关论文
共 50 条
  • [31] Accuracy of the CNAP™ monitor, a noninvasive continuous blood pressure device, in providing beat-to-beat blood pressure measurements during bariatric surgery in severely obese adolescents and young adults
    Joseph D. Tobias
    Chris McKee
    Dan Herz
    Steve Teich
    Paul Sohner
    Julie Rice
    N’Diris Barry
    Marc Michalsky
    Journal of Anesthesia, 2014, 28 : 861 - 865
  • [32] Accuracy of the CNAP™ monitor, a noninvasive continuous blood pressure device, in providing beat-to-beat blood pressure measurements during bariatric surgery in severely obese adolescents and young adults
    Tobias, Joseph D.
    McKee, Chris
    Herz, Dan
    Teich, Steve
    Sohner, Paul
    Rice, Julie
    Barry, N'Diris
    Michalsky, Marc
    JOURNAL OF ANESTHESIA, 2014, 28 (06) : 861 - 865
  • [33] Reliability of an automatic monitor for blood pressure measurement
    Cardona-Mueller, David
    Grover-Paez, Fernando
    Guzman-Saldivar, Victor
    Alanis-Sanchez, Guillermo A.
    Murguia-Soto, Cesar
    Totsuka-Sutto, Sylvia E.
    Quezada-Fernandez, Patricia
    Macias-Chumacera, Agustin
    Cardona-Munoz, Ernesto G.
    Ramos-Becerra, Carlos G.
    REVISTA MEDICA DE CHILE, 2018, 146 (02) : 190 - 195
  • [34] Atrial fibrillation detection in primary care during blood pressure measurements and using a smartphone cardiac monitor
    John D. Sluyter
    Robert Scragg
    Malakai ‘Ofanoa
    Ralph A. H. Stewart
    Scientific Reports, 11
  • [35] Atrial fibrillation detection in primary care during blood pressure measurements and using a smartphone cardiac monitor
    Sluyter, John D.
    Scragg, Robert
    Ofanoa, Malakai
    Stewart, Ralph A. H.
    SCIENTIFIC REPORTS, 2021, 11 (01)
  • [36] Noninvasive blood pressure monitor using strain gauges, a fastening band, and a wrist elasticity model
    Wang, Yu-Jen
    Chen, Tzung-Yu
    Tsai, Meng-Chiao
    Wu, Chih Hsun
    SENSORS AND ACTUATORS A-PHYSICAL, 2016, 252 : 198 - 208
  • [37] Blood pressure measurement in children and adolescents
    Kardas, Przemyslaw
    FAMILY MEDICINE AND PRIMARY CARE REVIEW, 2008, 10 (02): : 226 - 230
  • [38] Potential errors in the measurement of anesthetic partial pressure in blood
    Lockwood, GG
    Dziersk, J
    Sapsed-Byrne, SM
    ANESTHESIOLOGY, 1999, 91 (05) : 1550 - 1553
  • [39] NONINVASIVE BLOOD-PRESSURE MEASUREMENT ON THE TEMPORAL ARTERY USING THE AUSCULTATORY METHOD
    SHENOY, D
    VONMALTZAHN, WW
    BUCKEY, JC
    ANNALS OF BIOMEDICAL ENGINEERING, 1993, 21 (04) : 351 - 360
  • [40] Noninvasive Cuffless Estimation of Blood Pressure using Photoplethysmography without Electrocardiograph Measurement
    Samria, Rohan
    Jain, Ridhi
    Jha, Ankita
    Saini, Sandeep
    Chowdhury, Shubhajit Roy
    2014 IEEE REGION 10 SYMPOSIUM, 2014, : 254 - 257