Design of a 3D printed coronary artery model for CT optimization

被引:16
|
作者
Morup, S. D. [1 ,2 ,3 ]
Stowe, J. [3 ]
Precht, H. [1 ,4 ,5 ]
Gervig, M. H. [1 ]
Foley, S. [3 ]
机构
[1] UCL Univ Coll, Hlth Sci Res Ctr, Niels Bohrs Alle 1, DK-5230 Odense M, Denmark
[2] Odense Univ Hosp, Cardiol Res Dept, Baagoes Alle 15, DK-5700 Svendborg, Denmark
[3] Univ Coll Dublin, Sch Med, Radiog & Diagnost Imaging, Dublin, Ireland
[4] Univ Southern Denmark, Dept Clin Res, Winslowspk 5000, DK-5000 Odense C, Denmark
[5] Hosp Little Belt Kolding, Dept Radiol, Kolding, Denmark
关键词
Computed tomography; Phantom; Three-dimensional printing; Iterative reconstruction; Image quality; Patient safety;
D O I
10.1016/j.radi.2021.09.001
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Introduction: To design a custom phantom of the coronary arteries to optimize CT coronary angiography (CCTA) protocols. Methods: Characteristics of the left and right coronary arteries (mean Hounsfield Unit (HU) values and diameters) were collected from consecutive CCTA examinations (n 1/4 43). Four different materials (two mixtures of glycerine, gelatine and water, pig hearts, EcoflexTM silicone) were scanned inside a Lungman phantom using the CCTA protocol to find the closest model to in vivo data. A 3D printed model of the coronary artery tree was created using CCTA data by exporting a CT volume rendering into Autodesk MeshmixerTM software. The model was placed in an acid bath for 5 h, then covered in EcoflexTM, which was removed after drying. Both the EcoflexTM and pig heart were later filled with a mixture of contrast (Visipaque 320 mg I/ml), NaCl and gelatin and scanned with different levels of tube current and iterative reconstruction (ASiR-V). Objective (HU, noise and size (vessel diameter) and subjective analysis were performed on all scans. Results: The gelatine mixtures had HU values of 130 and 129, EcoflexTM 65 and the pig heart 56. At the different mA/ASiR-V levels the contrast filled EcoflexTM had a mean HU 318 +/- 4, noise 47 +/- 7HU and diameter of 4.4 mm. The pig heart had a mean HU of 209 +/- 5, noise 38 +/- 4HU and a diameter of 4.4 mm. With increasing iterative reconstruction level the visualisation of the pig heart arteries decreased so no measurements could be performed. Conclusion: The use of a 3D printed model of the arteries and casting with the EcoflexTM silicone is the most suitable solution for a custom-designed phantom. Implications for practice: Custom designed phantoms using 3D printing technology enable cost effective optimisation of CT protocols. (c) 2021 The College of Radiographers. Published by Elsevier Ltd. This is an open access article under the CC BY license (http://creativecommons.org/licenses/by/4.0/).
引用
收藏
页码:426 / 432
页数:7
相关论文
共 50 条
  • [1] Design and optimization of 3D fast printed cellular structures
    Collini L.
    Ursini C.
    Kumar A.
    Material Design and Processing Communications, 2021, 3 (04):
  • [2] CT-Derived 3D Printing for Coronary Artery Cannulation Simulator Design Manufacturing
    Etami, Helvina Vika
    Rismawanti, Rochmi Isnaini
    Hanifah, Vita Arfiana Nur
    Herianto, Herianto
    Yanuar, Yarabisa
    Kuswanto, Djoko
    Anggrahini, Dyah Wulan
    Gharini, Putrika Prastuti Ratna
    BIOENGINEERING-BASEL, 2022, 9 (08):
  • [3] Design optimization of 3D printed concrete elements considering buildability
    Mogra, Mihir
    Asaf, Ofer
    Sprecher, Aaron
    Amir, Oded
    ENGINEERING STRUCTURES, 2023, 294
  • [4] A new optimization model for 3D well design
    Shokir, EMEM
    Emera, MK
    Eid, SM
    Wally, AW
    OIL AND GAS SCIENCE AND TECHNOLOGY-REVUE D IFP ENERGIES NOUVELLES, 2004, 59 (03): : 255 - 266
  • [5] Personalized 3D printed coronary models in coronary stenting
    Sun, Zhonghua
    Jansen, Shirley
    QUANTITATIVE IMAGING IN MEDICINE AND SURGERY, 2019, 9 (08) : 1356 - 1367
  • [6] Computational optimization of a 3D printed collimator
    Islam, Fahima
    Lin, Jiao
    Huegle, Thomas
    Lumsden, Ian
    Anderson, David
    Elliott, Amy
    Haberl, Bianca
    Granroth, Garrett
    JOURNAL OF NEUTRON RESEARCH, 2020, 22 (2-3) : 155 - 168
  • [7] Design and Optimization of a 3D Printed Distal Flexible Joint for Endoscopic Surgery
    Song, Yongxiang
    Wang, Shuxin
    Luo, Xiangyu
    Shi, Chaoyang
    IEEE TRANSACTIONS ON MEDICAL ROBOTICS AND BIONICS, 2022, 4 (01): : 38 - 49
  • [8] A 3D printed photosynthesis model
    Smiar, Karen
    Mendez, James
    ABSTRACTS OF PAPERS OF THE AMERICAN CHEMICAL SOCIETY, 2018, 255
  • [9] The optimization of parameters and matching point pairs in the 3D reconstruction of coronary artery
    Tong, Jijun
    Wang, Fangliang
    Li, Meng
    Xia, Shudong
    Lin, Wenting
    BIOMEDICAL SIGNAL PROCESSING AND CONTROL, 2021, 67
  • [10] 3D printed models in patients with coronary artery fistulae: anatomical assessment and interventional planning
    Forte, Mari Nieves Velasco
    Byrne, Nick
    Perez, Israel Valverde
    Bell, Aaron
    Gomez-Ciriza, Gorka
    Krasemann, Thomas
    Sievert, Horst
    Simpson, John
    Pushparajah, Kuberan
    Razavi, Reza
    Qureshi, Shakeel
    Hussain, Tarique
    EUROINTERVENTION, 2017, 13 (09) : 1080 - 1083