Sensitivity calibration procedures in optical-CT scanning of BANG®3 polymer gel dosimeters

被引:25
|
作者
Xu, Y. [1 ]
Wuu, Cheng-Shie [1 ]
Maryanski, Marek J. [1 ,2 ]
机构
[1] Columbia Univ, Dept Radiat Oncol, New York, NY 10032 USA
[2] MGS Res Inc, Madison, CT 06443 USA
基金
美国国家卫生研究院;
关键词
calibration; dosimeters; dosimetry; optical tomography; phantoms; polymer gels; RADIATION-DOSE DISTRIBUTIONS; COMPUTED-TOMOGRAPHY; VERIFICATION; THERAPY; MRI; RADIOSURGERY; RADIOTHERAPY; TEMPERATURE; PERFORMANCE; PHANTOM;
D O I
10.1118/1.3298017
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
The dose response of the BANG (R) 3 polymer gel dosimeter (MGS Research Inc., Madison, CT) was studied using the OCTOPUS (TM) laser CT scanner (MGS Research Inc., Madison, CT). Six 17 cm diameter and 12 cm high Barex cylinders, and 18 small glass vials were used to house the gel. The gel phantoms were irradiated with 6 and 10 MV photons, as well as 12 and 16 MeV electrons using a Varian Clinac 2100EX. Three calibration methods were used to obtain the dose response curves: (a) Optical density measurements on the 18 glass vials irradiated with graded doses from 0 to 4 Gy using 6 or 10 MV large field irradiations; (b) optical-CT scanning of Barex cylinders irradiated with graded doses (0.5, 1, 1.5, and 2 Gy) from four adjacent 4x4 cm(2) photon fields or 6x6 cm(2) electron fields; and (c) percent depth dose (PDD) comparison of optical-CT scans with ion chamber measurements for 6x6 cm(2), 12 and 16 MeV electron fields. The dose response of the BANG (R) 3 gel was found to be linear and energy independent within the uncertainties of the experimental methods (about 3%). The slopes of the linearly fitted dose response curves (dose sensitivities) from the four field irradiations (0.0752 +/- 3%, 0.0756 +/- 3%, 0.0767 +/- 3%, and 0.0759 +/- 3% cm(-1) Gy(-1)) and the PDD matching methods (0.0768 +/- 3% and 0.0761 +/- 3% cm(-1) Gy(-1)) agree within 2.2%, indicating a good reproducibility of the gel dose response within phantoms of the same geometry. The dose sensitivities from the glass vial approach are different from those of the cylindrical Barex phantoms by more than 30%, owing probably to the difference in temperature inside the two types of phantoms during gel formation and irradiation, and possible oxygen contamination of the glass vial walls. The dose response curve obtained from the PDD matching approach with 16 MeV electron field was used to calibrate the gel phantom irradiated with the 12 MeV, 6x6 cm(2) electron field. Three-dimensional dose distributions from the gel measurement and the Eclipse planning system (Varian Corporation, Palo Alto, CA) were compared and evaluated using 3% dose difference and 2 mm distance-to-agreement criteria.
引用
收藏
页码:861 / 868
页数:8
相关论文
共 50 条
  • [41] Optical analysis of gel dosimeters: Comparison of Fricke and normoxic polymer gels
    Gambarini, G.
    Carrara, M.
    Mariani, A.
    Pirola, L.
    Tomatis, S.
    Valente, M.
    Vanossi, E.
    NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION B-BEAM INTERACTIONS WITH MATERIALS AND ATOMS, 2007, 263 (01): : 191 - 195
  • [42] A comparison of the dose-response behavior of AQUAJOINT®-based polymer gel and PAGAT gel dosimeters measured using Optical CT and MRI
    Takanashi, Takaoki
    Kawamura, Hiraku
    Fukasaku, Kazuaki
    Sahade, Daniel Antonio
    Hamada, Toshimasa
    9TH INTERNATIONAL CONFERENCE ON 3D RADIATION DOSIMETRY, 2017, 847
  • [43] Novel dual-wavelength optical-CT imaging method for gel dosimeter readout
    Du, Yi
    Wang, Xiangang
    Xiang, Xincheng
    Yves, De Deene
    10TH INTERNATIONAL CONFERENCE ON 3D RADIATION DOSIMETRY (IC3DDOSE), 2019, 1305
  • [44] The effect of x-ray CT dose on normoxic polymer gel dosimeters
    Baxter, P.
    Jirasek, A.
    Hilts, M.
    RADIOTHERAPY AND ONCOLOGY, 2007, 84 : S56 - S56
  • [45] Investigation of a 3D dosimetry system utilizing radio-chromic gel and telecentric optical-CT readout
    Hoopingarner, Scott
    Oldham, Mark
    Adamson, Justus
    Lakrad, Kawtar
    MEDICAL PHYSICS, 2023, 50 (09) : 5734 - 5744
  • [46] Optical-CT imaging of complex 3D dose distributions
    Oldham, M
    Kim, L
    Hugo, G
    Medical Imaging 2005: Physics of Medical Imaging, Pts 1 and 2, 2005, 5745 : 138 - 146
  • [47] Optical properties of polymerization N-(3-methoxypropyl) acrylamide polymer gel dosimeters for radiotherapy
    Rabaeh, Khalid
    Eyadeh, Molham
    PIGMENT & RESIN TECHNOLOGY, 2023, 52 (06) : 755 - 760
  • [48] Towards Optical CT scanning of radiochromic 3D dosimeters in mismatched refractive index solutions
    Dekker, K. H.
    Battista, J. J.
    Jordan, K. J.
    WORLD CONGRESS ON MEDICAL PHYSICS AND BIOMEDICAL ENGINEERING, 2015, VOLS 1 AND 2, 2015, 51 : 716 - 719
  • [49] Cosolvent-free polymer gel dosimeters with improved dose sensitivity and resolution for x-ray CT dose response
    Chain, J. N. M.
    Jirasek, A.
    Schreiner, L. J.
    McAuley, K. B.
    PHYSICS IN MEDICINE AND BIOLOGY, 2011, 56 (07): : 2091 - 2102
  • [50] A comprehensive evaluation of the PRESAGE/optical-CT 3D dosimetry system
    Sakhalkar, H. S.
    Adamovics, J.
    Ibbott, G.
    Oldham, M.
    MEDICAL PHYSICS, 2009, 36 (01) : 71 - 82