Sensitivity calibration procedures in optical-CT scanning of BANG®3 polymer gel dosimeters

被引:25
|
作者
Xu, Y. [1 ]
Wuu, Cheng-Shie [1 ]
Maryanski, Marek J. [1 ,2 ]
机构
[1] Columbia Univ, Dept Radiat Oncol, New York, NY 10032 USA
[2] MGS Res Inc, Madison, CT 06443 USA
基金
美国国家卫生研究院;
关键词
calibration; dosimeters; dosimetry; optical tomography; phantoms; polymer gels; RADIATION-DOSE DISTRIBUTIONS; COMPUTED-TOMOGRAPHY; VERIFICATION; THERAPY; MRI; RADIOSURGERY; RADIOTHERAPY; TEMPERATURE; PERFORMANCE; PHANTOM;
D O I
10.1118/1.3298017
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
The dose response of the BANG (R) 3 polymer gel dosimeter (MGS Research Inc., Madison, CT) was studied using the OCTOPUS (TM) laser CT scanner (MGS Research Inc., Madison, CT). Six 17 cm diameter and 12 cm high Barex cylinders, and 18 small glass vials were used to house the gel. The gel phantoms were irradiated with 6 and 10 MV photons, as well as 12 and 16 MeV electrons using a Varian Clinac 2100EX. Three calibration methods were used to obtain the dose response curves: (a) Optical density measurements on the 18 glass vials irradiated with graded doses from 0 to 4 Gy using 6 or 10 MV large field irradiations; (b) optical-CT scanning of Barex cylinders irradiated with graded doses (0.5, 1, 1.5, and 2 Gy) from four adjacent 4x4 cm(2) photon fields or 6x6 cm(2) electron fields; and (c) percent depth dose (PDD) comparison of optical-CT scans with ion chamber measurements for 6x6 cm(2), 12 and 16 MeV electron fields. The dose response of the BANG (R) 3 gel was found to be linear and energy independent within the uncertainties of the experimental methods (about 3%). The slopes of the linearly fitted dose response curves (dose sensitivities) from the four field irradiations (0.0752 +/- 3%, 0.0756 +/- 3%, 0.0767 +/- 3%, and 0.0759 +/- 3% cm(-1) Gy(-1)) and the PDD matching methods (0.0768 +/- 3% and 0.0761 +/- 3% cm(-1) Gy(-1)) agree within 2.2%, indicating a good reproducibility of the gel dose response within phantoms of the same geometry. The dose sensitivities from the glass vial approach are different from those of the cylindrical Barex phantoms by more than 30%, owing probably to the difference in temperature inside the two types of phantoms during gel formation and irradiation, and possible oxygen contamination of the glass vial walls. The dose response curve obtained from the PDD matching approach with 16 MeV electron field was used to calibrate the gel phantom irradiated with the 12 MeV, 6x6 cm(2) electron field. Three-dimensional dose distributions from the gel measurement and the Eclipse planning system (Varian Corporation, Palo Alto, CA) were compared and evaluated using 3% dose difference and 2 mm distance-to-agreement criteria.
引用
收藏
页码:861 / 868
页数:8
相关论文
共 50 条
  • [31] Optical Evaluation of normoxic PAGAT polymer gel dosimeters used to measure SWDP on diagnostic CT scanners
    Sarabipour, Sarvenaz
    Bosi, Stephen
    Hill, Brendan
    Baldock, Clive
    WORLD CONGRESS ON MEDICAL PHYSICS AND BIOMEDICAL ENGINEERING 2006, VOL 14, PTS 1-6, 2007, 14 : 1606 - 1608
  • [32] A New Optical-CT Apparatus for 3-D Radiotherapy Dosimetry: Is Free Space Scanning Feasible?
    Papadakis, Antonios E.
    Zacharakis, Giannis
    Maris, Thomas G.
    Ripoll, Jorge
    Damilakis, John
    IEEE TRANSACTIONS ON MEDICAL IMAGING, 2010, 29 (05) : 1204 - 1212
  • [33] Laser source compensator for optical CT imaging of gel dosimeters
    Dekker, Kurtis
    Battista, Jerry
    Jordan, Kevin
    MEDICAL PHYSICS, 2017, 44 (08) : 4383 - 4383
  • [34] A study of geometrical and reconstruction artifacts in 3D gel-dosimetry utilizing optical-CT
    Oldham, M
    Kim, L
    MEDICAL PHYSICS, 2003, 30 (06) : 1426 - 1426
  • [35] Light transport in optical-CT gel-dosimetry: Optical Monte-Carlo modeling
    Oldham, M
    MEDICAL PHYSICS, 2002, 29 (06) : 1370 - 1370
  • [36] Imaging dose: Calibration of polymer gel dosimeters for use in targeted radionuclide therapy
    Gear, Jonathan I.
    Flux, Glenn D.
    Charles-Edwards, Elizabeth
    Partridge, Mike
    Cook, Gary
    Ott, Robert J.
    2005 IEEE NUCLEAR SCIENCE SYMPOSIUM CONFERENCE RECORD, VOLS 1-5, 2005, : 2664 - 2668
  • [37] Preliminary investigation of the NMR, optical and x-ray CT dose-response of polymer gel dosimeters incorporating cosolvents to improve dose sensitivity
    Koeva, V. I.
    Olding, T.
    Jirasek, A.
    Schreiner, L. J.
    McAuley, K. B.
    PHYSICS IN MEDICINE AND BIOLOGY, 2009, 54 (09): : 2779 - 2790
  • [38] Characterization of a novel, fast, in-house, CCD-based optical-CT scanner for 3D dosimetry in radiochromic dosimeters
    Sakhalkar, H.
    Guo, P.
    Adamovics, J.
    Oldham, M.
    MEDICAL PHYSICS, 2007, 34 (06) : 2610 - 2610
  • [39] Implementation of 3D dose verification for IMRT using polymer gel dosimeters and laser-beam optical CT scanner
    Wuu, C
    Xu, Y
    Liang, R
    Witten, M
    Maryanski, M
    MEDICAL PHYSICS, 2002, 29 (06) : 1365 - 1366
  • [40] Radiation dose distributions in three dimensions from tomographic optical density scanning of polymer gels .2. Optical properties of the BANG polymer gel
    Maryanski, MJ
    Zastavker, YZ
    Gore, JC
    PHYSICS IN MEDICINE AND BIOLOGY, 1996, 41 (12): : 2705 - 2717