Sensitivity calibration procedures in optical-CT scanning of BANG®3 polymer gel dosimeters

被引:25
|
作者
Xu, Y. [1 ]
Wuu, Cheng-Shie [1 ]
Maryanski, Marek J. [1 ,2 ]
机构
[1] Columbia Univ, Dept Radiat Oncol, New York, NY 10032 USA
[2] MGS Res Inc, Madison, CT 06443 USA
基金
美国国家卫生研究院;
关键词
calibration; dosimeters; dosimetry; optical tomography; phantoms; polymer gels; RADIATION-DOSE DISTRIBUTIONS; COMPUTED-TOMOGRAPHY; VERIFICATION; THERAPY; MRI; RADIOSURGERY; RADIOTHERAPY; TEMPERATURE; PERFORMANCE; PHANTOM;
D O I
10.1118/1.3298017
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
The dose response of the BANG (R) 3 polymer gel dosimeter (MGS Research Inc., Madison, CT) was studied using the OCTOPUS (TM) laser CT scanner (MGS Research Inc., Madison, CT). Six 17 cm diameter and 12 cm high Barex cylinders, and 18 small glass vials were used to house the gel. The gel phantoms were irradiated with 6 and 10 MV photons, as well as 12 and 16 MeV electrons using a Varian Clinac 2100EX. Three calibration methods were used to obtain the dose response curves: (a) Optical density measurements on the 18 glass vials irradiated with graded doses from 0 to 4 Gy using 6 or 10 MV large field irradiations; (b) optical-CT scanning of Barex cylinders irradiated with graded doses (0.5, 1, 1.5, and 2 Gy) from four adjacent 4x4 cm(2) photon fields or 6x6 cm(2) electron fields; and (c) percent depth dose (PDD) comparison of optical-CT scans with ion chamber measurements for 6x6 cm(2), 12 and 16 MeV electron fields. The dose response of the BANG (R) 3 gel was found to be linear and energy independent within the uncertainties of the experimental methods (about 3%). The slopes of the linearly fitted dose response curves (dose sensitivities) from the four field irradiations (0.0752 +/- 3%, 0.0756 +/- 3%, 0.0767 +/- 3%, and 0.0759 +/- 3% cm(-1) Gy(-1)) and the PDD matching methods (0.0768 +/- 3% and 0.0761 +/- 3% cm(-1) Gy(-1)) agree within 2.2%, indicating a good reproducibility of the gel dose response within phantoms of the same geometry. The dose sensitivities from the glass vial approach are different from those of the cylindrical Barex phantoms by more than 30%, owing probably to the difference in temperature inside the two types of phantoms during gel formation and irradiation, and possible oxygen contamination of the glass vial walls. The dose response curve obtained from the PDD matching approach with 16 MeV electron field was used to calibrate the gel phantom irradiated with the 12 MeV, 6x6 cm(2) electron field. Three-dimensional dose distributions from the gel measurement and the Eclipse planning system (Varian Corporation, Palo Alto, CA) were compared and evaluated using 3% dose difference and 2 mm distance-to-agreement criteria.
引用
收藏
页码:861 / 868
页数:8
相关论文
共 50 条
  • [21] Fast laser scanning optical-CT apparatus for 3D radiation dosimetry
    Krstajic, Nikola
    Doran, Simon J.
    PHYSICS IN MEDICINE AND BIOLOGY, 2007, 52 (11): : N257 - N263
  • [22] Optical-CT gel-dosimetry II: Optical artifacts and geometrical distortion
    Oldham, M
    Kim, L
    MEDICAL PHYSICS, 2004, 31 (05) : 1093 - 1104
  • [23] An investigation comparing optical CT with MR scanning for polymer gel dosimetry
    Oldham, M
    Siewerdsen, JH
    Shetty, A
    Jaffray, DA
    PROCEEDINGS OF THE 22ND ANNUAL INTERNATIONAL CONFERENCE OF THE IEEE ENGINEERING IN MEDICINE AND BIOLOGY SOCIETY, VOLS 1-4, 2000, 22 : 1165 - 1171
  • [24] Study of the relative dose-response of BANG-3® polymer gel dosimeters in epithermal neutron irradiation
    Uusi-Simola, J
    Savolainen, S
    Kangasmäki, A
    Heikkinen, S
    Perkiö, J
    Ramadan, UA
    Seppälä, T
    Karila, J
    Serén, T
    Kotiluoto, P
    Sorvari, P
    Auterinen, I
    PHYSICS IN MEDICINE AND BIOLOGY, 2003, 48 (17): : 2895 - 2906
  • [25] NIPAM polymer gel dosimetry for IMRT four-field box irradiation using optical-CT scanner
    Yao, C. H.
    Hsu, W. T.
    Hsu, S. M.
    Ma, P. Y. L.
    Hsieh, B. T.
    Chang, Y. J.
    7TH INTERNATIONAL CONFERENCE ON 3D RADIATION DOSIMETRY (IC3DDOSE), 2013, 444
  • [26] A Characteristic Study on NIPAM Gel Dosimetry Using Optical-CT Scanner
    Yao, Chun-Hsu
    Hsu, Wang-Ting
    Lee, Jia-Jung
    Hsu, Shin-Ming
    Ma, Patrick Yuk-lun
    Hsieh, Bor-Tsung
    Chang, Yuan-Jen
    JOURNAL OF MEDICAL AND BIOLOGICAL ENGINEERING, 2014, 34 (04) : 327 - 332
  • [27] Water equivalence of NIPAM based polymer gel dosimeters with enhanced sensitivity for x-ray CT
    Gorjiara, Tina
    Hill, Robin
    Bosi, Stephen
    Kuncic, Zdenka
    Baldock, Clive
    RADIATION PHYSICS AND CHEMISTRY, 2013, 91 : 60 - 69
  • [28] Dual wavelength optical CT scanning of anthropomorphic shaped 3D dosimeters
    De Deene, Yves
    8TH INTERNATIONAL CONFERENCE ON 3D RADIATION DOSIMETRY (IC3DDOSE), 2015, 573
  • [29] An Investigation of a Low-Cost 'dry' Optical-CT Scanning System for 3D Dosimetry
    Bache, S.
    Malcolm, J.
    Adamovics, J.
    Oldham, M.
    MEDICAL PHYSICS, 2014, 41 (06) : 194 - 194
  • [30] Artifacts suppression in optical CT for gel dosimeters by iterative reconstruction
    Du Yi
    Wang Xiangang
    Xiang Xincheng
    8TH INTERNATIONAL CONFERENCE ON 3D RADIATION DOSIMETRY (IC3DDOSE), 2015, 573