Evaluation of dose calculation accuracy of treatment planning systems at hip prosthesis interfaces

被引:23
|
作者
Paulu, David [1 ]
Alaei, Parham [1 ]
机构
[1] Univ Minnesota, Dept Radiat Oncol, Minneapolis, MN 55455 USA
来源
关键词
algorithms; prosthesis; treatment planning; PELVIC IRRADIATION; CT; RADIOTHERAPY; BEAMS;
D O I
10.1002/acm2.12060
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
There are an increasing number of radiation therapy patients with hip prosthesis. The common method of minimizing treatment planning inaccuracies is to avoid radiation beams to transit through the prosthesis. However, the beams often exit through them, especially when the patient has a double-prosthesis. Modern treatment planning systems employ algorithms with improved dose calculation accuracies but even these algorithms may not predict the dose accurately at high atomic number interfaces. The current study evaluates the dose calculation accuracy of three common dose calculation algorithms employed in two commercial treatment planning systems. A hip prosthesis was molded inside a cylindrical phantom and the dose at several points within the phantom at the interface with prosthesis was measured using thermoluminescent dosimeters. The measured doses were then compared to the predicted ones by the planning systems. The results of the study indicate all three algorithms underestimate the dose at the prosthesis interface, albeit to varying degrees, and for both low-and high-energy x rays. The measured doses are higher than calculated ones by 5-22% for Pinnacle Collapsed Cone Convolution algorithm, 2-23% for Eclipse Acuros XB, and 6-25% for Eclipse Analytical Anisotropic Algorithm. There are generally better agreements for AXB algorithm and the worst results are for the AAA.
引用
收藏
页码:9 / 15
页数:7
相关论文
共 50 条
  • [1] Evaluation of the accuracy of various dose calculation algorithms of a commercial treatment planning system in the presence of hip prosthesis and comparison with Monte Carlo
    Mohammadi, Kheirollah
    Hassani, Mohsen
    Ghorbani, Mahdi
    Farhood, Bagher
    Knaup, Courtney
    [J]. JOURNAL OF CANCER RESEARCH AND THERAPEUTICS, 2017, 13 (03) : 501 - 509
  • [2] Evaluation of photon beam dose calculation accuracy of treatment planning systems using in vivo dosimetry
    Mohamed Ibrahim
    Wahib M.Attia
    Khaled M.Elshahat
    [J]. Oncology and Translational Medicine, 2016, 2 (05) : 234 - 238
  • [3] Evaluation of Treatment Planning Dose Calculation Accuracy at the Interface of Prosthetic Devices
    Paulu, D.
    Alaei, P.
    [J]. MEDICAL PHYSICS, 2012, 39 (06) : 3823 - 3823
  • [4] Accuracy assessment of absorbed dose rate calculation: evaluation of two treatment planning systems for molecular radiotherapy
    Negrin, J. Fragoso
    Gil, A. Vergara
    Santoro, L.
    Vauclin, S.
    Pouget, J.
    Kotzki, P.
    Deshayes, E.
    Bardies, M.
    [J]. EUROPEAN JOURNAL OF NUCLEAR MEDICINE AND MOLECULAR IMAGING, 2022, 49 (SUPPL 1) : S394 - S394
  • [5] Accuracy evaluation of dose calculation of ISOgray treatment planning system in wedged treatment fields
    Raghavi, S.
    Sadoughi, H. R.
    Ravari, M. E.
    Mansoury, M. A. Tajik
    Behmadi, M.
    [J]. INTERNATIONAL JOURNAL OF RADIATION RESEARCH, 2020, 22 (02): : 303 - 308
  • [6] Experimental evaluation of the accuracy of skin dose calculation for a commercial treatment planning system
    Court, Laurence E.
    Tishler, Roy B.
    Allen, Aaron M.
    Xiang, Hong
    Makrigiorgos, Mike
    Chin, Lee
    [J]. JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2008, 9 (01): : 29 - 35
  • [7] Accuracy of calculation of volumes for the evaluation of dose-volume histograms. Comparison of different treatment planning systems
    Fellner, C
    Sommer, B
    Siedhoff, N
    Potter, R
    [J]. STRAHLENTHERAPIE UND ONKOLOGIE, 1998, 174 (07) : 345 - 349
  • [8] The effect of dose calculation accuracy on inverse treatment planning
    Jeraj, R
    Keall, PJ
    Siebers, JV
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 2002, 47 (03): : 391 - 407
  • [9] Dose calculation accuracy of radiotherapy treatment planning systems in out-of-field regions
    Farhood, B.
    Ghorbani, M.
    [J]. Journal of Biomedical Physics and Engineering, 2019, 9 (02): : 133 - 136
  • [10] Dose calculation accuracy in proximity of a pacemaker: A multicenter study with threecommercial treatment planning systems
    Delana, Anna
    Barbareschi, Agnese
    Consorti, Rita
    Falco, Maria Daniela
    [J]. PHYSICA MEDICA-EUROPEAN JOURNAL OF MEDICAL PHYSICS, 2020, 80 : 201 - 208