Clinical implementation of varian enhanced dynamic wedges into the pinnacle treatment planning system: Monte Carlo dosimetry and patient-specific quality assurance techniques

被引:0
|
作者
Ahmad, M.
Deng, J.
Chen, Z.
Lund, M.
Kimmett, J.
Moran, M.
Nath, R.
机构
[1] Yale Univ, Sch Med, New Haven, CT USA
[2] Yale New Haven Med Ctr, New Haven, CT 06504 USA
[3] William W Backus Hosp, Norwich, NY USA
关键词
D O I
10.1118/1.2760762
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose: To present a commissioning study of enhanced dynamic wedges (EDWs) into the Pinnacle3 treatment planning system (TPS) and to evaluate various QA techniques for validating the treatment plans using EDWs. Method and Materials: Dose distribution modeling for EDW fields in the Pinnacle3 TPS is based on a combination of open‐field beam data and the Varian “Golden Segmented Treatment Table” (GSTT) unique to each photon beam. To validate the EDW models, dose profiles of 6− and 10−MV photon beams from a Clinac 2100C/D were measured in water for depths from near‐surface to 31.5 cm for a wide range of field sizes and wedge angles using a linear detector array of 25 energy‐compensated diodes (LDA‐25) to integrate the dose at multiple points in the EDW field for each exposure. The EDW output factors for square fields from 4 to 20 cm wide were measured in solid water using a cylindrical ionization chamber at a depth of 10 cm on central axis. The 6− and 10−MV photon beams emerging from the treatment head of the Clinac 2100C/D were fully simulated using the BEAMnrc and DOSXYZnrc was used to calculate the central‐axis percentage depth doses and dose profiles for the open and dynamically‐wedged fields in a water phantom. Film and MapCHECK data were compared with Pinnacle3 and MC predictions. Results: (1) Pinnacle collapsed‐cone convolution algorithm and MC simulations agreed with the measured EDW dose distributions to an accuracy of better than 3%. (2) Measured EDW output factors used for monitor‐unit calculation in Pinnacle3 TPS agreed with the MC predictions within 1–2%. (3) EDW‐based treatment plans were satisfactorily validated using MapCHECK and chamber measurements. Conclusions: We have successfully modeled EDWs into the Pinnacle3 TPS and demonstrated that MapCHECK and films dosimetry constitute valuable QA tools for validating planned dose distributions resulting from dynamically‐wedged beams. © 2007, American Association of Physicists in Medicine. All rights reserved.
引用
收藏
页码:2425 / 2425
页数:1
相关论文
共 50 条
  • [11] Quantitative Evaluation of Patient-Specific Quality Assurance Using Online Dosimetry System
    Jung, Jae-Yong
    Shin, Young-Ju
    Sohn, Seung-Chang
    Min, Jung-Whan
    Kim, Yon-Lae
    Kim, Dong-Su
    Choe, Bo-Young
    Suh, Tae-Suk
    JOURNAL OF THE KOREAN PHYSICAL SOCIETY, 2018, 72 (02) : 312 - 319
  • [12] Quantitative evaluation of patient-specific quality assurance using online dosimetry system
    Jae-Yong Jung
    Young-Ju Shin
    Seung-Chang Sohn
    Jung-Whan Min
    Yon-Lae Kim
    Dong-Su Kim
    Bo-Young Choe
    Tae-Suk Suh
    Journal of the Korean Physical Society, 2018, 72 : 312 - 319
  • [13] Treatment plan prescreening for patient-specific quality assurance measurements using independent Monte Carlo dose calculations
    Xu, Yuan
    Zhang, Ke
    Liu, Zhiqiang
    Liang, Bin
    Ma, Xiangyu
    Ren, Wenting
    Men, Kuo
    Dai, Jianrong
    FRONTIERS IN ONCOLOGY, 2022, 12
  • [14] Quality assurance of stereotactic treatment planning system: implementation of Monte Carlo technique on Leksell Gamma Knife
    Cheung, YC
    Yu, KN
    Yu, CP
    Ho, TK
    PATRAS MEDICAL PHYSICS 99, 1999, : 95 - 99
  • [15] MMCTP: a radiotherapy research environment for Monte Carlo and patient-specific treatment planning
    Alexander, A.
    DeBlois, F.
    Stroian, G.
    Al-Yahya, K.
    Heath, E.
    Seuntjens, J.
    PHYSICS IN MEDICINE AND BIOLOGY, 2007, 52 (13): : N297 - N308
  • [16] Commissioning Varian enhanced dynamic wedge in the PINNACLE treatment planning system using Gafchromic™ EBT film
    Fontanarosa, Davide
    Orlandini, Lucia Clara
    Andriani, Italo
    Bernardi, Luca
    MEDICAL PHYSICS, 2009, 36 (10) : 4504 - 4510
  • [17] MMCTP, a radiotherapy research environment for Monte Carlo and patient-specific treatment planning
    Alexander, A.
    DeBlois, F.
    Seuntjens, J.
    MEDICAL PHYSICS, 2006, 33 (06) : 2293 - 2293
  • [18] Verification of Varian enhanced dynamic wedge implementation in MasterPlan treatment planning system
    Pasquino, Massimo
    Borca, Valeria Casanova
    Tofani, Santi
    Ozzello, F.
    JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2009, 10 (02): : 11 - 20
  • [19] Analysis of Portal Dosimetry Patient-Specific Quality Assurance Results of 1203 Patients Treated with Varian Halcyon 1.0 Accelerator
    Wang, Z.
    Yang, B.
    Yu, L.
    Pang, T.
    Wang, B.
    Zhang, F.
    Qiu, J.
    MEDICAL PHYSICS, 2020, 47 (06) : E503 - E503
  • [20] Patient-Specific Quality Assurance for Monte Carlo-Calculated Lung SBRT On Cyberknife - Is It Necessary?
    Fabien, J.
    Zhang, Y.
    Brindle, J.
    Dobbins, D.
    Podder, T.
    Wessels, B.
    MEDICAL PHYSICS, 2013, 40 (06)