Clinical implementation of varian enhanced dynamic wedges into the pinnacle treatment planning system: Monte Carlo dosimetry and patient-specific quality assurance techniques

被引:0
|
作者
Ahmad, M.
Deng, J.
Chen, Z.
Lund, M.
Kimmett, J.
Moran, M.
Nath, R.
机构
[1] Yale Univ, Sch Med, New Haven, CT USA
[2] Yale New Haven Med Ctr, New Haven, CT 06504 USA
[3] William W Backus Hosp, Norwich, NY USA
关键词
D O I
10.1118/1.2760762
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose: To present a commissioning study of enhanced dynamic wedges (EDWs) into the Pinnacle3 treatment planning system (TPS) and to evaluate various QA techniques for validating the treatment plans using EDWs. Method and Materials: Dose distribution modeling for EDW fields in the Pinnacle3 TPS is based on a combination of open‐field beam data and the Varian “Golden Segmented Treatment Table” (GSTT) unique to each photon beam. To validate the EDW models, dose profiles of 6− and 10−MV photon beams from a Clinac 2100C/D were measured in water for depths from near‐surface to 31.5 cm for a wide range of field sizes and wedge angles using a linear detector array of 25 energy‐compensated diodes (LDA‐25) to integrate the dose at multiple points in the EDW field for each exposure. The EDW output factors for square fields from 4 to 20 cm wide were measured in solid water using a cylindrical ionization chamber at a depth of 10 cm on central axis. The 6− and 10−MV photon beams emerging from the treatment head of the Clinac 2100C/D were fully simulated using the BEAMnrc and DOSXYZnrc was used to calculate the central‐axis percentage depth doses and dose profiles for the open and dynamically‐wedged fields in a water phantom. Film and MapCHECK data were compared with Pinnacle3 and MC predictions. Results: (1) Pinnacle collapsed‐cone convolution algorithm and MC simulations agreed with the measured EDW dose distributions to an accuracy of better than 3%. (2) Measured EDW output factors used for monitor‐unit calculation in Pinnacle3 TPS agreed with the MC predictions within 1–2%. (3) EDW‐based treatment plans were satisfactorily validated using MapCHECK and chamber measurements. Conclusions: We have successfully modeled EDWs into the Pinnacle3 TPS and demonstrated that MapCHECK and films dosimetry constitute valuable QA tools for validating planned dose distributions resulting from dynamically‐wedged beams. © 2007, American Association of Physicists in Medicine. All rights reserved.
引用
收藏
页码:2425 / 2425
页数:1
相关论文
共 50 条
  • [21] The 'equivalent wedge' implementation of the Varian Enhanced Dynamic Wedge (EDW) into a treatment planning system
    Papatheodorou, S
    Zefkili, S
    Rosenwald, JC
    PHYSICS IN MEDICINE AND BIOLOGY, 1999, 44 (02): : 509 - 524
  • [22] Correction factors for commissioning and patient specific quality assurance of stereotactic fields in a Monte Carlo based treatment planning system TPS correction factors
    Goodall, Simon K.
    Rowshanfarzad, Pejman
    Ebert, Martin A.
    PHYSICAL AND ENGINEERING SCIENCES IN MEDICINE, 2023, 46 (02) : 735 - 745
  • [23] Clinical Implementation of the New Dosimetry System for TomoTherapy Patient Delivery Quality Assurance (DQA)
    Chung, E.
    Kwon, D.
    Park, T.
    Kang, H.
    Chung, Y.
    MEDICAL PHYSICS, 2017, 44 (06) : 2894 - 2894
  • [24] The development of patient-specific quality assurance (PSQA) for a proton wobbling nozzle using PTSim Monte Carlo code
    Oanh, Luu Dang Hoang
    Aso, Tsukasa
    Wang, Chun-Chieh
    Chao, Tsi-Chian
    Cho, I-Chun
    RADIATION PHYSICS AND CHEMISTRY, 2024, 223
  • [25] Clinical Implementation of a Commercial Monte Carlo Treatment Planning System for Electron Beams
    Vandervoort, E.
    Cygler, J.
    MEDICAL PHYSICS, 2011, 38 (06)
  • [26] PATIENT-SPECIFIC MONTE CARLO SIMULATION FOR THE DOSIMETRIC VALIDATION OF AN ADVANCED 1921R BRACHYTHERAPY TREATMENT PLANNING SYSTEM
    Zourari, K.
    Pantelis, E.
    Sakelliou, L.
    Georgiou, E.
    Karaiskos, P.
    Papagiannis, P.
    RADIOTHERAPY AND ONCOLOGY, 2010, 96 : S491 - S491
  • [27] Deep Hybrid Learning Prediction of Patient-Specific Quality Assurance in Radiotherapy: Implementation in Clinical Routine
    Moreau, Noemie
    Bonnor, Laurine
    Jaudet, Cyril
    Lechippey, Laetitia
    Falzone, Nadia
    Batalla, Alain
    Bertaut, Cindy
    Corroyer-Dulmont, Aurelien
    DIAGNOSTICS, 2023, 13 (05)
  • [28] Clinical Implementation of Monte Carlo Designed Patient-Specific Compensators for Co-60 TBI Treatments
    Serban, M.
    Seuntjens, J. P.
    Roussin, E.
    Wierzbicki, W.
    INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2011, 81 (02): : S897 - S898
  • [29] Clinical implementation and patient-specific quality assurance solutions for real-time target tracking and dynamic delivery in Radixact synchrony
    Trujillo-Bastidas, Christian D.
    Taylor, Michael J.
    Diaz-Londono, Gloria M.
    JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2025, 26 (01):
  • [30] Correction factors for commissioning and patient specific quality assurance of stereotactic fields in a Monte Carlo based treatment planning systemTPS correction factors
    Simon K. Goodall
    Pejman Rowshanfarzad
    Martin A. Ebert
    Physical and Engineering Sciences in Medicine, 2023, 46 : 735 - 745