Intrinsic and routine quality of serum total potassium measurement as investigated by split sample measurement with an ion chromatography candidate reference method

被引:0
|
作者
Thienpont, LM [1 ]
Van Nuwenborg, JE [1 ]
Stöckl, D [1 ]
机构
[1] State Univ Ghent, Fac Farmaceut Wetenschappen, B-9000 Ghent, Belgium
关键词
D O I
暂无
中图分类号
R446 [实验室诊断]; R-33 [实验医学、医学实验];
学科分类号
1001 ;
摘要
We evaluated the intrinsic quality of eight routine test systems for the measurement of serum total potassium (K+), as well as the routine quality of four of these systems, using a group of 60 single-donation serum samples that had been certified with an ion chromatography reference method. The intrinsic quality of the tests was evaluated by analysis of the sera in the manufacturers' application laboratories under strict internal quality control. The routine quality was evaluated by analysis of the same sera in five (per system) routine laboratories under daily working conditions. The results of the study were interpreted in light of the most stringent specifications derived from the biological variation of K+, which require limits of 6.3% for total error and 1.6% for systematic error. The study revealed that the intrinsic quality of all systems was excellent. None of the test systems yielded a substantial number of results outside the 6.3% total error limit, and only one test system exceeded the 1.6% systematic error limit. The majority of the routine laboratories reproduced the manufacturers' intrinsic quality. In particular, most laboratories satisfied the 6.3% total error limit. However, several laboratories exceeded the 1.6% systematic error limit. Generally, there was a considerable difference in quality between the participating laboratories. This showed that the major problems for serum K+ analysis (for samples with no unusual matrices and with concentrations within the reference interval) are at the routine laboratory.
引用
收藏
页码:849 / 857
页数:9
相关论文
共 50 条
  • [31] An improved reference measurement procedure for triglycerides and total glycerides in human serum by isotope dilution gas chromatography-mass spectrometry
    Chen, Yizhao
    Liu, Qinde
    Yong, Sharon
    Teo, Hui Ling
    Lee, Tong Kooi
    CLINICA CHIMICA ACTA, 2014, 428 : 20 - 25
  • [32] A candidate reference measurement procedure for quantification of glycocholic acid in human serum based on isotope dilution liquid chromatography-tandem mass spectrometry
    Zhang, Pingping
    Wang, Huimin
    Liang, Man
    Wang, Zhifang
    Liu, Chunlong
    Han, Yanlin
    ANALYTICAL AND BIOANALYTICAL CHEMISTRY, 2024, 416 (23) : 5121 - 5131
  • [33] Evaluation of a candidate reference measurement procedure for serum free testosterone based on ultrafiltration and isotope dilution-gas chromatography-mass spectrometry
    Van Uytfanghe, K
    Stöckl, D
    Kaufman, JM
    Fiers, T
    Ross, HA
    De Leenheer, AP
    Thienpont, LM
    CLINICAL CHEMISTRY, 2004, 50 (11) : 2101 - 2110
  • [34] Evaluation of a candidate reference measurement procedure for serum free testosterone based on ultrafiltration and isotope dilution-gas chromatography-mass spectrometry
    Thienpont, LM
    Van Uytfanghe, K
    Stöckl, D
    Kaufman, JM
    Fiers, T
    Ross, HA
    CLINICAL CHEMISTRY, 2004, 50 (06) : A93 - A93
  • [35] Development of a Candidate Reference Measurement Procedure for the Analysis of Cortisol in Human Serum Samples by Isotope Dilution-Gas Chromatography-Mass Spectrometry
    Kawaguchi, Migaku
    Takatsu, Akiko
    ANALYTICAL SCIENCES, 2009, 25 (08) : 989 - 992
  • [36] Development and evaluation of a candidate reference measurement procedure for the determination of testosterone in human serum using isotope dilution liquid chromatography/tandem mass spectrometry
    Susan S.-C. Tai
    Bei Xu
    Michael J. Welch
    Karen W. Phinney
    Analytical and Bioanalytical Chemistry, 2007, 388 : 1087 - 1094
  • [37] Development and evaluation of a candidate reference measurement procedure for the determination of testosterone in human serum using isotope dilution liquid chromatography/tandem mass spectrometry
    Tai, Susan S. -C.
    Xu, Bei
    Welch, Michael J.
    Phinney, Karen W.
    ANALYTICAL AND BIOANALYTICAL CHEMISTRY, 2007, 388 (5-6) : 1087 - 1094
  • [38] Development of a Candidate Reference Measurement Procedure for the Analysis of Cortisol in Human Serum Samples by Isotope Dilution-Gas Chromatography-Mass Spectrometry
    Migaku Kawaguchi
    Akiko Takatsu
    Analytical Sciences, 2009, 25 : 989 - 992
  • [39] Liquid chromatography-isotope dilution tandem mass spectrometry method for the measurement of urea in human serum and assignment of reference values to external quality assessment samples
    Yong, Sharon
    Liu, Hong
    Wong, Lingkai
    Teo, Hui Ling
    Chen, Yizhao
    Liu, Qinde
    Teo, Tang Lin
    Lee, Tong Kooi
    INTERNATIONAL JOURNAL OF MASS SPECTROMETRY, 2017, 414 : 87 - 93
  • [40] Candidate reference method for total thyroxine in human serum: Use of isotope-dilution liquid chromatography-mass spectrometry with electrospray ionization
    Tai, SSC
    Sniegoski, LT
    Welch, MJ
    CLINICAL CHEMISTRY, 2002, 48 (04) : 637 - 642