Intrinsic and routine quality of serum total potassium measurement as investigated by split sample measurement with an ion chromatography candidate reference method

被引:0
|
作者
Thienpont, LM [1 ]
Van Nuwenborg, JE [1 ]
Stöckl, D [1 ]
机构
[1] State Univ Ghent, Fac Farmaceut Wetenschappen, B-9000 Ghent, Belgium
关键词
D O I
暂无
中图分类号
R446 [实验室诊断]; R-33 [实验医学、医学实验];
学科分类号
1001 ;
摘要
We evaluated the intrinsic quality of eight routine test systems for the measurement of serum total potassium (K+), as well as the routine quality of four of these systems, using a group of 60 single-donation serum samples that had been certified with an ion chromatography reference method. The intrinsic quality of the tests was evaluated by analysis of the sera in the manufacturers' application laboratories under strict internal quality control. The routine quality was evaluated by analysis of the same sera in five (per system) routine laboratories under daily working conditions. The results of the study were interpreted in light of the most stringent specifications derived from the biological variation of K+, which require limits of 6.3% for total error and 1.6% for systematic error. The study revealed that the intrinsic quality of all systems was excellent. None of the test systems yielded a substantial number of results outside the 6.3% total error limit, and only one test system exceeded the 1.6% systematic error limit. The majority of the routine laboratories reproduced the manufacturers' intrinsic quality. In particular, most laboratories satisfied the 6.3% total error limit. However, several laboratories exceeded the 1.6% systematic error limit. Generally, there was a considerable difference in quality between the participating laboratories. This showed that the major problems for serum K+ analysis (for samples with no unusual matrices and with concentrations within the reference interval) are at the routine laboratory.
引用
收藏
页码:849 / 857
页数:9
相关论文
共 50 条
  • [41] Development and evaluation of a candidate reference measurement procedure for the determination of progesterone in human serum using isotope-dilution liquid chromatography/tandem mass spectrometry
    Tai, Susan S. -C.
    Xu, Bei
    Welch, Michael J.
    ANALYTICAL CHEMISTRY, 2006, 78 (18) : 6628 - 6633
  • [42] Reference method for serum total cholesterol measurement: Does substituting enzymatic step for Liebermann-Burchard reaction improve specificity?
    Modenese, A
    Carobene, A
    Ferrero, C
    Ceriotti, F
    Franzini, C
    DeGiorgi, E
    Franzin, M
    Giorgio, A
    Baldo, L
    CLINICAL CHEMISTRY, 1996, 42 (03) : 475 - 477
  • [43] Development and evaluation of a candidate reference method for the determination of total cortisol in human serum using isotope dilution liquid chromatography/mass spectrometry and liquid chromatography/tandem mass spectrometry
    Tai, SSC
    Welch, MJ
    ANALYTICAL CHEMISTRY, 2004, 76 (04) : 1008 - 1014
  • [44] Comparison of Abbott IMx total homocysteine assay with a high pressure liquid chromatography method for the measurement of total homocysteine in plasma and serum from a Norwegian population
    Mansoor, MA
    SCANDINAVIAN JOURNAL OF CLINICAL & LABORATORY INVESTIGATION, 1999, 59 (05): : 369 - 374
  • [45] Inductively Coupled Plasma Mass Spectrometry as a Reference Method to Evaluate Serum Calcium Measurement Bias and the Commutability of Processed Materials during Routine Measurements
    Zhao HaiJian
    Ge MengLei
    Yan Yin
    Zhang TianJiao
    Zeng Jie
    Zhou WeiYan
    Wang YuFei
    Meng QingHui
    Zhang ChuanBao
    National Center for Clinical Laboratories Beijing Hospital National Center of Gerontology Beijing Engineering Research Center of Laboratory Medicine Beijing China
    Department of Clinical Laboratory Childrens Hospital Capital Institute of Pediatrics Beijing China
    National Center for Clinical Laboratories Beijing Hospital National Center of Gerontology Beijing Engineering Research Center of Laboratory Medicine Beijing China
    Graduate School Peking Union Medical College Chinese Academy of Medical Sciences Beijing China
    中华医学杂志英文版, 2018, 131 (13) : 1584 - 1590
  • [46] Inductively Coupled Plasma Mass Spectrometry as a Reference Method to Evaluate Serum Calcium Measurement Bias and the Commutability of Processed Materials during Routine Measurements
    Zhao, Hai-Jian
    Ge, Meng-Lei
    Yu, Yin
    Zhang, Tian-Jiao
    Zang, Jie
    Zhou, Wei-Yan
    Wang, Yu-Fei
    Meng, Qing-Hui
    Zhang, Chuan-Bao
    CHINESE MEDICAL JOURNAL, 2018, 131 (13) : 1584 - 1590
  • [47] An isotope dilution-liquid chromatography-tandem mass spectrometry-based candidate reference measurement procedure for the quantification of cortisol in human serum and plasma
    Ott, Myriam
    Singh, Neeraj
    Bauland, Friederike
    Koeppl, Daniel
    Kandler, Kerstin
    Gaudl, Alexander
    Geistanger, Andrea
    Ceglarek, Uta
    Rauh, Manfred
    Geletneky, Christian
    Taibon, Judith
    CLINICAL CHEMISTRY AND LABORATORY MEDICINE, 2025,
  • [48] Development and validation of an isotope dilution-liquid chromatography (ID-LC-MS/MS)-based candidate reference measurement procedure for total 3,3′,5-triiodothyronine in human serum
    Li, Wenya
    Gao, Bingling
    Yang, Tao
    Tang, Liping
    Song, Dan
    Li, Hongmei
    Sun, Keqi
    Xiao, Peng
    ANALYTICAL AND BIOANALYTICAL CHEMISTRY, 2025,
  • [49] Evaluation and clinical application of a bracketing calibration-based isotope dilution liquid chromatography-tandem mass spectrometry candidate reference measurement procedure for serum theophylline
    Du, Yuanyuan
    Yan, Qiaofang
    Zhan, Min
    Zhang, Qiaoxuan
    Huang, Di
    Zhang, Pengwei
    Yan, Jun
    Wang, Jianbing
    Huang, Xianzhang
    Han, Liqiao
    ANALYTICAL AND BIOANALYTICAL CHEMISTRY, 2024, 416 (22) : 4897 - 4906
  • [50] EXTERNAL QUALITY ASSESSMENT OF URINARY-FREE CORTISOL MEASUREMENT IN THE UK AGAINST A GAS-CHROMATOGRAPHY MASS-SPECTROSCOPY REFERENCE METHOD
    HOLDER, G
    ANNALS OF CLINICAL BIOCHEMISTRY, 1995, 32 : 84 - 90