Comparison of four different immunoassays and a rapid isotope-dilution liquid chromatography-tandem mass spectrometry assay for serum folate

被引:2
|
作者
Jin, Lizi [1 ,2 ]
Lu, Youli [3 ,4 ,5 ]
Yi, Xilian [1 ,2 ]
Zhang, Meiwei [3 ,4 ,5 ]
Zhang, Jiangtao [1 ,2 ]
Zhou, Weiyan [1 ,2 ]
Zeng, Jie [1 ,2 ]
Zhang, Tianjiao [1 ,2 ]
Zhang, Chuanbao [1 ,2 ]
机构
[1] Chinese Acad Med Sci, Beijing Hosp, Natl Ctr Clin Labs, Inst Geriatr Med,Natl Ctr Gerontol, Beijing, Peoples R China
[2] Chinese Acad Med Sci & Peking Union Med Coll, 1 Dahua Rd, Beijing 100730, Peoples R China
[3] Fudan Univ, Shanghai Xuhui Cent Hosp, Cent Lab, Zhongshan Xuhui Hosp, Shanghai, Peoples R China
[4] Shanghai Engn Res Ctr Phase & Clin Res & Qual Con, Shanghai, Peoples R China
[5] Shanghai Inst Clin Mass Spectrometry, Shanghai, Peoples R China
关键词
immunoassays; isotope-dilution liquid chromatography-tandem mass spectrometry (ID-LC-MS; MS); method comparison; serum folate;
D O I
10.1515/cclm-2021-1283
中图分类号
R446 [实验室诊断]; R-33 [实验医学、医学实验];
学科分类号
1001 ;
摘要
Objectives Accurate measurement of serum folate is essential for the diagnosis and management of various disorders. This study aims to investigate the between-method differences of four immunoassays and a rapid isotope-dilution liquid chromatography-tandem mass spectrometry (ID-LC-MS/MS) method. Methods Roche Cobas (USA), Abbott Alinity i2000 (USA), Beckman Coulter Access (USA), Mindray CL-6000i (China), and the ID-LC-MS/MS method were compared using 46 human serum samples. The results were analysed by Passing-Bablok regressions and Bland-Altman plots. A bias of 13.31% based on biological variation was used as the bias criterion. Results All the within-run and total coefficients of variation (CVs) met the specification. The folate concentrations determined by all the assays were significantly different (p=0.0028). All assays had correlation coefficients over 0.97 with each other. The 95% confidence intervals (CIs) for the slope seldom contained 1 and few 95% CIs for the intercept contained 0 in the regression equations. Compared to ID-LC-MS/MS, the biases of all assays ranged from -20.91 to 13.56 nmol/L, and the mean relative biases ranged from -9.85 to 40.33%. The predicted mean relative biases at the medical decision levels rarely met the criterion. Conclusions Assays for serum folate had good correlations with each other but lacked good agreement. The accuracy and consistency of assays for serum folate should be measured and assessed routinely. Standardization work to improve the accuracy of serum folate assays, such as the extension of traceability to reference methods or materials, calibration standardization efforts, and assay-adjusted cut-offs should be promoted.
引用
收藏
页码:1393 / 1402
页数:10
相关论文
共 50 条
  • [31] Stable isotope-dilution liquid chromatography/tandem mass spectrometry method for determination of thyroxine in saliva
    Higashi, Tatsuya
    Ichikawa, Takuya
    Shimizu, Chikara
    Nagai, So
    Inagaki, Shinsuke
    Min, Jun Zhe
    Chiba, Hitoshi
    Ikegawa, Shigeo
    Toyo'oka, Toshimasa
    JOURNAL OF CHROMATOGRAPHY B-ANALYTICAL TECHNOLOGIES IN THE BIOMEDICAL AND LIFE SCIENCES, 2011, 879 (13-14): : 1013 - 1017
  • [32] Determination of Perfluorochemicals in Human Milk Using Isotope-dilution Liquid Chromatography Tandem Mass Spectrometry
    Wang, Yu-Chen
    Lin, Angela Yu-Chen
    Wang, Shu-Li
    Ding, Wang-Hsien
    JOURNAL OF THE CHINESE CHEMICAL SOCIETY, 2012, 59 (04) : 544 - 549
  • [33] Determination of Pentachlorophenol in Milk Powder by Isotope Dilution Liquid Chromatography-Tandem Mass Spectrometry
    Qi Y.
    Li J.
    Li C.
    Xu Q.
    Shipin Kexue/Food Science, 2023, 44 (16): : 324 - 330
  • [34] Quantification of estrogens in infant formulas by isotope dilution liquid chromatography-tandem mass spectrometry
    Liu, Shuan
    Li, Xiaomin
    Zhao, Yingchen
    Yin, Xiong
    Zhang, Qinghe
    Li, Xiuqin
    Tang, Guangshi
    Li, Hongmei
    ANALYTICAL METHODS, 2018, 10 (32) : 3968 - 3975
  • [35] DETERMINATION OF REDUCED AND OXIDIZED GLUTATHIONE BY ISOTOPE DILUTION LIQUID CHROMATOGRAPHY-TANDEM MASS SPECTROMETRY
    Yu, W. T.
    Kao, C. Y.
    Shiesh, S. C.
    CLINICAL CHEMISTRY AND LABORATORY MEDICINE, 2011, 49 : S783 - S783
  • [36] Analysis of acrylamide in coffee and cocoa by isotope dilution liquid chromatography-tandem mass spectrometry
    Aguas, Patricia C.
    Fitzhenry, Matthew J.
    Giannikopoulos, Georgina
    Varelis, Peter
    ANALYTICAL AND BIOANALYTICAL CHEMISTRY, 2006, 385 (08) : 1526 - 1531
  • [37] Rapid quantitation of digoxin in human plasma and urine using isotope dilution liquid chromatography-tandem mass spectrometry
    Salvador, A
    Sagan, C
    Denouel, J
    JOURNAL OF LIQUID CHROMATOGRAPHY & RELATED TECHNOLOGIES, 2006, 29 (13) : 1917 - 1932
  • [38] RAPID-DETERMINATION OF METHANDROSTENOLONE IN EQUINE URINE BY ISOTOPE-DILUTION LIQUID-CHROMATOGRAPHY TANDEM MASS-SPECTROMETRY
    EDLUND, PO
    BOWERS, L
    HENION, J
    COVEY, TR
    JOURNAL OF CHROMATOGRAPHY-BIOMEDICAL APPLICATIONS, 1989, 497 : 49 - 57
  • [39] Quantification of Polybrominated and Polychlorinated Biphenyls in Human Matrices by Isotope-Dilution Gas Chromatography-Tandem Mass Spectrometry
    Marder, M. Elizabeth
    Panuwet, Parinya
    Hunter, Ronald E.
    Ryan, P. Barry
    Marcus, Michele
    Barr, Dana Boyd
    JOURNAL OF ANALYTICAL TOXICOLOGY, 2016, 40 (07) : 511 - 518
  • [40] State-of-the-art of serum testosterone measurement by isotope dilution-liquid chromatography-tandem mass spectrometry
    Thienpont, Linda M.
    Van Uytfanghe, Katleen
    Blincko, Stuart
    Ramsay, Carol S.
    Xie, Hui
    Doss, Robert C.
    Keevil, Brian G.
    Owen, Laura J.
    Rockwood, Alan L.
    Kushnir, Mark M.
    Chun, Kelly Y.
    Chandler, Donald W.
    Field, Helen P.
    Sluss, Patrick M.
    CLINICAL CHEMISTRY, 2008, 54 (08) : 1290 - 1297