Comparative performance analysis for abdominal phantom ROI detectability according to CT reconstruction algorithm: ADMIRE

被引:2
|
作者
Shin, Jun-Bong [1 ,2 ]
Yoon, Do-Kun [1 ,2 ]
Pak, Seongyong [3 ]
Kwon, Yang-Ho [3 ]
Suh, Tae Suk [1 ,2 ]
机构
[1] Catholic Univ Korea, Dept Biomed Engn, Coll Med, Seoul, South Korea
[2] Catholic Univ Korea, Coll Med, Res Inst Biomed Engn, Seoul, South Korea
[3] Siemens Healthineers Ltd, Seoul, South Korea
来源
关键词
abdomen phantom; ADMIRE; FBP; MDCT; SAFIRE; MODELED ITERATIVE RECONSTRUCTION; LOW-CONTRAST DETECTABILITY; RADIATION-DOSE REDUCTION; COMPUTED-TOMOGRAPHY;
D O I
10.1002/acm2.12765
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose We compared and analyzed the detectability performance pertaining to an abdominal phantom including a region of interest (ROI) according to a computed tomography (CT) reconstruction algorithm. Methods Three types of reconstruction algorithms (FBP, SAFIRE, and ADMIRE) were used to evaluate the detectability performance using the abdominal phantom (phantom size: 25 x 18 x 28 cm(3)). The vendor default settings for routine multi-detector computed tomography abdominal scans were used. As the quantitative evaluation method, the contrast-to-noise ratio (CNR), difference in coefficient of variation (COV) with the normalization based on the FBP data, and the noise power spectrum (NPS) were measured. Results The characteristic of the ADMIRE-3 reconstructed image was higher than those of the FBP and SAFIRE-3 reconstructed images. The CNR values of the SAFIRE and ADMIRE images were much higher than the corresponding values of the FBP images. The difference in COV values for the ADMIRE images was similar to 1.2 times lower than the corresponding values of the SAFIRE images. Conclusion The comparative analysis of the abdominal phantom low-contrast resolution differences for each CT exposure parameters showed that ADMIRE demonstrated better results than SAFIRE and FBP in terms of contrast, CNR, COV difference, and 1D NPS. This indicates that ADMIRE can provide a clearer observation even with the same number of contrast objects as compared to SAFIRE and FBP owing to its better contrast resolution in the central part of the contrast hole at low kV.
引用
收藏
页码:136 / 143
页数:8
相关论文
共 50 条
  • [1] CT Reconstruction Algorithm and Low Contrast Detectability of Phantom Study: A Systematic Review and Meta-Analysis
    Yusof, Nur Aimi Adibah
    Karim, Muhammad Khalis Abdul
    Asikin, Nursyazalina Mohd
    Paiman, Suriati
    Kechik, Mohd Mustafa Awang
    Rahman, Mohd Amiruddin Abdul
    Noor, Noramaliza Mohd
    CURRENT MEDICAL IMAGING, 2023, 19 (10) : 1105 - 1113
  • [2] Advanced Modeled Iterative Reconstruction (ADMIRE) Facilitates Radiation Dose Reduction in Abdominal CT
    Ellmann, Stephan
    Kammerer, Ferdinand
    Allmendinger, Thomas
    Hammon, Matthias
    Janka, Rolf
    Lell, Michael
    Uder, Michael
    Kramer, Manuel
    ACADEMIC RADIOLOGY, 2018, 25 (10) : 1277 - 1284
  • [3] Detectability of Hypoattenuating Liver Lesions with Deep Learning CT Reconstruction: A Phantom and Patient Study
    Cao, Jinjin
    Mroueh, Nayla
    Mercaldo, Nathaniel
    Lennartz, Simon
    Kongboonvijit, Sasiprang
    Rao, Shravya Srinivas
    Pisuchpen, Nisanard
    Baliyan, Vinit
    Pierce, Theodore T.
    Anderson, Mark A.
    Sertic, Madeleine
    Shenoy-Bhangle, Anuradha S.
    Kambadakone, Avinash R.
    RADIOLOGY, 2024, 313 (01)
  • [4] Improved image quality and dose reduction in abdominal CT with deep-learning reconstruction algorithm: a phantom study
    Greffier, Joel
    Durand, Quentin
    Frandon, Julien
    Si-Mohamed, Salim
    Loisy, Maeliss
    de Oliveira, Fabien
    Beregi, Jean-Paul
    Dabli, Djamel
    EUROPEAN RADIOLOGY, 2023, 33 (01) : 699 - 710
  • [5] Improved image quality and dose reduction in abdominal CT with deep-learning reconstruction algorithm: a phantom study
    Joël Greffier
    Quentin Durand
    Julien Frandon
    Salim Si-Mohamed
    Maeliss Loisy
    Fabien de Oliveira
    Jean-Paul Beregi
    Djamel Dabli
    European Radiology, 2023, 33 : 699 - 710
  • [6] Evaluation of low-contrast detectability for iterative reconstruction in pediatric abdominal computed tomography: a phantom study
    Nicholas Rubert
    Richard Southard
    Susan M. Hamman
    Ryan Robison
    Pediatric Radiology, 2020, 50 : 345 - 356
  • [7] Evaluation of low-contrast detectability for iterative reconstruction in pediatric abdominal computed tomography: a phantom study
    Rubert, Nicholas
    Southard, Richard
    Hamman, Susan M.
    Robison, Ryan
    PEDIATRIC RADIOLOGY, 2020, 50 (03) : 345 - 356
  • [8] ROC analysis of PET imaging performance of two PET/CT scanners based on lesion detectability in a torso phantom
    Bernstein, Kenneth R.
    Matthews, Kenneth L.
    Morrow, Andrew N.
    Smith, Blair M.
    Bujenovic, L. Steven
    2006 IEEE NUCLEAR SCIENCE SYMPOSIUM CONFERENCE RECORD, VOL 1-6, 2006, : 2636 - 2643
  • [9] Hybrid Iterative Reconstruction Technique for Abdominal CT Protocols in Obese Patients: Assessment of Image Quality, Radiation Dose, and Low-Contrast Detectability in a Phantom
    Schindera, Sebastian T.
    Odedra, Devang
    Mercer, Diego
    Thipphavong, Seng
    Chou, Paul
    Szucs-Farkas, Zsolt
    Rogalla, Patrik
    AMERICAN JOURNAL OF ROENTGENOLOGY, 2014, 202 (02) : W146 - W152
  • [10] IMAGE QUALITY AND POTENTIAL DOSE REDUCTION USING ADVANCED MODELED ITERATIVE RECONSTRUCTION (ADMIRE) IN ABDOMINAL CT - A REVIEW
    Kataria, B.
    Althen, J. Nilsson
    Smedby, O.
    Persson, A.
    Sokjer, H.
    Sandborg, M.
    RADIATION PROTECTION DOSIMETRY, 2021, 195 (3-4) : 177 - 187