Evaluation of low-contrast detectability for iterative reconstruction in pediatric abdominal computed tomography: a phantom study

被引:3
|
作者
Rubert, Nicholas [1 ]
Southard, Richard [1 ,2 ]
Hamman, Susan M. [3 ]
Robison, Ryan [1 ,4 ]
机构
[1] Phoenix Childrens Hosp, Dept Radiol, 1919 E Thomas Rd, Phoenix, AZ 85016 USA
[2] Univ Arizona, Coll Med, Phoenix, AZ USA
[3] Univ Michigan Hlth Syst, Dept Radiol, CS Mott Childrens Hosp, Sect Pediat Radiol, Ann Arbor, MI USA
[4] Arizona State Univ, Coll Nursing & Hlth Innovat, Phoenix, AZ USA
关键词
Children; Computed tomography; Dose reduction; Iterative reconstruction; Low-contrast detectability; Phantom; TASK-BASED PERFORMANCE; DOSE REDUCTION; IMAGE QUALITY; CT; ALGORITHM; CAMPAIGN;
D O I
10.1007/s00247-019-04561-8
中图分类号
R72 [儿科学];
学科分类号
100202 ;
摘要
Background Iterative reconstruction is offered by all vendors to achieve similar or better CT image quality at lower doses than images reconstructed with filtered back-projection. Objective The purpose of this study was to investigate the dose-reduction potential for pediatric abdominal CT imaging when using either a commercially available hybrid or a commercially available model-based iterative reconstruction algorithm from a single manufacturer. Materials and methods A phantom containing four low-contrast inserts and a uniform background with total attenuation equivalent to the abdomen of an average 8-year-old child was imaged on a CT scanner (IQon; Philips Healthcare, Cleveland, OH). We reconstructed images using both hybrid iterative reconstruction (iDose(4)) and model-based iterative reconstruction (Iterative Model Reconstruction). The four low-contrast inserts had circular cross-section with diameters of 3 mm, 5 mm, 7 mm and 10 mm and contrasts of 14 Hounsfield units (HU), 7 HU, 5 HU and 3 HU, respectively. Helical scans with identical kilovoltage (kV), pitch, rotation time, and collimation were repeated on the phantom at volume CT dose index (CTDIvol) of 2.0 milligrays (mGy), 3.0 mGy, 4.5 mGy and 6.0 mGy. We measured the contrast-to-noise ratio (CNR) in each rod across scans. Additionally, we collected sub-images containing each rod and sub-images containing the background and used them in two-alternative forced choice observer experiments with four observers (two radiologists and two physicists). We calculated the dose-reduction potential of both iterative reconstruction algorithms relative to a scan performed at 6 mGy and reconstructed with filtered back-projection. Results We calculated dose-reduction potential by either matching average equal observer performance in the two-alternative forced choice experiments or matching CNR. When matching CNR, the dose-reduction potential was 34% to 45% for hybrid iterative reconstruction and 89% to 95% for model-based iterative reconstruction. When matching average observer performance, the dose-reduction potential was 9% to 30% for hybrid iterative reconstruction and 57% to 74% for model-based iterative reconstruction. The range in dose-reduction potential depended on the rod size and contrast level. Conclusion Observer performance in this phantom study indicates that the dose-reduction potential indicated by an observer study is less than that of CNR; extrapolating the results to clinical studies suggests that the dose-reduction potential would also be less.
引用
收藏
页码:345 / 356
页数:12
相关论文
共 50 条
  • [1] Evaluation of low-contrast detectability for iterative reconstruction in pediatric abdominal computed tomography: a phantom study
    Nicholas Rubert
    Richard Southard
    Susan M. Hamman
    Ryan Robison
    [J]. Pediatric Radiology, 2020, 50 : 345 - 356
  • [2] Iterative reconstruction with multifrequency signal recognition technology to improve low-contrast detectability: A phantom study
    Funama, Yoshinori
    Shirasaka, Takashi
    Goto, Taiga
    Aoki, Yuko
    Tanaka, Kana
    Yoshida, Ryo
    [J]. ACTA RADIOLOGICA OPEN, 2022, 11 (06)
  • [3] Objective assessment of low-contrast computed tomography images with iterative reconstruction
    Urikura, Atsushi
    Hara, Takanori
    Ichikawa, Katsuhiro
    Nishimaru, Eiji
    Hoshino, Takashi
    Yoshida, Tsukasa
    Nakaya, Yoshihiro
    [J]. PHYSICA MEDICA-EUROPEAN JOURNAL OF MEDICAL PHYSICS, 2016, 32 (08): : 992 - 998
  • [4] DETECTABILITY OF LOW-CONTRAST FEATURES IN COMPUTED-TOMOGRAPHY
    WARREN, RC
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 1984, 29 (01): : 1 - 13
  • [5] Influence of the adaptive iterative dose reduction 3D algorithm on the detectability of low-contrast lesions and radiation dose repeatability in abdominal computed tomography: a phantom study
    Jeong Hee Yoon
    Jeong Min Lee
    Bo Yun Hur
    Jeehyun Baek
    Hackjoon Shim
    Joon Koo Han
    Byung Ihn Choi
    [J]. Abdominal Imaging, 2015, 40 : 1843 - 1852
  • [6] Influence of the adaptive iterative dose reduction 3D algorithm on the detectability of low-contrast lesions and radiation dose repeatability in abdominal computed tomography: a phantom study
    Yoon, Jeong Hee
    Lee, Jeong Min
    Hur, Bo Yun
    Baek, Jeehyun
    Shim, Hackjoon
    Han, Joon Koo
    Choi, Byung Ihn
    [J]. ABDOMINAL IMAGING, 2015, 40 (06): : 1843 - 1852
  • [7] NOISE, LOW-CONTRAST DETECTABILITY, AND PATIENT DOSAGE IN COMPUTED TOMOGRAPHY
    PAYNE, JT
    MCCULLOUGH, EC
    [J]. JOURNAL OF COMPUTER ASSISTED TOMOGRAPHY, 1977, 1 (02) : 252 - 252
  • [8] Low-Contrast Detectability for X-Ray Computed Tomography
    Hsieh, J.
    Toth, T.
    [J]. MEDICAL PHYSICS, 2008, 35 (06)
  • [9] Improving Low -contrast Detectability and Noise Texture Pattern for Computed Tomography Using Iterative Reconstruction Accelerated with Machine Learning Method: A Phantom Study
    Funama, Yoshinori
    Takahashi, Hisashi
    Goto, Taiga
    Aoki, Yuko
    Yoshida, Ryo
    Kumagai, Yukio
    Awai, Kazuo
    [J]. ACADEMIC RADIOLOGY, 2020, 27 (07) : 929 - 936
  • [10] Hybrid Iterative Reconstruction Technique for Abdominal CT Protocols in Obese Patients: Assessment of Image Quality, Radiation Dose, and Low-Contrast Detectability in a Phantom
    Schindera, Sebastian T.
    Odedra, Devang
    Mercer, Diego
    Thipphavong, Seng
    Chou, Paul
    Szucs-Farkas, Zsolt
    Rogalla, Patrik
    [J]. AMERICAN JOURNAL OF ROENTGENOLOGY, 2014, 202 (02) : W146 - W152