Improved image quality and dose reduction in abdominal CT with deep-learning reconstruction algorithm: a phantom study

被引:18
|
作者
Greffier, Joel [1 ]
Durand, Quentin [1 ]
Frandon, Julien [1 ]
Si-Mohamed, Salim [2 ,3 ]
Loisy, Maeliss [1 ]
de Oliveira, Fabien [1 ]
Beregi, Jean-Paul [1 ]
Dabli, Djamel [1 ]
机构
[1] Univ Montpellier, Dept Med Imaging, Nimes Med Imaging Grp, CHU Nimes,EA 2992, Bd Prof Robert Debre, F-30029 Montpellier 9, France
[2] Univ Lyon, Univ Claude Bernard Lyon 1, UJM St Etienne, INSA Lyon,CNRS,Inserm,CREATIS UMR 5220,U1206, 7 Ave Jean Capelle O, F-69621 Villeurbanne, France
[3] Hosp Civils Lyon, Louis Pradel Hosp, Dept Radiol, 59 Blvd Pinel, F-69500 Bron, France
关键词
Artificial intelligence; Deep learning; Multidetector computed tomography; Image enhancement; Image processing; computer-assisted; ITERATIVE RECONSTRUCTION; TEXTURE; CHEST;
D O I
10.1007/s00330-022-09003-y
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Objectives To assess the impact of a new artificial intelligence deep-learning reconstruction (Precise Image; AI-DLR) algorithm on image quality against a hybrid iterative reconstruction (IR) algorithm in abdominal CT for different clinical indications. Methods Acquisitions on phantoms were performed at 5 dose levels (CTDIvol: 13/11/9/6/1.8 mGy). Raw data were reconstructed using level 4 of iDose(4) (i4) and 3 levels of AI-DLR (Smoother/Smooth/Standard). Noise power spectrum (NPS), task-based transfer function (TTF) and detectability index (d ') were computed: d ' modelled detection of a liver metastasis (LM) and hepatocellular carcinoma at portal (HCCp) and arterial (HCCa) phases. Image quality was subjectively assessed on an anthropomorphic phantom by 2 radiologists. Results From Standard to Smoother levels, noise magnitude and average NPS spatial frequency decreased and the detectability (d ') of all simulated lesions increased. For both inserts, TTF values were similar for all three AI-DLR levels from 13 to 6 mGy but decreased from Standard to Smoother levels at 1.8 mGy. Compared to the i4 used in clinical practice, d ' values were higher using the Smoother and Smooth levels and close for the Standard level. For all dose levels, except at 1.8 mGy, radiologists considered images satisfactory for clinical use for the 3 levels of AI-DLR, but rated images too smooth using the Smoother level. Conclusion Use of the Smooth and Smoother levels of AI-DLR reduces the image noise and improves the detectability of lesions and spatial resolution for standard and low-dose levels. Using the Smooth level is apparently the best compromise between the lowest dose level and adequate image quality.
引用
收藏
页码:699 / 710
页数:12
相关论文
共 50 条
  • [1] Improved image quality and dose reduction in abdominal CT with deep-learning reconstruction algorithm: a phantom study
    Joël Greffier
    Quentin Durand
    Julien Frandon
    Salim Si-Mohamed
    Maeliss Loisy
    Fabien de Oliveira
    Jean-Paul Beregi
    Djamel Dabli
    [J]. European Radiology, 2023, 33 : 699 - 710
  • [2] Impact of an artificial intelligence deep-learning reconstruction algorithm for CT on image quality and potential dose reduction: A phantom study
    Greffier, Joel
    Si-Mohamed, Salim
    Frandon, Julien
    Loisy, Maeliss
    de Oliveira, Fabien
    Beregi, Jean Paul
    Dabli, Djamel
    [J]. MEDICAL PHYSICS, 2022, 49 (08) : 5052 - 5063
  • [3] Image quality and dose reduction opportunity of deep learning image reconstruction algorithm for CT: a phantom study
    Greffier, Joel
    Hamard, Aymeric
    Pereira, Fabricio
    Barrau, Corinne
    Pasquier, Hugo
    Beregi, Jean Paul
    Frandon, Julien
    [J]. EUROPEAN RADIOLOGY, 2020, 30 (07) : 3951 - 3959
  • [4] Image quality and dose reduction opportunity of deep learning image reconstruction algorithm for CT: a phantom study
    Joël Greffier
    Aymeric Hamard
    Fabricio Pereira
    Corinne Barrau
    Hugo Pasquier
    Jean Paul Beregi
    Julien Frandon
    [J]. European Radiology, 2020, 30 : 3951 - 3959
  • [5] Comparison of two versions of a deep learning image reconstruction algorithm on CT image quality and dose reduction: A phantom study
    Greffier, Joel
    Dabli, Djamel
    Frandon, Julien
    Hamard, Aymeric
    Belaouni, Asmaa
    Akessoul, Philippe
    Fuamba, Yannick
    Le Roy, Julien
    Guiu, Boris
    Beregi, Jean-Paul
    [J]. MEDICAL PHYSICS, 2021, 48 (10) : 5743 - 5755
  • [6] Deep learning image reconstruction algorithm for abdominal multidetector CT at different tube voltages: assessment of image quality and radiation dose in a phantom study
    Hye Joo Park
    Seo-Youn Choi
    Ji Eun Lee
    Sanghyeok Lim
    Min Hee Lee
    Boem Ha Yi
    Jang Gyu Cha
    Ji Hye Min
    Bora Lee
    Yunsub Jung
    [J]. European Radiology, 2022, 32 : 3974 - 3984
  • [7] Deep learning image reconstruction algorithm for abdominal multidetector CT at different tube voltages: assessment of image quality and radiation dose in a phantom study
    Park, Hye Joo
    Choi, Seo-Youn
    Lee, Ji Eun
    Lim, Sanghyeok
    Lee, Min Hee
    Yi, Boem Ha
    Cha, Jang Gyu
    Min, Ji Hye
    Lee, Bora
    Jung, Yunsub
    [J]. EUROPEAN RADIOLOGY, 2022, 32 (06) : 3974 - 3984
  • [8] Comparison of image quality of two versions of deep-learning image reconstruction algorithm on a rapid kV-switching CT: a phantom study
    Dabli, Djamel
    Loisy, Maeliss
    Frandon, Julien
    de Oliveira, Fabien
    Meerun, Azhar Mohamad
    Guiu, Boris
    Beregi, Jean-Paul
    Greffier, Joel
    [J]. EUROPEAN RADIOLOGY EXPERIMENTAL, 2023, 7 (01)
  • [9] Comparison of image quality of two versions of deep-learning image reconstruction algorithm on a rapid kV-switching CT: a phantom study
    Djamel Dabli
    Maeliss Loisy
    Julien Frandon
    Fabien de Oliveira
    Azhar Mohamad Meerun
    Boris Guiu
    Jean-Paul Beregi
    Joël Greffier
    [J]. European Radiology Experimental, 7
  • [10] Dose reduction and image quality in CT examinations using an iterative reconstruction algorithm: a phantom study
    Guariglia, S.
    Meliado, G.
    Zivelonghi, E.
    Pinali, L.
    Montemezzi, S.
    Cavedon, C.
    [J]. BIOMEDICAL PHYSICS & ENGINEERING EXPRESS, 2015, 1 (04):