Comparative performance analysis for abdominal phantom ROI detectability according to CT reconstruction algorithm: ADMIRE

被引:2
|
作者
Shin, Jun-Bong [1 ,2 ]
Yoon, Do-Kun [1 ,2 ]
Pak, Seongyong [3 ]
Kwon, Yang-Ho [3 ]
Suh, Tae Suk [1 ,2 ]
机构
[1] Catholic Univ Korea, Dept Biomed Engn, Coll Med, Seoul, South Korea
[2] Catholic Univ Korea, Coll Med, Res Inst Biomed Engn, Seoul, South Korea
[3] Siemens Healthineers Ltd, Seoul, South Korea
来源
关键词
abdomen phantom; ADMIRE; FBP; MDCT; SAFIRE; MODELED ITERATIVE RECONSTRUCTION; LOW-CONTRAST DETECTABILITY; RADIATION-DOSE REDUCTION; COMPUTED-TOMOGRAPHY;
D O I
10.1002/acm2.12765
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose We compared and analyzed the detectability performance pertaining to an abdominal phantom including a region of interest (ROI) according to a computed tomography (CT) reconstruction algorithm. Methods Three types of reconstruction algorithms (FBP, SAFIRE, and ADMIRE) were used to evaluate the detectability performance using the abdominal phantom (phantom size: 25 x 18 x 28 cm(3)). The vendor default settings for routine multi-detector computed tomography abdominal scans were used. As the quantitative evaluation method, the contrast-to-noise ratio (CNR), difference in coefficient of variation (COV) with the normalization based on the FBP data, and the noise power spectrum (NPS) were measured. Results The characteristic of the ADMIRE-3 reconstructed image was higher than those of the FBP and SAFIRE-3 reconstructed images. The CNR values of the SAFIRE and ADMIRE images were much higher than the corresponding values of the FBP images. The difference in COV values for the ADMIRE images was similar to 1.2 times lower than the corresponding values of the SAFIRE images. Conclusion The comparative analysis of the abdominal phantom low-contrast resolution differences for each CT exposure parameters showed that ADMIRE demonstrated better results than SAFIRE and FBP in terms of contrast, CNR, COV difference, and 1D NPS. This indicates that ADMIRE can provide a clearer observation even with the same number of contrast objects as compared to SAFIRE and FBP owing to its better contrast resolution in the central part of the contrast hole at low kV.
引用
收藏
页码:136 / 143
页数:8
相关论文
共 50 条
  • [41] Semiquantitative analysis of amyloid PET/CT and the performance of its own CT images for ROI's delimitation
    Segovia Roman, F.
    Testart Dardel, N.
    Sanchez Vano, R.
    Sopena-Navales, P.
    Gonzalez-Jimenez, A.
    Sanchez Sanchez, R.
    Trivino-Ibanez, E.
    Ramirez Perez de Inestrosa, J.
    Gomez-Rio, M.
    EUROPEAN JOURNAL OF NUCLEAR MEDICINE AND MOLECULAR IMAGING, 2017, 44 : S596 - S597
  • [42] Deep learning reconstruction improves radiomics feature stability and discriminative power in abdominal CT imaging: a phantom study
    Michallek, Florian
    Genske, Ulrich
    Niehues, Stefan Markus
    Hamm, Bernd
    Jahnke, Paul
    EUROPEAN RADIOLOGY, 2022, 32 (07) : 4587 - 4595
  • [43] Deep learning reconstruction improves radiomics feature stability and discriminative power in abdominal CT imaging: a phantom study
    Florian Michallek
    Ulrich Genske
    Stefan Markus Niehues
    Bernd Hamm
    Paul Jahnke
    European Radiology, 2022, 32 : 4587 - 4595
  • [44] Comparative analysis of the radiation shield effect in an abdominal CT scan
    Kim, Seon-Chil
    Kim, Young-Jae
    Lee, Joon-Seok
    Dong, Kyung-Rae
    Chung, Woon-Kwan
    Lim, Chang-Seon
    JOURNAL OF THE KOREAN PHYSICAL SOCIETY, 2014, 64 (06) : 929 - 935
  • [45] Comparative analysis of the radiation shield effect in an abdominal CT scan
    Seon-Chil Kim
    Young-Jae Kim
    Joon-Seok Lee
    Kyung-Rae Dong
    Woon-Kwan Chung
    Chang-Seon Lim
    Journal of the Korean Physical Society, 2014, 64 : 929 - 935
  • [46] Feasibility of MMWF noise reduction algorithm in brain SPECT images according to various reconstruction methods: A phantom study
    Kim, Eun Soo
    Lee, Youngjin
    Park, Chan Rok
    OPTIK, 2021, 247
  • [47] Iterative Reconstruction Algorithm for CT: Can Radiation Dose Be Decreased While Low-Contrast Detectability Is Preserved?
    Schindera, Sebastian T.
    Odedra, Devang
    Raza, Syed Arsalan
    Kim, Tae Kyoung
    Jang, Hyun-Jung
    Szucs-Farkas, Zsolt
    Rogalla, Patrik
    RADIOLOGY, 2013, 269 (02) : 510 - 517
  • [48] On the Relationship of Minimum Required Dose to Low Contrast Detectability and Blending Fractions With Adaptive Statistical Iterative Reconstruction (ASIR) in Abdominal CT
    Zhou, Y.
    Scott, A.
    Allahverdian, J.
    Lee, C.
    MEDICAL PHYSICS, 2015, 42 (06) : 3544 - 3544
  • [49] A comparative analysis of the dependences of the hemodynamic parameters on changes in ROI’s position in perfusion CT scans
    Yong-Seok Choi
    Jae-Hwan Cho
    Jang-Sun Namgung
    Hyo-Jin Kim
    Dae-Young Yoon
    Han-Joo Lee
    Journal of the Korean Physical Society, 2013, 62 : 1340 - 1345
  • [50] Performance evaluation of iterative reconstruction algorithms for achieving CT radiation dose reduction - a phantom study
    Dodge, Cristina T.
    Tamm, Eric P.
    Cody, Dianna D.
    Liu, Xinming
    Jensen, Corey T.
    Wei, Wei
    Kundra, Vikas
    Rong, X. John
    JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2016, 17 (02): : 511 - 531