Rapid 2D variable flip angle method for accurate and precise T1 measurements over a wide range of T1 values

被引:1
|
作者
Lena, Beatrice [1 ,2 ]
Bos, Clemens [2 ]
Ferrer, Cyril J. [2 ]
Moonen, Chrit T. W. [2 ]
Viergever, Max A. [1 ]
Bartels, Lambertus W. [1 ,2 ]
机构
[1] Univ Med Ctr Utrecht, Image Sci Inst, Imaging & Oncol Div, Utrecht, Netherlands
[2] Univ Med Ctr Utrecht, Imaging & Oncol Div, Image Guided Mol Intervent Grp, Utrecht, Netherlands
关键词
slice excitation profile; T-1; mapping; variable flip angle;
D O I
10.1002/nbm.4542
中图分类号
Q6 [生物物理学];
学科分类号
071011 ;
摘要
Purpose To perform dynamic T-1 mapping using a 2D variable flip angle (VFA) method, a correction for the slice profile effect is needed. In this work we investigated the impact of flip angle selection and excitation RF pulse profile on the performance of slice profile correction when applied to T-1 mapping over a range of T-1 values. Methods A correction of the slice profile effect is proposed, based on Bloch simulation of steady-state signals. With this correction, Monte Carlo simulations were performed to assess the accuracy and precision of 2D VFA T-1 mapping in the presence of noise, for RF pulses with time-bandwidth products of 2, 3 and 10 and with flip angle pairs in the range [1 degrees-90 degrees]. To evaluate its performance over a wide range of T-1, maximum errors were calculated for six T-1 values between 50 ms and 1250 ms. The method was demonstrated using in vitro and in vivo experiments. Results Without corrections, 2D VFA severely underestimates T-1. Slice profile errors were effectively reduced with the correction based on simulations, both in vitro and in vivo. The precision and accuracy of the method depend on the nominal T-1 values, the FA pair, and the RF pulse shape. FA pairs leading to T-1 can be identified for the common RF shapes, for T-1 values between 50 ms and 1250 ms. Conclusions 2D VFA T-1 mapping with Bloch-simulation-based correction can deliver T-1 estimates that are accurate and precise to within 5% over a wide T-1 range.
引用
收藏
页数:14
相关论文
共 50 条
  • [1] Simultaneous proton resonance frequency shift thermometry and T1 measurements using a single reference variable flip angle T1 method
    Svedin, Bryant T.
    Payne, Allison
    Parker, Dennis L.
    MAGNETIC RESONANCE IN MEDICINE, 2019, 81 (05) : 3138 - 3152
  • [2] Rapid high-resolution T1 mapping by variable flip angles:: Accurate and precise measurements in the presence of radiofrequency field inhomogeneity
    Cheng, HLM
    Wright, GA
    MAGNETIC RESONANCE IN MEDICINE, 2006, 55 (03) : 566 - 574
  • [3] RAPID CALCULATION OF T1 USING VARIABLE FLIP ANGLE GRADIENT REFOCUSED IMAGING
    FRAM, EK
    HERFKENS, RJ
    JOHNSON, GA
    GLOVER, GH
    KARIS, JP
    SHIMAKAWA, A
    PERKINS, TG
    PELC, NJ
    MAGNETIC RESONANCE IMAGING, 1987, 5 (03) : 201 - 208
  • [4] Uncertainty in T1 mapping using the variable flip angle method with two flip angles
    Schabel, Matthias C.
    Morrell, Glen R.
    PHYSICS IN MEDICINE AND BIOLOGY, 2009, 54 (01): : N1 - N8
  • [5] Technical Note: The effect of 2D excitation profile on T1 measurement accuracy using the variable flip angle method with an average flip angle assumption
    Svedin, Bryant T.
    Parker, Dennis L.
    MEDICAL PHYSICS, 2017, 44 (11) : 5930 - 5937
  • [6] T1 Mapping Using Variable Flip Angle SPGR Data With Flip Angle Correction
    Liberman, Gilad
    Louzoun, Yoram
    Ben Bashat, Dafna
    JOURNAL OF MAGNETIC RESONANCE IMAGING, 2014, 40 (01) : 171 - 180
  • [7] Repeatability and reproducibility of variable flip angle T1 quantification in the prostate at 3 T
    Zhong, Xinran
    Shakeri, Sepideh
    Liu, Dapeng
    Sayre, James
    Raman, Steven S.
    Wu, Holden H.
    Sung, Kyunghyun
    JOURNAL OF MAGNETIC RESONANCE IMAGING, 2019, 49 (06) : 1730 - 1735
  • [8] T1 measurements incorporating flip angle calibration and correction in vivo
    Wang, Jinghua
    Qiu, Maolin
    Kim, Hyeonjin
    Constable, R. Todd
    JOURNAL OF MAGNETIC RESONANCE, 2006, 182 (02) : 283 - 292
  • [9] Silent T1 mapping using the variable flip angle method with B1 correction
    Ljungberg, Emil
    Wood, Tobias
    Solana, Ana Beatriz
    Kolind, Shannon
    Williams, Steven C. R.
    Wiesinger, Florian
    Barker, Gareth J.
    MAGNETIC RESONANCE IN MEDICINE, 2020, 84 (02) : 813 - 824
  • [10] Estimating T1 from multichannel variable flip angle SPGR sequences
    Trzasko, Joshua D.
    Mostardi, Petrice M.
    Riederer, Stephen J.
    Manduca, Armando
    MAGNETIC RESONANCE IN MEDICINE, 2013, 69 (06) : 1787 - 1794