Technical Note: On GAFChromic EBT-XD film and the lateral response artifact

被引:55
|
作者
Lewis, David F. [1 ]
Chan, Maria F. [2 ]
机构
[1] RC Film Consulting LLC, 54 Benedict Rd, Monroe, CT 06468 USA
[2] Mem Sloan Kettering Canc Ctr, 136 Mt View Blvd, Basking Ridge, NJ 07920 USA
关键词
radiochromic film dosimetry; QA; IMRT; radiotherapy; lateral response; FLAT-BED SCANNERS; RADIOCHROMIC FILM; NONUNIFORMITY CORRECTION; DOSIMETRY; CALIBRATION;
D O I
10.1118/1.4939226
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose: The new radiochromic film, GAFChromic EBT-XD, contains the same active material, lithium-10,12-pentacosadiynoate, as GAFChromic EBT3, but the crystalline form is different. This work investigates the effect of this change on the well-known lateral response artifact when EBT-XD film is digitized on a flatbed scanner. Methods: The dose response of a single production lot of EBT-XD was characterized by scanning an unexposed film plus a set of films exposed to doses between 2.5 and 50 Gy using 6 MV photons. To characterize the lateral response artifact, the authors used the unexposed film plus a subset of samples exposed to doses between 20 and 50 Gy. Digital images of these films were acquired at seven discrete lateral locations perpendicular to the scan direction on three Epson 10000XL scanners. Using measurements at the discrete lateral positions, the scanner responses were determined as a function of the lateral position of the film From the data for each scanner, a set of coefficients were derived whereby measured response values could be corrected to remove the effects of the lateral response artifact. The EBT-XD data were analyzed as in their previous work and compared to results reported for EBT3 in that paper. Results: For films scanned in the same orientation and having equal responses, the authors found that the lateral response artifact for EBT-XD and EBT3 films was remarkably similar. For both films, the artifact increases with increased net response. However, as EBT-XD is less sensitive than EBT3, a greater exposure dose is required to reach the same net response. On this basis, the lower sensitivity of EBT-XD relative to EBT3 results in less net response change for equal exposure and a reduction in the impact of the lateral response artifact. Conclusions: The shape of the crystalline active component in EBT-XD and EBT3 does not affect the fundamental existence of the lateral response artifact when the films are digitized on flatbed scanners. Owing its lower sensitivity, EBT-XD film requires higher dose to reach the same response as EBT3, resulting in lesser impact of the lateral response artifact. For doses >10 Gy, the slopes of the EBT-XD red and green channel dose response curves are greater than the corresponding ones for EBT3. For these two reasons, the authors prefer EBT-XD for doses exceeding about 10 Gy. 2016 American Association of Physicists in Medicine.
引用
收藏
页码:643 / 649
页数:7
相关论文
共 50 条
  • [31] Investigating the method of selection of background pixel values for the calibration of EBT-XD film dosimetry
    Raj, Sathiya
    Shree, Nithya
    Kadirampatti, Ganesh
    PRECISION RADIATION ONCOLOGY, 2024, 8 (03): : 132 - 137
  • [32] Monte Carlo Calculated Stopping-Power of EBT3 and EBT-XD Film for Pristine Proton Beams
    Shi, C.
    Chen, C.
    Mah, D.
    Chan, M.
    MEDICAL PHYSICS, 2017, 44 (06)
  • [33] Neutron response of GafChromic® EBT2 film
    Hsiao, Ming-Chen
    Liu, Yuan-Hao
    Chen, Wei-Lin
    Jiang, Shiang-Huei
    PHYSICS IN MEDICINE AND BIOLOGY, 2013, 58 (05): : 1391 - 1413
  • [34] Technical Note: Potential errors in optical density measurements due to scanning side in EBT and EBT2 Gafchromic film dosimetry
    Desroches, Joannie
    Bouchard, Hugo
    Lacroix, Frederic
    MEDICAL PHYSICS, 2010, 37 (04) : 1565 - 1570
  • [35] An Evaluation of the Use of EBT-XD Film for SRS/SBRT Commissioning of Unity MR-Linac System
    Lim, S.
    Tyagi, N.
    Subashi, E.
    Liang, J.
    Chan, M.
    MEDICAL PHYSICS, 2021, 48 (06)
  • [36] AcurosXB dose verification of ultra-small lung lesions with EBT-XD film in a heterogeneous phantom
    Ollers, M.
    Swinnen, A.
    Verhaegen, F.
    RADIOTHERAPY AND ONCOLOGY, 2019, 133 : S479 - S479
  • [37] Dose mapping in Cesium-137 blood irradiator using novel Ge-doped silica optical fibres in comparison with gafchromic EBT-XD film: A preliminary study
    Bakar, K. S. A. K.
    Noor, N. M.
    Hassan, M. F.
    Idris, F.
    Bradley, D. A.
    RADIATION PHYSICS AND CHEMISTRY, 2023, 209
  • [38] Effect of glass compression plate on EBT-XD film dosimetry for pretreatment quality assurance of stereotactic body radiotherapy
    Raj, Sathiya
    Shree, Nithya
    Ganesh, K. M.
    REPORTS OF PRACTICAL ONCOLOGY AND RADIOTHERAPY, 2024, 29 (03) : 357 - 361
  • [39] Assessment of stereotactic high-resolution detectors for stereotactic body radiotherapy: comparative analysis between myQA® SRS and Gafchromic EBT-XD films
    Chinnaiya, Dinesan
    Mudhana, Gopinath
    RADIATION AND ENVIRONMENTAL BIOPHYSICS, 2024, 63 (02) : 203 - 214
  • [40] An evaluation of the use of EBT-XD film for SRS/SBRT commissioning of a 1.5 Tesla MR-Linac system
    Lim, Seng Boh
    Tyagi, Neelam
    Subashi, Ergys
    Liang, Jiayi
    Chan, Maria
    PHYSICA MEDICA-EUROPEAN JOURNAL OF MEDICAL PHYSICS, 2022, 96 : 9 - 17