Technical Note: On GAFChromic EBT-XD film and the lateral response artifact

被引:55
|
作者
Lewis, David F. [1 ]
Chan, Maria F. [2 ]
机构
[1] RC Film Consulting LLC, 54 Benedict Rd, Monroe, CT 06468 USA
[2] Mem Sloan Kettering Canc Ctr, 136 Mt View Blvd, Basking Ridge, NJ 07920 USA
关键词
radiochromic film dosimetry; QA; IMRT; radiotherapy; lateral response; FLAT-BED SCANNERS; RADIOCHROMIC FILM; NONUNIFORMITY CORRECTION; DOSIMETRY; CALIBRATION;
D O I
10.1118/1.4939226
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose: The new radiochromic film, GAFChromic EBT-XD, contains the same active material, lithium-10,12-pentacosadiynoate, as GAFChromic EBT3, but the crystalline form is different. This work investigates the effect of this change on the well-known lateral response artifact when EBT-XD film is digitized on a flatbed scanner. Methods: The dose response of a single production lot of EBT-XD was characterized by scanning an unexposed film plus a set of films exposed to doses between 2.5 and 50 Gy using 6 MV photons. To characterize the lateral response artifact, the authors used the unexposed film plus a subset of samples exposed to doses between 20 and 50 Gy. Digital images of these films were acquired at seven discrete lateral locations perpendicular to the scan direction on three Epson 10000XL scanners. Using measurements at the discrete lateral positions, the scanner responses were determined as a function of the lateral position of the film From the data for each scanner, a set of coefficients were derived whereby measured response values could be corrected to remove the effects of the lateral response artifact. The EBT-XD data were analyzed as in their previous work and compared to results reported for EBT3 in that paper. Results: For films scanned in the same orientation and having equal responses, the authors found that the lateral response artifact for EBT-XD and EBT3 films was remarkably similar. For both films, the artifact increases with increased net response. However, as EBT-XD is less sensitive than EBT3, a greater exposure dose is required to reach the same net response. On this basis, the lower sensitivity of EBT-XD relative to EBT3 results in less net response change for equal exposure and a reduction in the impact of the lateral response artifact. Conclusions: The shape of the crystalline active component in EBT-XD and EBT3 does not affect the fundamental existence of the lateral response artifact when the films are digitized on flatbed scanners. Owing its lower sensitivity, EBT-XD film requires higher dose to reach the same response as EBT3, resulting in lesser impact of the lateral response artifact. For doses >10 Gy, the slopes of the EBT-XD red and green channel dose response curves are greater than the corresponding ones for EBT3. For these two reasons, the authors prefer EBT-XD for doses exceeding about 10 Gy. 2016 American Association of Physicists in Medicine.
引用
收藏
页码:643 / 649
页数:7
相关论文
共 50 条
  • [21] Effect of Magnetic Field On the Response of EBT-XD and EBT3 Radiochromic Films in MRIgRT
    Darafsheh, A.
    Hao, Y.
    Maraghechi, B.
    Khan, R.
    MEDICAL PHYSICS, 2020, 47 (06) : E761 - E762
  • [22] Characterization of the Response of EBT-XD Radiochromic Films to Proton Therapy Beams
    Darafsheh, A.
    Zhao, T.
    Khan, R.
    MEDICAL PHYSICS, 2020, 47 (06) : E717 - E717
  • [23] Performance evaluation of an LED flatbed scanner for triple channel film dosimetry with EBT3 and EBT-XD film
    Marchant Van der Walt
    Linda Marsh
    John Baines
    Stephen Gibson
    Ariadne Shoobridge
    Glenn de Vine
    Physical and Engineering Sciences in Medicine, 2022, 45 : 901 - 914
  • [24] Evaluation of the Response of EBT3, EBT-XD, MD-V3, and HD-V2 GafChromic Films in Proton FLASH Fields
    Darafsheh, A.
    Hao, Y.
    Owen, D.
    Zwart, T.
    Wagner, M.
    Zhao, T.
    MEDICAL PHYSICS, 2022, 49 (06) : E193 - E193
  • [25] Performance evaluation of an LED flatbed scanner for triple channel film dosimetry with EBT3 and EBT-XD film
    Van der Walt, Marchant
    Marsh, Linda
    Baines, John
    Gibson, Stephen
    Shoobridge, Ariadne
    de Vine, Glenn
    PHYSICAL AND ENGINEERING SCIENCES IN MEDICINE, 2022, 45 (3) : 901 - 914
  • [26] Proton beam dosimetry at ultra-high dose rates (FLASH): Evaluation of GAFchromic™ (EBT3, EBT-XD) and OrthoChromic (OC-1) film performances
    Villoing, Daphnee
    Koumeir, Charbel
    Bongrand, Arthur
    Guertin, Arnaud
    Haddad, Ferid
    Metivier, Vincent
    Poirier, Freddy
    Potiron, Vincent
    Servagent, Noel
    Supiot, Stephane
    Delpon, Gregory
    Chiavassa, Sophie
    MEDICAL PHYSICS, 2022, 49 (04) : 2732 - 2745
  • [27] Influence of 0.35 T magnetic field on the response of EBT3 and EBT-XD radiochromic films
    Darafsheh, Arash
    Hao, Yao
    Maraghechi, Borna
    Cammin, Jochen
    Reynoso, Francisco J.
    Khan, Rao
    MEDICAL PHYSICS, 2020, 47 (09) : 4543 - 4552
  • [28] Response characterization of EBT-XD radiochromic films in megavoltage photon and electron beams
    Leon-Marroquin, E. Yazmin
    Mulrow, Daniel
    Darafsheh, Arash
    Khan, Rao
    MEDICAL PHYSICS, 2019, 46 (09) : 4246 - 4256
  • [29] The Effect of Scanning Orientation On New EBT-XD Film Using Vidar and Epson Scanners
    Chiu-Tsao, S.
    Chan, M.
    MEDICAL PHYSICS, 2016, 43 (06) : 3597 - 3597
  • [30] Is EBT-XD film suitable for linac and Gamma Knife radiosurgery dosimetry verification and audit?
    Nisbet, A.
    Dimitriadis, A.
    Palmer, A. L.
    Clark, C. H.
    RADIOTHERAPY AND ONCOLOGY, 2016, 119 : S723 - S723