Accuracy of Five Intraoral Scanners and Two Laboratory Scanners for a Complete Arch: A Comparative In Vitro Study

被引:18
|
作者
Kang, Byung-hyun [1 ]
Son, Keunbada [2 ,3 ]
Lee, Kyu-bok [1 ,3 ]
机构
[1] Kyungpook Natl Univ, Sch Dent, Dept Prosthodont, Daegu 41940, South Korea
[2] Kyungpook Natl Univ, Grad Sch, Dept Dent Sci, Daegu 41940, South Korea
[3] Kyungpook Natl Univ, Adv Dent Device Dev Inst, Daegu 41940, South Korea
来源
APPLIED SCIENCES-BASEL | 2020年 / 10卷 / 01期
关键词
intraoral scanner; accuracy; digital workflow; complete arch; dentistry; IMPRESSIONS; PRECISION; FIT;
D O I
10.3390/app10010074
中图分类号
O6 [化学];
学科分类号
0703 ;
摘要
This study aims to evaluate the accuracy of five different intraoral scanners and two different laboratory scanners for a complete arch. A computer-aided design (CAD) reference model (CRM) was obtained using industrial scanners. A CAD test model (CTM) was obtained using five types of intraoral scanners (CS3500, CS3600, Trios2, Trios3, and i500) and two types of laboratory scanners (3shape E1 and DOF) (N = 20). In addition, the CRM and CTM were superimposed using a 3D inspection software (Geomagic control X; 3D Systems) and 3D analysis was performed. In the 3D analysis, the accuracy was measured by the type of tooth, the anterior and posterior region, and the overall region. As for the statistical analysis of the accuracy, the differences were confirmed using the Kruskal-Wallis H test (alpha = 0.05). Also, the differences between the groups were analyzed by post-hoc tests including Mann-Whitney U-test and Bonferroni correction method (alpha = 0.0017). There was a significant difference in the scanning accuracy of the complete arch according to the type of scanner (P < 0.001). The i500 Group showed the lowest accuracy (143 +/- 69.6 mu m), while the 3Shape E1 Group was the most accurate (14.3 +/- 0.3 mu m). Also, the accuracy was lower in the posterior region than in the anterior region in all types of scanners (P < 0.001). Scanning accuracy of the complete arch differed depending on the type of scanner. While three types of intraoral scanners (CS3500, CS3600, Trios3) can be recommended for scanning of a complete arch, the two remaining types of intraoral scanners (Trios2 and i500) cannot be recommended.
引用
收藏
页数:11
相关论文
共 50 条
  • [1] Complete-Arch Accuracy of Four Intraoral Scanners: An In Vitro Study
    Celeghin, Giordano
    Franceschetti, Giulio
    Mobilio, Nicola
    Fasiol, Alberto
    Catapano, Santo
    Corsalini, Massimo
    Grande, Francesco
    [J]. HEALTHCARE, 2021, 9 (03):
  • [2] Complete-arch accuracy of intraoral scanners
    Treesh, Joshua C.
    Liacouras, Peter C.
    Taft, Robert M.
    Brooks, Daniel, I
    Raiciulescu, Sorana
    Ellert, Daniel O.
    Grant, Gerald T.
    Ye, Ling
    [J]. JOURNAL OF PROSTHETIC DENTISTRY, 2018, 120 (03): : 382 - 388
  • [3] Accuracy of intraoral scanners in different complete arch scan patterns
    Pattamavilai, Sakaorat
    Ongthiemsak, Chakree
    [J]. JOURNAL OF PROSTHETIC DENTISTRY, 2024, 131 (01): : 155 - 162
  • [4] Trueness and Precision of Two Intraoral Scanners: A Comparative In Vitro Study
    Rotar, Raul Nicolae
    Jivanescu, Anca
    Ille, Codruta
    Podariu, Angela Codruta
    Jumanca, Daniela Elisabeta
    Matichescu, Ana-Maria
    Balean, Octavia
    Rusu, Laura Cristina
    [J]. SCANNING, 2019, 2019
  • [5] Accuracy of four intraoral scanners in oral implantology: a comparative in vitro study
    Imburgia, Mario
    Logozzo, Silvia
    Hauschild, Uli
    Veronesi, Giovanni
    Mangano, Carlo
    Mangano, Francesco Guido
    [J]. BMC ORAL HEALTH, 2017, 17
  • [6] Accuracy of four intraoral scanners in oral implantology: a comparative in vitro study
    Mario Imburgia
    Silvia Logozzo
    Uli Hauschild
    Giovanni Veronesi
    Carlo Mangano
    Francesco Guido Mangano
    [J]. BMC Oral Health, 17
  • [7] Accuracy evaluation of two different intraoral scanners in implant prosthodontics. A comparative in vitro study
    Verniani, G.
    Casucci, A.
    Nosrati, N.
    D'Arienzo, L. F.
    Val, M.
    Cagidiaco, E. Ferrari
    [J]. JOURNAL OF OSSEOINTEGRATION, 2024, 16 (01) : 61 - 64
  • [8] Repeatability of Intraoral Scanners for Complete Arch Scan of Partially Edentulous Dentitions: An In Vitro Study
    Lee, Jae-Hyun
    Yun, Je-Hyeon
    Han, Jung-Suk
    Yeo, In-Sung Luke
    Yoon, Hyung-In
    [J]. JOURNAL OF CLINICAL MEDICINE, 2019, 8 (08)
  • [9] A new method to measure the accuracy of intraoral scanners along the complete dental arch: A pilot study
    Iturrate, Mikel
    Lizundia, Erlantz
    Amezua, Xabier
    Solaberrieta, Eneko
    [J]. JOURNAL OF ADVANCED PROSTHODONTICS, 2019, 11 (06): : 331 - 340
  • [10] The time efficiency of intraoral scanners An in vitro comparative study
    Patzelt, Sebastian B. M.
    Lamprinos, Christos
    Stampf, Susanne
    Att, Wael
    [J]. JOURNAL OF THE AMERICAN DENTAL ASSOCIATION, 2014, 145 (06): : 542 - 551