Dual-energy CT angiography of abdomen with routine concentration contrast agent in comparison with conventional single-energy CT with high concentration contrast agent

被引:25
|
作者
He, Jingzhen [1 ]
Wang, Qing [1 ]
Ma, Xiangxing [1 ]
Sun, Zhiyuan [1 ]
机构
[1] Shandong Univ, Qilu Hosp, Dept Radiol, Jinan 250012, Shandong, Peoples R China
关键词
Dual-energy CT; CT angiography; Contrast agent; REDUCED IODINE LOAD; IMAGE-QUALITY; SPECTRAL CT; PULMONARY ANGIOGRAPHY; RENAL-INSUFFICIENCY; DOSE REDUCTION; TUBE-VOLTAGE; ENHANCEMENT; OPTIMIZATION; 16-MDCT;
D O I
10.1016/j.ejrad.2014.11.025
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose: To compare the quantitative and subjective image quality in abdominal angiography between dual-energy CT (DECT) at the routine concentration of iodinated contrast agent (300 mg/mL) and conventional 120-kVp single-energy CT (SECT) at the high concentration of contrast agent (370 mg/mL). Materials and methods: Abdominal computed tomography angiography (CTA) was performed in 104 patients, including 56 with conventional 120-kVp SECT at the high concentration of contrast agent and 48 with DECT at the routine concentration of contrast agent. The monochromatic images at the optimal kiloelectron-voltage (key) of DECT that demonstrated the best contrast-to-noise ratio were reconstructed. The signal intensity and noise in abdominal arteries were comparatively analyzed between DECT and SECT. The image quality and visibility of the branch orders of superior mesenteric artery and renal arteries were further assessed. The radiation doses were recorded. Results: Compared with SECT, DECT demonstrated higher signal intensity, signal-to-noise ratio, and contrast-to-noise ratio (all P < 0.01) with moderately increased noise (40%, P < 0.01) in all abdominal arteries. The image quality of DECT was superior to that of SECT (P < 0.01) as evaluated with a subjective five-point scale system. Visualization of the branches of superior mesenteric artery and renal arteries was also better by DECT (P < 0.01) than SECT. The radiation dose of DECT was slight higher than that of SECT (P < 0.0001). Conclusion: DECT with image reconstruction at the optimal keV provides a high-quality angiographic technique, which allows use of a lower concentration of contrast agent compared with conventional 120-kVp SECT. (C) 2014 Elsevier Ireland Ltd. All rights reserved.
引用
收藏
页码:221 / 227
页数:7
相关论文
共 50 条
  • [21] Initial experience with single-source dual-energy CT abdominal angiography and comparison with single-energy CT angiography: image quality, enhancement, diagnosis and radiation dose
    Daniella F. Pinho
    Naveen M. Kulkarni
    Arun Krishnaraj
    Sanjeeva P. Kalva
    Dushyant V. Sahani
    European Radiology, 2013, 23 : 351 - 359
  • [22] Quantitative Imaging of a Gold Nanoparticle Contrast Agent with Dual Energy CT
    Ducote, J. L.
    Le, H. Q.
    Alivov, Y.
    Molloi, S.
    MEDICAL PHYSICS, 2010, 37 (06)
  • [23] Initial experience with single-source dual-energy CT abdominal angiography and comparison with single-energy CT angiography: image quality, enhancement, diagnosis and radiation dose
    Pinho, Daniella F.
    Kulkarni, Naveen M.
    Krishnaraj, Arun
    Kalva, Sanjeeva P.
    Sahani, Dushyant V.
    EUROPEAN RADIOLOGY, 2013, 23 (02) : 351 - 359
  • [24] Myocardial extracellular volume fraction analysis in doxorubicin-induced beagle models: comparison of dual-energy CT with equilibrium contrast-enhanced single-energy CT
    Zhou, Zhen
    Gao, Yifeng
    Wang, Hongwei
    Wang, Wenjing
    Zhang, Hongkai
    Wang, Sicong
    Sun, Zhonghua
    Xu, Lei
    CARDIOVASCULAR DIAGNOSIS AND THERAPY, 2021, 11 (01) : 102 - 110
  • [25] Aortic dual-energy CT angiography with low contrast medium injection rate
    Liu, Jie
    Lv, Pei-Jie
    Wu, Runze
    Zhang, Yong-Gao
    Hu, Li-Li
    Hou, Ping
    Gao, Jian-Bo
    JOURNAL OF X-RAY SCIENCE AND TECHNOLOGY, 2014, 22 (05) : 689 - 696
  • [26] INFLUENCE OF CALIBRATION MATERIALS IN SINGLE-ENERGY AND DUAL-ENERGY QUANTITATIVE CT
    STEENBEEK, JCM
    VANKUIJK, C
    GRASHUIS, JL
    RADIOLOGY, 1992, 183 (03) : 849 - 855
  • [27] Quantitative comparison of dual-energy CT and differential phase contrast CT imaging methods
    Zhang, Xin
    Su, Ting
    Xia, Dongmei
    Liang, Dong
    Ge, Yongshuai
    MEDICAL IMAGING 2022: PHYSICS OF MEDICAL IMAGING, 2022, 12031
  • [28] Robustness of CT radiomics features: consistency within and between single-energy CT and dual-energy CT
    Chen, Yong
    Zhong, Jingyu
    Wang, Lan
    Shi, Xiaomeng
    Lu, Wei
    Li, Jianying
    Feng, Jianxing
    Xia, Yihan
    Chang, Rui
    Fan, Jing
    Chen, Liwei
    Zhu, Ying
    Yan, Fuhua
    Yao, Weiwu
    Zhang, Huan
    EUROPEAN RADIOLOGY, 2022, 32 (08) : 5480 - 5490
  • [29] Robustness of CT radiomics features: consistency within and between single-energy CT and dual-energy CT
    Yong Chen
    Jingyu Zhong
    Lan Wang
    Xiaomeng Shi
    Wei Lu
    Jianying Li
    Jianxing Feng
    Yihan Xia
    Rui Chang
    Jing Fan
    Liwei Chen
    Ying Zhu
    Fuhua Yan
    Weiwu Yao
    Huan Zhang
    European Radiology, 2022, 32 : 5480 - 5490
  • [30] Dual-energy MDCT: Comparison of pulmonary artery enhancement on dedicated CT pulmonary angiography, routine and low contrast volume studies
    Godoy, Myrna C. B.
    Heller, Samantha L.
    Naidich, David P.
    Assadourian, Bernard
    Leidecker, Christianne
    Schmidt, Bernhard
    Vlahos, Ioannis
    EUROPEAN JOURNAL OF RADIOLOGY, 2011, 79 (02) : E11 - E17