Cuff-Less Blood Pressure Estimation Using Pulse Waveform Analysis and Pulse Arrival Time

被引:61
|
作者
Yoon, Young-Zoon [1 ]
Kang, Jae Min [1 ]
Kwon, Yongjoo [1 ]
Park, Sangyun [1 ]
Noh, Seungwoo [1 ]
Kim, Younho [1 ]
Park, Jongae [1 ]
Hwang, Sung Woo [1 ]
机构
[1] Samsung Adv Inst Technol, Suwon 16678, South Korea
关键词
Cuff-less blood pressure measurement; photoplethysmography; pulse arrival time; pulse waveform analysis; wearable device; TRANSIT-TIME; NONINVASIVE ESTIMATION; VELOCITY; EXERCISE; SIGNALS; MARKER; ECG; PPG;
D O I
10.1109/JBHI.2017.2714674
中图分类号
TP [自动化技术、计算机技术];
学科分类号
0812 ;
摘要
Using the massive MIMIC physiological data base, we tried to validate pulse wave analysis (PWA) based on multiparameters model whether it can continuously estimate blood pressure (BP) values on single site of one hand. In addition, to consider the limitation of insufficient data acquirement for home user, we used pulse arrival time (PAT) driven BP information to determine the individual scale factors of the PWA-BP estimation model. Experimental results indicate that the accuracy of the average regression model has error standard deviations of 10.6 +/- 3.3 mmHg (PAT), 8.7 +/- 3.2 mmHg (PWA) for SBP and 6.0 +/- 2.3 mmHg (PAT), 4.4 +/- 1.6 mmHg (PWA) for DBP on 23 subjects over a 1 day period. We defined a local-model which is extracted regression model from sparsely selected small dataset, contrast to full dataset for 24h (average-model). The limit of BP estimation accuracy from the local-model of PWA is lower than that of PAT-BP average-model. Whereas the error of the BP estimation local-model was reduced using more data for scaling, it required more than four times the 1 min data extracted over the 12 h calibration period to predict BP for 1 day. This study shows that PWA has possibility to estimate BP value and PAT-driven BP information could be used to determine the individual scale factors of the PWA-BP estimation model for home users.
引用
收藏
页码:1068 / 1074
页数:7
相关论文
共 50 条
  • [41] Estimation of pulse transit time using two diametric blood pressure waveform measurements
    Hahn, Jin-Oh
    Reisner, Andrew T.
    Asada, H. Harry
    MEDICAL ENGINEERING & PHYSICS, 2010, 32 (07) : 753 - 759
  • [42] Posture-Dependent Variability in Wrist Ballistocardiogram-Photoplethysmogram Pulse Transit Time: Implication to Cuff-Less Blood Pressure Tracking
    Shin, Sungtae
    Mousavi, Azin
    Lyle, Sophia
    Jang, Elisabeth
    Yousefian, Peyman
    Mukkamala, Ramakrishna
    Jang, Dae-Geun
    Kwon, Ui Kun
    Kim, Youn Ho
    Hahn, Jin-Oh
    IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, 2022, 69 (01) : 347 - 355
  • [43] Pulse waveform as an indicator of baseline offset in pulse transit time based blood pressure estimation
    Lin, Chen
    Zhou, Yuan
    Wang, Hu
    Wang, Yao
    2017 IEEE-NIH HEALTHCARE INNOVATIONS AND POINT OF CARE TECHNOLOGIES (HI-POCT), 2017, : 26 - 31
  • [44] Pulse arrival time as a surrogate of blood pressure
    Finnegan, Eoin
    Davidson, Shaun
    Harford, Mirae
    Jorge, Joao
    Watkinson, Peter
    Young, Duncan
    Tarassenko, Lionel
    Villarroel, Mauricio
    SCIENTIFIC REPORTS, 2021, 11 (01)
  • [45] Pulse arrival time as a surrogate of blood pressure
    Eoin Finnegan
    Shaun Davidson
    Mirae Harford
    João Jorge
    Peter Watkinson
    Duncan Young
    Lionel Tarassenko
    Mauricio Villarroel
    Scientific Reports, 11
  • [46] Cuff-less blood pressure measurement from dual-channel photoplethysmographic signals via peripheral pulse transit time with singular spectrum analysis
    Wang, Yang
    Liu, Zhiwen
    Ma, Shaodong
    PHYSIOLOGICAL MEASUREMENT, 2018, 39 (02)
  • [47] Cuff-Less Blood Pressure Estimation via Small Convolutional Neural Networks
    Wang, Weinan
    Mohseni, Pedram
    Kilgore, Kevin
    Najafizadeh, Laleh
    2021 43RD ANNUAL INTERNATIONAL CONFERENCE OF THE IEEE ENGINEERING IN MEDICINE & BIOLOGY SOCIETY (EMBC), 2021, : 1031 - 1034
  • [48] Continuous Cuff-Less Blood Pressure Estimation Using Lightweight Binary Tree Neural Network
    Li, Yongjian
    Chen, Meng
    Wang, Chunyuan
    Li, Yixue
    Wang, Yuchen
    Wei, Shoushui
    IEEE TRANSACTIONS ON INSTRUMENTATION AND MEASUREMENT, 2025, 74
  • [49] Nonlinear Cuffless Blood Pressure Estimation of Healthy Subjects Using Pulse Transit Time and Arrival Time
    Esmaili, Amirhossein
    Kachuee, Mohammad
    Shabany, Mahdi
    IEEE TRANSACTIONS ON INSTRUMENTATION AND MEASUREMENT, 2017, 66 (12) : 3299 - 3308
  • [50] A Novel Method for Calibration-Based Cuff-Less Blood Pressure Estimation
    Li, Zhenqi
    Yan, Cong
    Zhao, Wei
    Hu, Jing
    Jia, Dongya
    Wang, Hongmei
    You, Tianyuan
    2019 41ST ANNUAL INTERNATIONAL CONFERENCE OF THE IEEE ENGINEERING IN MEDICINE AND BIOLOGY SOCIETY (EMBC), 2019, : 4266 - 4269