Stability and Performance of the Binary Compensation Unit for Cryogenic Current Comparator Bridges

被引:6
|
作者
Gotz, Martin [1 ]
Drung, Dietmar [2 ]
机构
[1] Phys Tech Bundesanstalt, D-38116 Braunschweig, Germany
[2] Phys Tech Bundesanstalt, D-10587 Berlin, Germany
关键词
Calibration; current comparator; electrical resistance measurement; measurement uncertainty; precision measurement;
D O I
10.1109/TIM.2017.2659998
中图分类号
TM [电工技术]; TN [电子技术、通信技术];
学科分类号
0808 ; 0809 ;
摘要
As a key module of the redesigned cryogenic current comparator setup, the binary compensation unit is now in everyday use for resistance calibration at Physikalisch-Technische Bundesanstalt (PTB), as well as in a number of national metrology institutes. In a first step, its excellent performance was demonstrated in consistency checks involving the quantum Hall resistance and normal resistors with nominal values 100 Omega and 12.9 k Omega. Meanwhile, the stability of this compensation unit could be evaluated, too. Our analysis is based on the results of calibrations at regular intervals of 20 weeks covering more than three years. Also the influence of temperature has been studied experimentally. From the data presented here, we conclude that the device under test can be considered to be stable within 1 part in 10(4) over ten years.
引用
收藏
页码:1467 / 1474
页数:8
相关论文
共 50 条
  • [21] Improved Cryogenic Current Comparator Setup With Digital Current Sources
    Goetz, Martin
    Drung, Dietmar
    Pesel, Eckart
    Barthelmess, Henry-Jobes
    Hinnrichs, Colmar
    Assmann, Cornelia
    Peters, Margret
    Scherer, Hansjoerg
    Schumacher, Bernd
    Schurig, Thomas
    [J]. IEEE TRANSACTIONS ON INSTRUMENTATION AND MEASUREMENT, 2009, 58 (04) : 1176 - 1182
  • [22] Ultralow noise current amplifier based on a cryogenic current comparator
    Gay, F
    Piquemal, F
    Genevès, G
    [J]. REVIEW OF SCIENTIFIC INSTRUMENTS, 2000, 71 (12): : 4592 - 4595
  • [23] ISOLATED RAMPING CURRENT SOURCES FOR A CRYOGENIC CURRENT COMPARATOR BRIDGE
    ELMQUIST, RE
    DZIUBA, RF
    [J]. REVIEW OF SCIENTIFIC INSTRUMENTS, 1991, 62 (10): : 2457 - 2460
  • [24] AC resistance bridge based on the cryogenic current comparator
    Seppa, H
    Satrapinski, A
    [J]. IEEE TRANSACTIONS ON INSTRUMENTATION AND MEASUREMENT, 1997, 46 (02) : 463 - 466
  • [25] HIGH-TEMPERATURE SUPERCONDUCTOR CRYOGENIC CURRENT COMPARATOR
    ELMQUIST, RE
    DZIUBA, RF
    [J]. IEEE TRANSACTIONS ON INSTRUMENTATION AND MEASUREMENT, 1995, 44 (02) : 262 - 264
  • [26] Operation of a Cryogenic Current Comparator in Presence of Mechanical Vibrations
    Bierzychudek, Marcos E.
    Goetz, Martin
    Schmelzer, Martin
    [J]. 2016 CONFERENCE ON PRECISION ELECTROMAGNETIC MEASUREMENTS (CPEM 2016), 2016,
  • [27] A quantum Hall cryogenic current comparator resistance bridge
    Jones, K
    Early, MD
    [J]. 2000 CONFERENCE ON PRECISION ELECTROMAGNETIC MEASUREMENTS DIGEST, 2000, : 92 - 93
  • [28] Compact Cryogenic Current Comparator and Null Detector Assembly
    Williams, Jonathan M.
    Shelly, Connor D.
    King, Becky J.
    Edmonds, Kieran J.
    [J]. 2018 CONFERENCE ON PRECISION ELECTROMAGNETIC MEASUREMENTS (CPEM 2018), 2018,
  • [29] Optimum sensitivity of an externally shielded cryogenic current comparator
    Early, MD
    Jones, K
    [J]. IEEE TRANSACTIONS ON INSTRUMENTATION AND MEASUREMENT, 1997, 46 (02) : 459 - 462
  • [30] A cryogenic current comparator for the absolute measurement of nA beams
    Peters, A
    Vodel, W
    Koch, H
    Neubert, R
    Reeg, H
    Schroeder, CH
    [J]. BEAM INSTRUMENTATION WORKSHOP, 1998, 451 : 163 - 180