Ultralow noise current amplifier based on a cryogenic current comparator

被引:35
|
作者
Gay, F [1 ]
Piquemal, F [1 ]
Genevès, G [1 ]
机构
[1] Bur Natl Metrol, Lab Cent Ind Elect, F-92260 Fontenay Aux Roses, France
来源
REVIEW OF SCIENTIFIC INSTRUMENTS | 2000年 / 71卷 / 12期
关键词
D O I
10.1063/1.1326054
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
This article describes the development of a current amplifier operating at 4.2 K designed for very low current measurements. A cryogenic current comparator with a winding ratio of 10 000/1, a low noise dc superconducting quantum interference device, and a very efficient magnetic shield are combined to reach an equivalent input noise of 4 fA/root Hz in the signal bandwidth and a flicker noise corner around 0.5 Hz. (C) 2000 American Institute of Physics. [S0034-6748(00)05012-7].
引用
收藏
页码:4592 / 4595
页数:4
相关论文
共 50 条
  • [1] Medium size Cryogenic Current Comparator as very low current amplifier
    Gay, F
    Piquemal, F
    Geneves, G
    1998 CONFERENCE ON PRECISION ELECTROMAGNETIC MEASUREMENTS DIGEST, 1998, : 217 - 218
  • [2] Ultimate current resolution of a cryogenic current comparator
    Sesé, J
    Camón, A
    Rillo, C
    Rietveld, G
    IEEE TRANSACTIONS ON INSTRUMENTATION AND MEASUREMENT, 1999, 48 (06) : 1306 - 1313
  • [4] AC resistance bridge based on the cryogenic current comparator
    Seppa, H
    Satrapinski, A
    IEEE TRANSACTIONS ON INSTRUMENTATION AND MEASUREMENT, 1997, 46 (02) : 463 - 466
  • [5] Calibrating Ultrastable Low-Noise Current Amplifiers of the Second Generation With a Cryogenic Current Comparator
    Goetz, Martin
    Drung, Dietmar
    Krause, Christian
    Becker, Ulrich
    Scherer, Hansjoerg
    IEEE TRANSACTIONS ON INSTRUMENTATION AND MEASUREMENT, 2019, 68 (06) : 2027 - 2033
  • [6] LEAKAGE CURRENT DETECTION IN CRYOGENIC CURRENT COMPARATOR BRIDGES
    ELMQUIST, RE
    IEEE TRANSACTIONS ON INSTRUMENTATION AND MEASUREMENT, 1993, 42 (02) : 167 - 169
  • [7] CRYOGENIC CURRENT COMPARATOR WITH INCREASED RESOLUTION
    VANDERWEL, W
    MOOIJ, JE
    HARMANS, CJPM
    REVIEW OF SCIENTIFIC INSTRUMENTS, 1988, 59 (04): : 624 - 626
  • [8] Cryogenic current comparator implementation at CENAM
    Sánchez, C
    Avilés, D
    Hernández, D
    Durán, G
    2000 CONFERENCE ON PRECISION ELECTROMAGNETIC MEASUREMENTS DIGEST, 2000, : 96 - 97
  • [9] Calibration of the Second Generation of Ultrastable Low-Noise Current Amplifiers Using a Cryogenic Current Comparator
    Goetz, Martin
    Drung, Dietmar
    Krause, Christian
    Becker, Ulrich
    Scherer, Hansjorg
    2018 CONFERENCE ON PRECISION ELECTROMAGNETIC MEASUREMENTS (CPEM 2018), 2018,
  • [10] Calibration of ultrastable low-noise current amplifiers without direct use of a cryogenic current comparator
    Goetz, M.
    Drung, D.
    Krause, C.
    Mortara, A.
    METROLOGIA, 2020, 57 (05)