The effect of air within the infusion syringe on drug delivery of syringe pump infusion systems

被引:0
|
作者
Schulz, G [1 ]
Fischer, J [1 ]
Neff, T [1 ]
Bänziger, O [1 ]
Weiss, M [1 ]
机构
[1] Univ Zurich, Kinderklin, Abt Neonatol, CH-8032 Zurich, Switzerland
来源
ANAESTHESIST | 2000年 / 49卷 / 12期
关键词
syringe pumps; air; infusion line occlusion; drug delivery; complications;
D O I
10.1007/s001010070016
中图分类号
R614 [麻醉学];
学科分类号
100217 ;
摘要
Application of highly concentrated short-acting vasoactive drugs in the critically ill patient requires precisely working syringe pump systems for continuous intravenous drug delivery. We performed a bench study to investigate the consequences of small amounts of air entrapped within a 50-ml infusion syringe. In particular we studied the effect of entrapped air on drug delivery after moderate vertical displacement of the pump by 50 cm (e. g, in preparation for transport) and the effect on the time required to trigger the pressure alarm after occlusion of the infusion line. At a flow rate of 1 ml/h,lowering the syringe pump prolonged the zero-drug delivery time from (mean+/-SD) 4.1+/-0.8 min (without air) to 6.2+/-0.9 (with 1 mi air) and to 13.1+/-0.9 min (with 2 mi of air, p<0.001 for all comparisons). Entrapping of 2 mi of air within the syringe resulted in a 2.6-fold prolongation of the occlusion alarm time after accidental occlusion of the infusion line and a 3-fold increase of the resulting infusion bolus after occlusion. Enclosed air within infusion syringes considerably affects the syringe compliance. It increases the susceptibility of constant drug delivery to vertical displacement of syringe pumps and impairs the occlusion alarm function. Therefore, any air in syringe of infusion pump systems should be carefully removed. To avoid infusion boluses of short-acting vasoactive drugs after accidental occlusions, the occluded infusion line should be released to ambient pressure first.
引用
收藏
页码:1018 / 1023
页数:6
相关论文
共 50 条
  • [1] ALTERED DRUG INFUSION RATES DURING SYRINGE-PUMP INFUSION
    LEFF, RD
    HOLSTAD, SG
    ANN, C
    KUBAJAK, M
    ROBERTS, RJ
    AMERICAN JOURNAL OF HOSPITAL PHARMACY, 1986, 43 (09): : 2241 - 2242
  • [2] Air bubbles worsen flow irregularities in syringe pump infusion systems
    Neff, TA
    Schulz, G
    Fischer, J
    Baenziger, O
    Weiss, M
    ANESTHESIOLOGY, 2000, 93 (3A) : U165 - U165
  • [3] The effect of syringe size on the performance of infusion pump
    Kim, DW
    Steward, DJ
    PAEDIATRIC ANAESTHESIA, 1999, 9 (04): : 335 - 337
  • [4] FACTORS AFFECTING DRUG DELIVERY FROM A SYRINGE-PUMP INFUSION SET
    GEATER, REM
    LEFF, RD
    ROBERTS, RJ
    AMERICAN JOURNAL OF HOSPITAL PHARMACY, 1985, 42 (11): : 2510 - 2513
  • [5] An Intelligent Infusion Flow Controlled Syringe Infusion Pump
    Merhi, Nour
    Mohamad, Nour
    HajjMoussa, George
    ElSayed, Ahmad
    Bamashmos, Saeed H.
    Hamawy, Lara
    HajjHassan, Mohamad
    Abou Ali, Mohamad
    Kassem, Abdallah
    31ST INTERNATIONAL CONFERENCE ON MICROELECTRONICS (IEEE ICM 2019), 2019, : 48 - 52
  • [6] Potential hazard with syringe infusion pump
    Kannan, S
    ANAESTHESIA, 2001, 56 (09) : 913 - 914
  • [7] Experience in development of an infusion syringe pump
    Aronov A.M.
    Mel'nichenko Z.N.
    Stepkin A.V.
    Biomedical Engineering, 2007, 41 (4) : 203 - 204
  • [8] COST AND ACCEPTABILITY OF 3 SYRINGE-PUMP INFUSION SYSTEMS
    JOHNSON, MS
    PESKO, LJ
    WOOD, CF
    REINDERS, TP
    AMERICAN JOURNAL OF HOSPITAL PHARMACY, 1990, 47 (08): : 1794 - 1798
  • [9] Reliability of syringe infusion pump data channel
    Markevicius, Vytautas
    Navikas, Dangirutis
    PROCEEDINGS OF THE ITI 2008 30TH INTERNATIONAL CONFERENCE ON INFORMATION TECHNOLOGY INTERFACES, 2008, : 733 - 738
  • [10] Developing the control system of a syringe infusion pump
    Assuncao, R.
    Barbosa, P.
    Ruge, R.
    Guimaraes, P. S.
    Alves, J.
    Silva, I.
    Marques, M. A.
    2014 11TH INTERNATIONAL CONFERENCE ON REMOTE ENGINEERING AND VIRTUAL INSTRUMENTATION (REV), 2014, : 254 - 255