The effect of air within the infusion syringe on drug delivery of syringe pump infusion systems

被引:0
|
作者
Schulz, G [1 ]
Fischer, J [1 ]
Neff, T [1 ]
Bänziger, O [1 ]
Weiss, M [1 ]
机构
[1] Univ Zurich, Kinderklin, Abt Neonatol, CH-8032 Zurich, Switzerland
来源
ANAESTHESIST | 2000年 / 49卷 / 12期
关键词
syringe pumps; air; infusion line occlusion; drug delivery; complications;
D O I
10.1007/s001010070016
中图分类号
R614 [麻醉学];
学科分类号
100217 ;
摘要
Application of highly concentrated short-acting vasoactive drugs in the critically ill patient requires precisely working syringe pump systems for continuous intravenous drug delivery. We performed a bench study to investigate the consequences of small amounts of air entrapped within a 50-ml infusion syringe. In particular we studied the effect of entrapped air on drug delivery after moderate vertical displacement of the pump by 50 cm (e. g, in preparation for transport) and the effect on the time required to trigger the pressure alarm after occlusion of the infusion line. At a flow rate of 1 ml/h,lowering the syringe pump prolonged the zero-drug delivery time from (mean+/-SD) 4.1+/-0.8 min (without air) to 6.2+/-0.9 (with 1 mi air) and to 13.1+/-0.9 min (with 2 mi of air, p<0.001 for all comparisons). Entrapping of 2 mi of air within the syringe resulted in a 2.6-fold prolongation of the occlusion alarm time after accidental occlusion of the infusion line and a 3-fold increase of the resulting infusion bolus after occlusion. Enclosed air within infusion syringes considerably affects the syringe compliance. It increases the susceptibility of constant drug delivery to vertical displacement of syringe pumps and impairs the occlusion alarm function. Therefore, any air in syringe of infusion pump systems should be carefully removed. To avoid infusion boluses of short-acting vasoactive drugs after accidental occlusions, the occluded infusion line should be released to ambient pressure first.
引用
收藏
页码:1018 / 1023
页数:6
相关论文
共 50 条
  • [31] Performance of the atom 235 syringe infusion pump under hyperbaric conditions
    Story, DA
    Houston, JJ
    Millar, IL
    ANAESTHESIA AND INTENSIVE CARE, 1998, 26 (02) : 193 - 195
  • [32] SYRINGE DESIGN INFLUENCES INFUSION ERORS DUE TO CHANGES IN PUMP HEIGHT
    Murphy, Robert
    Feldman, Jeffrey
    ANESTHESIA AND ANALGESIA, 2012, 115
  • [33] In vitro delivery accuracy of Computer Controlled Infusion Pump software linked to an Alaris® GH syringe pump for propofol target-controlled infusion in dogs
    Cuniberti, Barbara
    Huuskonen, Vilhelmiina
    Hughes, J. M. Lynne
    MEDYCYNA WETERYNARYJNA-VETERINARY MEDICINE-SCIENCE AND PRACTICE, 2018, 74 (07): : 466 - 472
  • [34] Semi-Automated Self-Monitored Syringe Infusion Pump
    ElKheshen, Hasnaa
    Deni, Ibrahim
    Baalbaky, Alaa
    Dib, Mohamad
    Hamawy, Lara
    Abou Ali, Mohamad
    2018 INTERNATIONAL CONFERENCE ON COMPUTER AND APPLICATIONS (ICCA), 2018, : 331 - 335
  • [35] Flow irregularities from syringe infusion pumps caused by syringe stiction
    Peduzzi, Neris
    Batliner, Martin
    Grass, Beate
    Buehler, Philipp K.
    Daners, Marianne Schmid
    Meboldt, Mirko
    Weiss, Markus
    PEDIATRIC ANESTHESIA, 2020, 30 (08) : 885 - 891
  • [36] Evaluation of the FASTSTART mode for reducing start-up delay in syringe pump infusion systems
    Neff, T
    Fischer, J
    Fehr, S
    Baenziger, O
    Weiss, M
    SWISS MEDICAL WEEKLY, 2001, 131 (15-16) : 219 - 222
  • [37] A LOADING PORT FOR SYRINGE INFUSION PUMPS
    OFLYNN, RP
    SILER, JN
    ANESTHESIA AND ANALGESIA, 1991, 72 (03): : 407 - 408
  • [38] Effect of infusion line connection and infusion line characteristics on start-up fluid delivery from syringe infusion pumps used for microinfusion
    Weiss, Markus
    Wendel-Garcia, Pedro David
    Cannizzaro, Vincenzo
    Kleine-Brueggeney, Maren
    PEDIATRIC ANESTHESIA, 2025, 35 (02) : 118 - 124
  • [39] Evaluation of a novel flow-controlled syringe infusion pump for precise and continuous drug delivery at low flow rates: a laboratory study
    Batliner, M.
    Weiss, M.
    Dual, S. A.
    Grass, B.
    Meboldt, M.
    Daners, M. Schmid
    ANAESTHESIA, 2019, 74 (11) : 1425 - 1431
  • [40] USE OF A NEW SYRINGE PUMP (SPRINGFUSOR(R)) FOR MUSCLE-RELAXANT INFUSION
    EAGLE, CCP
    CAPES, DF
    ANAESTHESIA AND INTENSIVE CARE, 1993, 21 (04) : 444 - 446