The effect of air within the infusion syringe on drug delivery of syringe pump infusion systems

被引:0
|
作者
Schulz, G [1 ]
Fischer, J [1 ]
Neff, T [1 ]
Bänziger, O [1 ]
Weiss, M [1 ]
机构
[1] Univ Zurich, Kinderklin, Abt Neonatol, CH-8032 Zurich, Switzerland
来源
ANAESTHESIST | 2000年 / 49卷 / 12期
关键词
syringe pumps; air; infusion line occlusion; drug delivery; complications;
D O I
10.1007/s001010070016
中图分类号
R614 [麻醉学];
学科分类号
100217 ;
摘要
Application of highly concentrated short-acting vasoactive drugs in the critically ill patient requires precisely working syringe pump systems for continuous intravenous drug delivery. We performed a bench study to investigate the consequences of small amounts of air entrapped within a 50-ml infusion syringe. In particular we studied the effect of entrapped air on drug delivery after moderate vertical displacement of the pump by 50 cm (e. g, in preparation for transport) and the effect on the time required to trigger the pressure alarm after occlusion of the infusion line. At a flow rate of 1 ml/h,lowering the syringe pump prolonged the zero-drug delivery time from (mean+/-SD) 4.1+/-0.8 min (without air) to 6.2+/-0.9 (with 1 mi air) and to 13.1+/-0.9 min (with 2 mi of air, p<0.001 for all comparisons). Entrapping of 2 mi of air within the syringe resulted in a 2.6-fold prolongation of the occlusion alarm time after accidental occlusion of the infusion line and a 3-fold increase of the resulting infusion bolus after occlusion. Enclosed air within infusion syringes considerably affects the syringe compliance. It increases the susceptibility of constant drug delivery to vertical displacement of syringe pumps and impairs the occlusion alarm function. Therefore, any air in syringe of infusion pump systems should be carefully removed. To avoid infusion boluses of short-acting vasoactive drugs after accidental occlusions, the occluded infusion line should be released to ambient pressure first.
引用
收藏
页码:1018 / 1023
页数:6
相关论文
共 50 条
  • [41] Levels of Hemolysis Markers in Erythrocyte Concentrates Administered Using a Syringe Infusion Pump
    Gannam, Fernanda Figueiredo
    Belela-Anacleto, Aline Santa Cruz
    Kusahara, Denise Miyuki
    Pedreira, Mavilde Goncalves
    JOURNAL OF INFUSION NURSING, 2018, 41 (03) : 180 - 188
  • [42] Iloprost infusion by a new device as a portable syringe pump: safety, tolerability and agreement
    Faggioli, Paola
    Sciascera, Alba
    Giani, Leopoldo
    Mazzone, Antonino
    ITALIAN JOURNAL OF MEDICINE, 2012, 6 (04) : 311 - 314
  • [43] INCIDENCE OF INFUSION PHLEBITIS WITH A SYRINGE-PUMP SYSTEM VERSUS MINIBOTTLES - REPLY
    BAPTISTA, RJ
    DRISCOLL, DF
    CLINICAL PHARMACY, 1987, 6 (11): : 836 - 836
  • [44] An in vitro evaluation of infusion methods using a syringe pump to improve noradrenaline administration
    Genay, S.
    Decaudin, B.
    Scoccia, S.
    Barthelemy, C.
    Debaene, B.
    Lebuffe, G.
    Odou, P.
    ACTA ANAESTHESIOLOGICA SCANDINAVICA, 2015, 59 (02) : 197 - 204
  • [45] Lag times to steady state drug delivery by continuous intravenous infusion: direct comparison of peristaltic and syringe pump performance identifies contributions from infusion system dead volume and pump startup characteristics
    Lauren E. Gibson
    Anders S. Knudsen
    David Arney
    Hao Deng
    Nathaniel M. Sims
    Robert A. Peterfreund
    Journal of Clinical Monitoring and Computing, 2022, 36 : 1489 - 1498
  • [46] Lag times to steady state drug delivery by continuous intravenous infusion: direct comparison of peristaltic and syringe pump performance identifies contributions from infusion system dead volume and pump startup characteristics
    Gibson, Lauren E.
    Knudsen, Anders S.
    Arney, David
    Deng, Hao
    Sims, Nathaniel M.
    Peterfreund, Robert A.
    JOURNAL OF CLINICAL MONITORING AND COMPUTING, 2022, 36 (05) : 1489 - 1498
  • [47] Microparticle Drug Delivery Syringe
    Susan M, Hamilla
    Michael, Heusser
    Ashley, Murphy
    2011 IEEE 37TH ANNUAL NORTHEAST BIOENGINEERING CONFERENCE (NEBEC), 2011,
  • [48] Correction to: Lag times to steady state drug delivery by continuous intravenous infusion: direct comparison of peristaltic and syringe pump performance identifies contributions from infusion system dead volume and pump startup characteristics
    Lauren E. Gibson
    Anders S. Knudsen
    David Arney
    Hao Deng
    Nathaniel M. Sims
    Robert A. Peterfreund
    Journal of Clinical Monitoring and Computing, 2022, 36 : 1235 - 1236
  • [49] Preparing Drugs for Infusion Via Syringe Pump: A Key Step to Ensure Homogeneous Concentration
    Garrigue, Bruno
    Dehu, Yann
    Girault, Fabrice
    Figadere, Bruno
    Leblanc, Karine
    Briole, Nicolas
    Capitani, Georges Antoine
    Lagadec, Steven
    Laborne, Francois-Xavier
    CRITICAL CARE NURSE, 2016, 36 (04) : 36 - 44
  • [50] In-line filter included into the syringe infusion pump assembly reduces flow irregularities
    B. Brotschi
    B. Grass
    M. Weiss
    C. Doell
    V. Bernet
    Intensive Care Medicine, 2012, 38 : 518 - 522