An Intelligent Infusion Flow Controlled Syringe Infusion Pump

被引:0
|
作者
Merhi, Nour [1 ]
Mohamad, Nour [1 ]
HajjMoussa, George [1 ]
ElSayed, Ahmad [1 ]
Bamashmos, Saeed H. [1 ]
Hamawy, Lara [1 ]
HajjHassan, Mohamad [1 ]
Abou Ali, Mohamad [1 ]
Kassem, Abdallah [2 ]
机构
[1] Lebanese Int Univ, Dept Biomed Engn, Beirut, Lebanon
[2] Notre Dame Univ, Fac Engn, ECCE Dept, Zouk Mosbeh, Lebanon
关键词
Arduino Microcontroller; IEC; MEGA; 2560; Drug Library; Intelligent Flow Sensor; LD20; Flow rate; Syringe Pump;
D O I
10.1109/icm48031.2019.9021516
中图分类号
TM [电工技术]; TN [电子技术、通信技术];
学科分类号
0808 ; 0809 ;
摘要
Commercial intelligent syringe infusion pumps nowadays have many advanced features such as drug libraries, wireless communication, and a barcode scanner. However, such intelligent devices lack the capability of measuring an infusion rate (flow) directly. The rate is a variation of volume divided by a variation of time. Thus, commercial syringe pumps measure the diameter of disposable syringe through a linear potentiometer. Together with a distance information provided via a stepper motor's drive, a volume can be calculated. Lastly, a flow will be calculated by a division of this volume via spent time. In this paper, a direct measurement of an infused flow will be made via an LD20 (SF04) flow sensor manufactured by Sensiron company. This will improve the safety of a delivery of high risk medications through a direct flow measurement, an air-bubble detection, an occlusion detection, a temperature reading of infused drug and other features provided via the usage of LD20 flow sensor.
引用
收藏
页码:48 / 52
页数:5
相关论文
共 50 条
  • [1] The effect of air within the infusion syringe on drug delivery of syringe pump infusion systems
    Schulz, G
    Fischer, J
    Neff, T
    Bänziger, O
    Weiss, M
    ANAESTHESIST, 2000, 49 (12): : 1018 - 1023
  • [2] Potential hazard with syringe infusion pump
    Kannan, S
    ANAESTHESIA, 2001, 56 (09) : 913 - 914
  • [3] Experience in development of an infusion syringe pump
    Aronov A.M.
    Mel'nichenko Z.N.
    Stepkin A.V.
    Biomedical Engineering, 2007, 41 (4) : 203 - 204
  • [4] Air bubbles worsen flow irregularities in syringe pump infusion systems
    Neff, TA
    Schulz, G
    Fischer, J
    Baenziger, O
    Weiss, M
    ANESTHESIOLOGY, 2000, 93 (3A) : U165 - U165
  • [5] ALTERED DRUG INFUSION RATES DURING SYRINGE-PUMP INFUSION
    LEFF, RD
    HOLSTAD, SG
    ANN, C
    KUBAJAK, M
    ROBERTS, RJ
    AMERICAN JOURNAL OF HOSPITAL PHARMACY, 1986, 43 (09): : 2241 - 2242
  • [6] Effects of infusion tubing line lengths and syringe sizes on infusion system compliance: an experimental study using a syringe-type infusion pump at low flow rate
    Hong, Kwan Young
    Lee, Eun Kyung
    Kim, Youyoung
    Choi, Dan-Cheong
    Min, Jeong-Jin
    JOURNAL OF CLINICAL MONITORING AND COMPUTING, 2023, 37 (05) : 1379 - 1386
  • [7] Effects of infusion tubing line lengths and syringe sizes on infusion system compliance: an experimental study using a syringe-type infusion pump at low flow rate
    Kwan Young Hong
    Eun Kyung Lee
    Youyoung Kim
    Dan-Cheong Choi
    Jeong-Jin Min
    Journal of Clinical Monitoring and Computing, 2023, 37 : 1379 - 1386
  • [8] Developing the control system of a syringe infusion pump
    Assuncao, R.
    Barbosa, P.
    Ruge, R.
    Guimaraes, P. S.
    Alves, J.
    Silva, I.
    Marques, M. A.
    2014 11TH INTERNATIONAL CONFERENCE ON REMOTE ENGINEERING AND VIRTUAL INSTRUMENTATION (REV), 2014, : 254 - 255
  • [9] Reliability of syringe infusion pump data channel
    Markevicius, Vytautas
    Navikas, Dangirutis
    PROCEEDINGS OF THE ITI 2008 30TH INTERNATIONAL CONFERENCE ON INFORMATION TECHNOLOGY INTERFACES, 2008, : 733 - 738
  • [10] Inadvertent subcutaneous infusion of propofol by a syringe pump
    Rumiko Uda
    Journal of Anesthesia, 2000, 14 (3) : 169 - 170