X-ray micro-CT scanner for small animal imaging based on Timepix detector technology

被引:16
|
作者
Dudak, Jan [1 ,2 ]
Zemlicka, Jan [1 ]
Krejci, Frantisek [1 ]
Polansky, Stepan [1 ]
Jakubek, Jan [1 ]
Mrzilkova, Jana [3 ]
Patzelt, Matej [3 ]
Trnka, Jan [3 ]
机构
[1] Czech Tech Univ, Inst Expt & Appl Phys, Prague 12800 2, Czech Republic
[2] Czech Tech Univ, Fac Biomed Engn, Kladno 27200, Czech Republic
[3] Charles Univ Prague, Fac Med 3, Prague 10000, Czech Republic
关键词
Computerized tomography; Hybrid detector; Timepix; Small animal imaging; COMPUTED TOMOGRAPHY; MICROTOMOGRAPHY; RESOLUTION;
D O I
10.1016/j.nima.2014.10.076
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
We describe a newly developed compact micro-CT scanner with rotating gantry equipped with a Timepix Quad hybrid pixel semiconductor detector and a micro-focus X-ray tube providing spatial resolution down to 30 mu m. The resolving power of the device in relation to soft tissue sensitivity is demonstrated using a tissue-equivalent phantom and different types of biological samples. The results demonstrate that the use of noiseless particle counting detectors is a promising way to achieve sufficient soft tissue contrast even without any contrast agents. (C) 2014 Elsevier B.V. All rights reserved.
引用
收藏
页码:81 / 86
页数:6
相关论文
共 50 条
  • [21] A flat-panel detector based micro-CT system: performance evaluation for small-animal imaging
    Lee, SC
    Kim, HK
    Chun, IK
    Cho, MH
    Lee, SY
    Cho, MH
    PHYSICS IN MEDICINE AND BIOLOGY, 2003, 48 (24): : 4173 - 4185
  • [22] Effect of experimental parameters on X-ray Micro-CT imaging for Spectral CT applications
    Lima, Mateus Gruener
    Jussiani, Eduardo Inocente
    RADIATION PHYSICS AND CHEMISTRY, 2025, 226
  • [23] Investigations on x-ray luminescence CT for small animal imaging
    Badea, C. T.
    Stanton, I. N.
    Johnston, S. M.
    Johnson, G. A.
    Therien, M. J.
    MEDICAL IMAGING 2012: PHYSICS OF MEDICAL IMAGING, 2012, 8313
  • [24] Ultrafast laser-based micro-CT system for small animal imaging
    Krol, A
    Kieffer, JC
    Nees, J
    Chen, L
    Toth, R
    Hou, B
    Kincaid, RE
    Coman, IL
    Lipson, ED
    Mourou, G
    MEDICAL IMAGING 2004: PHYSICS OF MEDICAL IMAGING, PTS 1 AND 2, 2004, 5368 : 265 - 271
  • [25] X-ray strain tensor imaging: FEM simulation and experiments with a micro-CT
    Kim, Jae G.
    Park, So E.
    Lee, Soo Y.
    JOURNAL OF X-RAY SCIENCE AND TECHNOLOGY, 2014, 22 (01) : 63 - 75
  • [26] A perfusion procedure for imaging of the mouse cerebral vasculature by X-ray micro-CT
    Ghanavati, Sahar
    Yu, Lisa X.
    Lerch, Jason P.
    Sled, John G.
    JOURNAL OF NEUROSCIENCE METHODS, 2014, 221 : 70 - 77
  • [27] A realistic projection simulator for laboratory based X-ray micro-CT
    Dhaene, Jelle
    Pauwels, Elin
    De Schryver, Thomas
    De Muynck, Amelie
    Dierick, Manuel
    Van Hoorebeke, Luc
    NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION B-BEAM INTERACTIONS WITH MATERIALS AND ATOMS, 2015, 342 : 170 - 178
  • [28] Design, optimization and testing of a multi-beam micro-CT scanner based on multi-beam field emission x-ray technology
    Peng, R.
    Zhang, J.
    Calderon-Colon, X.
    Wang, S.
    Sultana, S.
    Wang, P.
    Yang, G.
    Chang, S.
    Lu, J.
    Zhou, O.
    MEDICAL IMAGING 2010: PHYSICS OF MEDICAL IMAGING, 2010, 7622
  • [29] Assessment of x-ray Scatter for the Micro-CT Subsystem of the FLEX Triumph™ Preclinical PET-CT Scanner
    Gutierrez, Daniel
    Zaidi, Habib
    2010 IEEE NUCLEAR SCIENCE SYMPOSIUM CONFERENCE RECORD (NSS/MIC), 2010, : 2216 - 2223
  • [30] An experimental micro-CT system for X-ray NDT
    Rossi, M
    Casali, F
    Bettuzzi, M
    Morigi, MP
    Romani, A
    Golovkin, SV
    Govorun, VN
    DEVELOPMENTS IN X-RAY TOMOGRAPHY III, 2002, 4503 : 338 - 348