Technical Note: Homogeneity of Gafchromic® EBT2 film

被引:105
|
作者
Hartmann, Bernadette [1 ,2 ]
Martisikova, Maria [1 ]
Jaekel, Oliver [1 ,3 ]
机构
[1] German Canc Res Ctr, Dept Med Phys Radiat Oncol, D-69120 Heidelberg, Germany
[2] Univ Halle Wittenberg, Inst Phys, D-06120 Halle, Germany
[3] Univ Klinikum Heidelberg, Heidelberger Lonenstrahl Therapiezentrum HIT, D-69120 Heidelberg, Germany
关键词
dosimeters; dosimetry; radiation therapy; FLAT-BED SCANNER; RADIOCHROMIC FILM; DOSIMETRY;
D O I
10.1118/1.3368601
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Methods: A scan protocol established for EBT film and published previously was used. The uniformity of the film coloration was investigated for unirradiated and irradiated EBT2 film sheets. The dose response of EBT2 film was measured and the influence of film inhomogeneities on dose determination was evaluated. Results: Inhomogeneities in pixel values of up to +/- 3.7% within one film were detected. The relative inhomogeneities were found to be approximately independent of the dose. Nonuniformities of the film response lead to uncertainties in dose determination of +/- 8.7% at 1 Gy. When using net optical densities for dose calibration, uncertainties in dose determination amount to more than +/- 6%. Conclusions: EBT2 films from the lot investigated in this study show response inhomogeneities, which lead to uncertainties in dose determination exceeding the commonly accepted tolerance levels. It is important to test further EBT2 lots regarding homogeneity before using the film in clinical routine.
引用
收藏
页码:1753 / 1756
页数:4
相关论文
共 50 条
  • [21] Evaluation of a Gafchromic EBT2 film dosimetry system for radiotherapy quality assurance
    T. Aland
    T. Kairn
    J. Kenny
    Australasian Physical & Engineering Sciences in Medicine, 2011, 34 : 251 - 260
  • [22] Correction of nonuniformity error of Gafchromic EBT2 and EBT3
    Katsuda, Toshizo
    Gotanda, Rumi
    Gotanda, Tatsuhiro
    Akagawa, Takuya
    Tanki, Nobuyoshi
    Kuwano, Tadao
    Yabunaka, Kouichi
    JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2016, 17 (03): : 41 - 51
  • [23] Technical Note: Initial characterization of the new EBT-XD Gafchromic film
    Grams, Michael P.
    Gustafson, Jon M.
    Long, Kenneth M.
    de los Santos, Luis E. Fong
    MEDICAL PHYSICS, 2015, 42 (10) : 5782 - 5786
  • [24] Technical Note: On GAFChromic EBT-XD film and the lateral response artifact
    Lewis, David F.
    Chan, Maria F.
    MEDICAL PHYSICS, 2016, 43 (02) : 643 - 649
  • [25] Gafchromic EBT2 dosimetry via robust optimization
    Alves, Victor G. L.
    Cardoso, Simone C.
    da Silva, Ademir X.
    COMPUTER PHYSICS COMMUNICATIONS, 2013, 184 (07) : 1708 - 1716
  • [26] Variationsin Dose Response of Gafchromic EBT2 Film Scanned in Transmission and Reflection Mode
    Park, S.
    Kang, S.
    Cheong, K.
    Hwang, T.
    Lee, M.
    Kim, K.
    Oh, D.
    Bae, H.
    MEDICAL PHYSICS, 2011, 38 (06) : 3524 - +
  • [27] Submillimeter Dose Measurement with Gafchromic EBT2 and EBT3 Films
    Corradini, N.
    Presilla, S.
    MEDICAL PHYSICS, 2013, 40 (06)
  • [28] High dose per fraction dosimetry of small fields with Gafchromic EBT2 film
    Hardcastle, Nicholas
    Basavatia, Amar
    Bayliss, Adam
    Tome, Wolfgang A.
    MEDICAL PHYSICS, 2011, 38 (07) : 4081 - 4085
  • [29] Response of the modified GAFCHROMIC EBT2 radiochromic film to DC glow discharge plasma
    Farag, Omar F.
    Kotb, Omar M.
    El Ghazaly, M.
    EL-Sayed, Naglaa M.
    SCIENTIFIC REPORTS, 2024, 14 (01)
  • [30] Dosimetric evaluation of Gafchromic EBT2 film for breast intraoperative electron radiotherapy verification
    Baghani, Hamid Reza
    Aghamiri, S. Mahmoud Reza
    Mahdavi, S. Rabi
    Robatjazi, Mostafa
    Zadeh, Arezo Rahim
    Akbari, Mohammad Esmail
    Mirzaei, Hamid Reza
    Nafissi, Nahid
    Samsami, Majid
    PHYSICA MEDICA-EUROPEAN JOURNAL OF MEDICAL PHYSICS, 2015, 31 (01): : 37 - 42