Liquid ionization chamber measurements of dose distributions in small 6 MV photon beams

被引:19
|
作者
Dasu, A
Lofroth, PO
Wickman, G
机构
[1] Umea Univ, Dept Radiat Phys, S-90185 Umea, Sweden
[2] Alexandru Ioan Cuza Univ, Dept Med Phys & Biophys, R-6600 Iasi 7, Romania
来源
PHYSICS IN MEDICINE AND BIOLOGY | 1998年 / 43卷 / 01期
关键词
D O I
10.1088/0031-9155/43/1/002
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
A new liquid ionization chamber (LIC) design optimized for high spatial resolution was used for measurements of dose distributions in radiation fields intended for stereotactic radiosurgery (SRS). This work was mainly focused on the properties of this detector in radiation fields from linear accelerators for clinical radiotherapy (pulsed radiation with dose rates from approximately 0.5 to 5 Gy min(-1) and beam diameters down to 8 mm). The narrow beams used in stereotactic radiosurgery require detectors with small sizes in order to provide a good spatial resolution. The LIC is investigated to see whether it can be used as a detector for dose measurements in beams currently used for stereotactic radiosurgery. Its properties are compared with those of silicon diodes. The comparisons include output factor (OF), depth dose and profile measurements in 6 MV photon fields of different sizes. For OF measurements, an NACP air ionization chamber was also used in the comparison. The dependence of the response on the detector orientation in the photon beam is also investigated for the diodes and the LIC. The results suggest that LICs can provide better properties than diodes for measuring dose distributions in narrow photon beams.
引用
收藏
页码:21 / 36
页数:16
相关论文
共 50 条
  • [1] MEASUREMENTS OF DOSE DISTRIBUTIONS IN SMALL BEAMS OF 6 MV X-RAYS
    RICE, RK
    HANSEN, JL
    SVENSSON, GK
    SIDDON, RL
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 1987, 32 (09): : 1087 - 1099
  • [2] Cylindrical ionization chamber response in static and dynamic 6 and 15 MV photon beams
    Andersson, P.
    Swanpalmer, J.
    Palm, A.
    Bath, M.
    Chakarova, R.
    [J]. BIOMEDICAL PHYSICS & ENGINEERING EXPRESS, 2023, 9 (02)
  • [3] LIQUID IONIZATION-CHAMBER FOR ABSORBED DOSE DETERMINATIONS IN PHOTON AND ELECTRON-BEAMS
    JOHANSSON, KA
    SVENSSON, H
    [J]. ACTA RADIOLOGICA ONCOLOGY, 1982, 21 (05): : 359 - 367
  • [4] Surface dose measurements with GafChromic EBT film for 6 and 18 MV photon beams
    Bilge, Hatice
    Cakir, Aydin
    Okutan, Murat
    Acar, Hilal
    [J]. PHYSICA MEDICA, 2009, 25 (02) : 101 - 104
  • [5] Comparison of MCNP5 Calculations in the Buildup Region with Plane Parallel Ionization Chamber Measurements for 6 and 18 MV Photon Beams
    Abbas, H.
    Satti, J.
    MacDonald, C.
    [J]. MEDICAL PHYSICS, 2012, 39 (06) : 3706 - 3706
  • [6] Volume dose ratios relevant for alanine dosimetry in small, 6 MV photon beams
    Cronholm, Rickard O.
    Andersen, Claus E.
    Behrens, Claus F.
    Helt-Hansen, Jakob
    [J]. RADIATION MEASUREMENTS, 2012, 47 (10) : 1014 - 1017
  • [7] Peripheral dose measurements for 6 and 18 MV photon beams on a linear accelerator with multileaf collimator
    Mazonakis, Michalis
    Zacharopoulou, Fotini
    Varveris, Haralambos
    Damilakis, John
    [J]. MEDICAL PHYSICS, 2008, 35 (10) : 4396 - 4403
  • [8] IS REFERENCE CHAMBER CALIBRATION IN MV PHOTON BEAMS NECESARRY FOR ABSOLUTE DOSE DETERMINATION?
    Vestergaard, A.
    Ronde, H. S.
    Berg, M.
    Praestegaard, L. H.
    [J]. RADIOTHERAPY AND ONCOLOGY, 2011, 99 : S412 - S412
  • [9] Comparison of a Liquid Ionization Chamber with a Diamond Detector for Relatively Small Photon Field Measurements
    Yip, C.
    Cheng, H.
    Cheung, F.
    Chan, C.
    [J]. MEDICAL PHYSICS, 2011, 38 (06)
  • [10] Evaluation of a liquid ionization chamber for relative dosimetry in small and large fields of radiotherapy photon beams
    Benitez, E. M.
    Casado, F. J.
    Garcia-Pareja, S.
    Martin-Viera, J. A.
    Moreno, C.
    Parra, V.
    [J]. RADIATION MEASUREMENTS, 2013, 58 : 79 - 86