Cylindrical ionization chamber response in static and dynamic 6 and 15 MV photon beams

被引:0
|
作者
Andersson, P. [1 ,3 ]
Swanpalmer, J. [1 ,2 ]
Palm, A. [2 ]
Bath, M. [1 ,2 ]
Chakarova, R. [1 ,2 ]
机构
[1] Univ Gothenburg, Inst Clin Sci, Sahlgrenska Acad, Dept Med Radiat Sci, Gothenburg, Sweden
[2] Sahlgrens Univ Hosp, Dept Med Phys & Biomed Engn, Gothenburg, Sweden
[3] RISE Res Inst Sweden, Mat & Prod, Gothenburg, Sweden
来源
关键词
Ionization chamber dosimetry; non-reference conditions; Monte Carlo simulations; radiotherapy; VMAT; REFERENCE DOSIMETRY; LINEAR ACCELERATORS; MONTE-CARLO;
D O I
10.1088/2057-1976/acb553
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose. To investigate the response of the CC13 ionization chamber under non-reference photon beam conditions, focusing on penumbra and build-up regions of static fields and on dynamic intensity-modulated beams. Methods. Measurements were performed in 6 MV 100 x 100, 20 x 100, and 20 x 20 mm(2) static fields. Monte Carlo calculations were performed for the static fields and for 6 and 15 MV dynamic beam sequences using a Varian multi-leaf collimator. The chamber was modelled using EGSnrc egs_chamber software. Conversion factors were calculated by relating the absorbed dose to air in the chamber air cavity to the absorbed dose to water. Correction and point-dose correction factors were calculated to quantify the conversion factor variations. Results. The correction factors for positions on the beam central axis and at the penumbra centre were 0.98-1.02 for all static fields and depths investigated. The largest corrections were obtained for chamber positions beyond penumbra centre in the off-axis direction. Point-dose correction factors were 0.54-0.71 at 100 mm depth and their magnitude increased with decreasing field size and measurement depth. Factors of 0.99-1.03 were obtained inside and near the integrated penumbra of the dynamic field at 100 mm depth, and of 0.92-0.94 beyond the integrated penumbra centre. The variations in the ionization chamber response across the integrated dynamic penumbra qualitatively followed the behaviour across penumbra of static fields. Conclusions. Without corrections, the CC13 chamber was of limited usefulness for profile measurements in 20-mm-wide fields. However, measurements in dynamic small irregular beam openings resembling the conditions of pre-treatment patient quality assurance were feasible. Uncorrected ionization chamber response could be applied for dose verification at 100 mm depth inside and close to large gradients of dynamically accumulating high- and low-dose regions assuming 3% tolerance between measured and calculated doses.
引用
收藏
页数:9
相关论文
共 50 条
  • [1] Liquid ionization chamber measurements of dose distributions in small 6 MV photon beams
    Dasu, A
    Lofroth, PO
    Wickman, G
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 1998, 43 (01): : 21 - 36
  • [2] Dosimetrical properties of the new IBA RAZOR NanoChamber ionization chamber in MV photon beams
    Casar, B.
    Gershkevitsh, E.
    Mendez, I.
    [J]. RADIOTHERAPY AND ONCOLOGY, 2018, 127 : S919 - S919
  • [3] Ionization chamber dosimetry in non-reference conditions for 15 MV narrow photon fields
    Montaño, EG
    Sánchez-Doblado, F
    Capote, R
    Hermoso, E
    Arrans, R
    Andreo, P
    [J]. RADIOTHERAPY AND ONCOLOGY, 2005, 76 : S226 - S226
  • [4] Feasibility of linear diode array based small field data acquisition for 6 MV & 15 MV photon beams - An intercomparison with micro ion chamber
    Gul, Attia
    Kakakhel, M. Basim
    Amjad, Nauman
    Razzaq, Aasia
    Mirza, Sikander M.
    [J]. RADIATION PHYSICS AND CHEMISTRY, 2019, 162 : 146 - 152
  • [5] Comparison of MCNP5 Calculations in the Buildup Region with Plane Parallel Ionization Chamber Measurements for 6 and 18 MV Photon Beams
    Abbas, H.
    Satti, J.
    MacDonald, C.
    [J]. MEDICAL PHYSICS, 2012, 39 (06) : 3706 - 3706
  • [6] Study of the ionization chamber response to flattening-filter-free photon beams
    Bresolin, A.
    Bonfantini, F.
    Stucchi, C. G.
    Mongioj, V.
    Carrara, M.
    Pignoli, E.
    [J]. RADIATION MEASUREMENTS, 2017, 97 : 47 - 53
  • [7] Spectral differences in 6 MV beams with matched PDDs and the effect on chamber response
    Lye, J. E.
    Butler, D. J.
    Ramanathan, G.
    Franich, R. D.
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 2012, 57 (22): : 7599 - 7614
  • [8] Recombination factors for the cylindrical FC65-G ionization chamber in pulsed photon beams and the plane-parallel Roos ionization chamber in pulsed electron beams
    Berg, M
    Noerrevang, O
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 2004, 49 (23): : 5309 - 5318
  • [9] Monte Carlo study of conversion factors for ionization chamber dosimetry in solid slab phantoms for MV photon beams
    Park, Dong-wook
    Lee, Jai-ki
    [J]. JOURNAL OF THE KOREAN PHYSICAL SOCIETY, 2016, 69 (04) : 647 - 652
  • [10] Monte Carlo study of conversion factors for ionization chamber dosimetry in solid slab phantoms for MV photon beams
    Dong-wook Park
    Jai-ki Lee
    [J]. Journal of the Korean Physical Society, 2016, 69 : 647 - 652