Effective dose to staff from interventional procedures: estimations from single and double dosimetry

被引:14
|
作者
Kuipers, Gerritjan [1 ]
Velders, Xandra L. [1 ]
机构
[1] Acad Med Ctr, NL-1100 DD Amsterdam, Netherlands
关键词
CARDIAC-CATHETERIZATION; FOLLOW-UP; RADIOLOGY;
D O I
10.1093/rpd/ncp155
中图分类号
X [环境科学、安全科学];
学科分类号
08 ; 0830 ;
摘要
The exposure of 11 physicians performing interventional procedures was measured by means of two personal dosemeters. One personal dosemeter was worn outside the lead apron and an additional under the lead apron. The study was set up in order to determine the added value of a dosemeter worn under the lead apron. With the doses measured, the effective doses of the physicians were estimated using an algorithm for single dosimetry and two algorithms for double dosimetry. The effective doses calculated with the single dosimetry algorithm ranged from 0.11 to 0.85 mSv in 4 weeks. With the double dosimetry algorithms, the effective doses ranged from 0.02 mSv to 0.47 mSv. The statistical analysis revealed no significant differences in the accuracy of the effective doses calculated with single or double dosimetry algorithms. It was concluded that the effective dose cannot be considered a more accurate estimate when two dosemeters are used instead of one.
引用
收藏
页码:95 / 100
页数:6
相关论文
共 50 条
  • [41] Estimating the whole-body effective dose and health risks as well as introducing a new easy method for eye lens dosimetry in interventional cardiology procedures
    Hatami, Alireza
    Bagheri, Mahmoud
    Falahati, Farzaneh
    Banaei, Amin
    Abedi-Firouzjah, Razzagh
    Zamani, Hamed
    Kiapour, Mohammad
    Momeni, Farideh
    METHODSX, 2020, 7
  • [42] Relationships between entrance skin dose, effective dose and dose area product for patients in diagnostic and interventional cardiac procedures
    Hansson, B
    Karambatsakidou, A
    RADIATION PROTECTION DOSIMETRY, 2000, 90 (1-2) : 141 - 144
  • [43] EXPERIMENTAL VALIDATION OF ALGORITHMS USED TO ESTIMATE EFFECTIVE DOSE DURING INTERVENTIONAL RADIOLOGY PROCEDURES
    Negri, P.
    Campi, F.
    De Crescenzo, S.
    Garlati, L.
    Tambussi, O.
    RADIATION PROTECTION DOSIMETRY, 2019, 187 (01) : 42 - 49
  • [44] Entrance skin dose estimates derived from dose-area product measurements in interventional radiological procedures
    McParland, BJ
    BRITISH JOURNAL OF RADIOLOGY, 1998, 71 (852): : 1288 - 1295
  • [45] Using a convolutional neural network for human recognition in a staff dose management software for fluoroscopic interventional procedures
    Troville, J.
    Dhonde, R. S.
    Rudin, S.
    Bednarek, D. R.
    MEDICAL IMAGING 2021: PHYSICS OF MEDICAL IMAGING, 2021, 11595
  • [46] SCATTER DOSE VALUES IN LOWER EXTREMITIES FOR STAFF DURING PAEDIATRIC INTERVENTIONAL CARDIOLOGY PROCEDURES: AN EXPERIMENTAL APPROACH
    Ubeda, Carlos
    Miranda, Patricia
    Aguirre, Daniel
    Riquelme, Nemorino
    Dalmazzo, Dandaro
    Vega, Juan
    RADIATION PROTECTION DOSIMETRY, 2017, 174 (01) : 88 - 93
  • [47] Radiation-Induced Noncancer Risks in Interventional Cardiology: Optimisation of Procedures and Staff and Patient Dose Reduction
    Sun, Zhonghua
    AbAziz, Aini
    Yusof, Ahmad Khairuddin Md
    BIOMED RESEARCH INTERNATIONAL, 2013, 2013
  • [48] Skin dose to patients from interventional radiology and cardiology procedures with potentially long fluoroscopy times
    Mooney, RB
    RADIATION PROTECTION DOSIMETRY, 2000, 90 (1-2) : 123 - 126
  • [49] Patient dose during interventional neuroradiology procedures: Results from a multi-center study
    Kien, N.
    Rehel, J. -L.
    Etard, C.
    Aubert, B.
    JOURNAL DE RADIOLOGIE, 2011, 92 (12): : 1101 - 1112
  • [50] A Significant Reduction in the Frequency of Substantial-Dose Interventional Coronary Procedures from 2007 to 2017
    Balter, S.
    Myoung, Sarah
    Doshi, Darshan
    Karmpaliotis, Dimitrios
    Kirtane, Ajay Jayant
    Parikh, Manish
    Grossberg, Yonina
    Liao, Ming
    Moses, Jeffrey W.
    JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 2018, 72 (13) : B102 - B102