Evaluation of laboratory scanner accuracy by a novel calibration block for complete-arch implant rehabilitation

被引:14
|
作者
Pan, Yu [1 ]
Tam, Juliana M. Y. [2 ]
Tsoi, James K. H. [1 ]
Lam, Walter Y. H. [1 ]
Huang, Ruoxuan [3 ]
Chen, Zhuofan [3 ]
Pow, Edmond H. N. [1 ]
机构
[1] Univ Hong Kong, Fac Dent, Hong Kong, Peoples R China
[2] Hong Kong Polytech Univ, Ind Ctr, Hong Kong, Peoples R China
[3] Sun Yat Sen Univ, Guangdong Prov Key Lab Stornatol, Guanghua Sch Stomatol, Hosp Stomatol, Guangzhou, Peoples R China
关键词
Accuracy; Complete-arch; Laboratory scanner; Calibration block; Tilted implant; Inter-implant distance; EXTRAORAL DENTAL SCANNERS; IN-VITRO; 3-DIMENSIONAL ACCURACY; IMPRESSIONS; PRECISION;
D O I
10.1016/j.jdent.2020.103476
中图分类号
R78 [口腔科学];
学科分类号
1003 ;
摘要
Objectives: To compare the accuracies of 4 laboratory scanners using a new custom-made block for complete-arch implant rehabilitation. Materials and Methods: A block comprised 4 cylinders, with 2 in the anterior (0 degrees angulation) and 2 in the posterior region (45 degrees distal angulation) (Experimental group) and a standard block with 2 parallel cones in 16 degrees taper (ISO group), were fabricated. Both blocks were scanned consecutively for 15 times by 4 laboratory scanners: IScan, Zfx, 3Shape, and KaVo. Measurements were also made by a coordinate measuring machine (CMM) as the reference. Acquired digital models were inspected with a metrology software. Linear and angular distortions were computed evaluation of trueness, precision and expanded uncertainty of scanners. Effects of blocks and scanners on the scanning trueness and precision were analyzed by Two-way ANOVA (alpha = 0.05). Results: The linear trueness and precision of scanners was significantly poorer when the Experimental block instead of ISO block was used. Significant greater distortions were noticed at the 45 degrees sites than the 0 degrees sites and no significant effect of inter-implant distance on the scanning accuracy was found. Zfx, IScan and 3Shape exhibited comparable expanded uncertainties (10.6 similar to 11.8) but KaVo showed the greatest (19.3) in complete-arch implant scans. Conclusion: The ISO block might not be more suitable than the Experimental block for evaluating the accuracy of laboratory scanner for complete-arch implant scanning. All the scanners tested except one demonstrated acceptable accuracy for complete-arch implant digitization. Scanning accuracy was compromised by unfavourable angulation of implants. Clinical significance: It is questionable whether the laboratory scanners validated for tooth-borne cases could also perform well for implant cases. It might be necessary to develop a new calibration object specifically for complete-arch implant scenarios to verify the capability of laboratory scanners in the implant workflow.
引用
收藏
页数:9
相关论文
共 50 条
  • [1] The effect of scanner type and scan body position on the accuracy of complete-arch digital implant scans
    Cakmak, Gulce
    Yilmaz, Hakan
    Trevino, Alejandro
    Kokat, Ali Murat
    Yilmaz, Burak
    CLINICAL IMPLANT DENTISTRY AND RELATED RESEARCH, 2020, 22 (04) : 533 - 541
  • [2] Comparative Analysis of Intraoral Scanner Accuracy in a Six-Implant Complete-Arch Model: An In Vitro Study
    Ferrini, Francesco
    Mazzoleni, Federica
    Barbini, Matteo
    Coppo, Carlotta
    Di Domenico, Giovanna Laura
    Gherlone, Enrico Felice
    PROSTHESIS, 2024, 6 (02): : 401 - 412
  • [3] Accuracy of complete-arch implant impression made with occlusal registration material
    Papazoglou, Efstratios
    Wee, Alvin G.
    Carr, Alan B.
    Urban, Istvan
    Margaritis, Vasileios
    JOURNAL OF PROSTHETIC DENTISTRY, 2020, 123 (01): : 143 - 148
  • [4] Complete digital workflow in prosthesis prototype fabrication for complete-arch implant rehabilitation: A technique
    Papaspyridakos, Panos
    Chen, Yo-wei
    Gonzalez-Gusmao, Ilenia
    Att, Wael
    JOURNAL OF PROSTHETIC DENTISTRY, 2019, 122 (03): : 189 - 192
  • [5] Accuracy of complete-arch model using an intraoral video scanner: An in vitro study
    Jeong, Ii-Do
    Lee, Jae-Jun
    Jeon, Jin-Hun
    Kim, Ji-Hwan
    Kim, Hae-Young
    Kim, Woong-Chul
    JOURNAL OF PROSTHETIC DENTISTRY, 2016, 115 (06): : 755 - 759
  • [6] Accuracy of photogrammetry, intraoral scanning, and conventional impression techniques for complete-arch implant rehabilitation: an in vitro comparative study
    Bowen Ma
    Xinxin Yue
    Yujie Sun
    Lingyan Peng
    Wei Geng
    BMC Oral Health, 21
  • [7] Effect of Scanner Type and Scan Body Location on the Accuracy of Mandibular Complete-Arch Digital Implant Scans: An In Vitro Study
    Cakmak, Gulce
    Yilmaz, Hakan
    Santos, Alejandro Trevino
    Kokat, Ali Murat
    Yilmaz, Burak
    JOURNAL OF PROSTHODONTICS-IMPLANT ESTHETIC AND RECONSTRUCTIVE DENTISTRY, 2022, 31 (05): : 419 - 426
  • [8] Accuracy of photogrammetry, intraoral scanning, and conventional impression techniques for complete-arch implant rehabilitation: an in vitro comparative study
    Ma, Bowen
    Yue, Xinxin
    Sun, Yujie
    Peng, Lingyan
    Geng, Wei
    BMC ORAL HEALTH, 2021, 21 (01)
  • [9] Complete-arch accuracy of intraoral scanners
    Treesh, Joshua C.
    Liacouras, Peter C.
    Taft, Robert M.
    Brooks, Daniel, I
    Raiciulescu, Sorana
    Ellert, Daniel O.
    Grant, Gerald T.
    Ye, Ling
    JOURNAL OF PROSTHETIC DENTISTRY, 2018, 120 (03): : 382 - 388
  • [10] Accuracy of a new elastomeric impression material for complete-arch dental implant impressions
    Baig, Mirza R.
    Buzayan, Muaiyed M.
    Yunus, Norsiah
    JOURNAL OF INVESTIGATIVE AND CLINICAL DENTISTRY, 2018, 9 (02) : e12320