Evaluation of laboratory scanner accuracy by a novel calibration block for complete-arch implant rehabilitation

被引:14
|
作者
Pan, Yu [1 ]
Tam, Juliana M. Y. [2 ]
Tsoi, James K. H. [1 ]
Lam, Walter Y. H. [1 ]
Huang, Ruoxuan [3 ]
Chen, Zhuofan [3 ]
Pow, Edmond H. N. [1 ]
机构
[1] Univ Hong Kong, Fac Dent, Hong Kong, Peoples R China
[2] Hong Kong Polytech Univ, Ind Ctr, Hong Kong, Peoples R China
[3] Sun Yat Sen Univ, Guangdong Prov Key Lab Stornatol, Guanghua Sch Stomatol, Hosp Stomatol, Guangzhou, Peoples R China
关键词
Accuracy; Complete-arch; Laboratory scanner; Calibration block; Tilted implant; Inter-implant distance; EXTRAORAL DENTAL SCANNERS; IN-VITRO; 3-DIMENSIONAL ACCURACY; IMPRESSIONS; PRECISION;
D O I
10.1016/j.jdent.2020.103476
中图分类号
R78 [口腔科学];
学科分类号
1003 ;
摘要
Objectives: To compare the accuracies of 4 laboratory scanners using a new custom-made block for complete-arch implant rehabilitation. Materials and Methods: A block comprised 4 cylinders, with 2 in the anterior (0 degrees angulation) and 2 in the posterior region (45 degrees distal angulation) (Experimental group) and a standard block with 2 parallel cones in 16 degrees taper (ISO group), were fabricated. Both blocks were scanned consecutively for 15 times by 4 laboratory scanners: IScan, Zfx, 3Shape, and KaVo. Measurements were also made by a coordinate measuring machine (CMM) as the reference. Acquired digital models were inspected with a metrology software. Linear and angular distortions were computed evaluation of trueness, precision and expanded uncertainty of scanners. Effects of blocks and scanners on the scanning trueness and precision were analyzed by Two-way ANOVA (alpha = 0.05). Results: The linear trueness and precision of scanners was significantly poorer when the Experimental block instead of ISO block was used. Significant greater distortions were noticed at the 45 degrees sites than the 0 degrees sites and no significant effect of inter-implant distance on the scanning accuracy was found. Zfx, IScan and 3Shape exhibited comparable expanded uncertainties (10.6 similar to 11.8) but KaVo showed the greatest (19.3) in complete-arch implant scans. Conclusion: The ISO block might not be more suitable than the Experimental block for evaluating the accuracy of laboratory scanner for complete-arch implant scanning. All the scanners tested except one demonstrated acceptable accuracy for complete-arch implant digitization. Scanning accuracy was compromised by unfavourable angulation of implants. Clinical significance: It is questionable whether the laboratory scanners validated for tooth-borne cases could also perform well for implant cases. It might be necessary to develop a new calibration object specifically for complete-arch implant scenarios to verify the capability of laboratory scanners in the implant workflow.
引用
收藏
页数:9
相关论文
共 50 条
  • [31] Impact of color temperature and illuminance of ambient light conditions on the accuracy of complete-arch digital implant scans
    Ochoa-Lopez, Gaston
    Revilla-Leon, Marta
    Gomez-Polo, Miguel
    CLINICAL ORAL IMPLANTS RESEARCH, 2024, 35 (08) : 898 - 905
  • [32] Accuracy of intraoral optical scan versus stereophotogrammetry for complete-arch digital implant impression: An in vitro study
    Pozzi, Alessandro
    Agliardi, Enrico
    Lio, Fabrizio
    Nagy, Katalin
    Nardi, Alessandra
    Arcuri, Lorenzo
    JOURNAL OF PROSTHODONTIC RESEARCH, 2024, 68 (01) : 172 - 180
  • [33] ACCURACY OF IMPRESSION MATERIALS FOR COMPLETE-ARCH FIXED PARTIAL DENTURES
    LIN, CC
    ZIEBERT, GJ
    DONEGAN, SJ
    DHURU, VB
    JOURNAL OF PROSTHETIC DENTISTRY, 1988, 59 (03): : 288 - 291
  • [34] Complete-Arch Accuracy of Four Intraoral Scanners: An In Vitro Study
    Celeghin, Giordano
    Franceschetti, Giulio
    Mobilio, Nicola
    Fasiol, Alberto
    Catapano, Santo
    Corsalini, Massimo
    Grande, Francesco
    HEALTHCARE, 2021, 9 (03)
  • [35] Comparison of intraoral and laboratory scanners to an industrial-grade scanner while analyzing the fabrication trueness of polymer and titanium complete-arch implant-supported frameworks
    Yilmaz, Burak
    Dede, Dogu Omur
    Donmez, Mustafa Borga
    Kucukekenci, Ahmet Serkan
    Lu, Wei-En
    Schumacher, Fernanda Lang
    Cakmak, Gulce
    JOURNAL OF DENTISTRY, 2023, 138
  • [36] AN IN VITRO COMPARISON OF PHOTOGRAMMETRIC AND CONVENTIONAL COMPLETE-ARCH IMPLANT IMPRESSION TECHNIQUES
    Bergin, Junping Ma
    Rubenstein, Jeffrey E.
    Mancl, Lloyd
    Brudvik, James S.
    Raigrodski, Ariel J.
    JOURNAL OF PROSTHETIC DENTISTRY, 2013, 110 (04): : 243 - 251
  • [37] Influence of ambient light conditions on the accuracy and scanning time of seven intraoral scanners in complete-arch implant scans
    Ochoa-Lopez, Gaston
    Cascos, Rocio
    Antonaya-Martin, Jose Luis
    Revilla-Leon, Marta
    Gomez-Polo, Miguel
    JOURNAL OF DENTISTRY, 2022, 121
  • [38] Accuracy of Complete-Arch Implant Digital Scans: Effect of Scanning Protocol, Number of Implants, and Scan Body Splinting
    Denneulin, Thomas
    Rignon-Bret, Christophe
    Ravalec, Guy
    Tapie, Laurent
    Bouter, Denis
    Wulfman, Claudine
    INTERNATIONAL JOURNAL OF PROSTHODONTICS, 2023, 36 (02) : 219 - 227
  • [39] Passively fitting implant-supported complete-arch interim restoration
    Liu, Xiaoqiang
    Li, Man
    Liu, Jianzhang
    Tan, Jianguo
    JOURNAL OF PROSTHETIC DENTISTRY, 2019, 121 (05): : 733 - 736
  • [40] Accuracy of complete-arch digital implant impression with intraoral optical scanning and stereophotogrammetry: An in vivo prospective comparative study
    Pozzi, Alessandro
    Carosi, Paolo
    Gallucci, German O.
    Nagy, Katalin
    Nardi, Alessandra
    Arcuri, Lorenzo
    CLINICAL ORAL IMPLANTS RESEARCH, 2023, 34 (10) : 1106 - 1117