Novel x-ray source for dual-energy subtraction angiography

被引:6
|
作者
Tuffanelli, A [1 ]
Taibi, A [1 ]
Baldazzi, G [1 ]
Bollini, D [1 ]
Gombia, M [1 ]
Ramello, L [1 ]
Gambaccini, M [1 ]
机构
[1] Univ Ferrara, Dipartimento Fis, I-44100 Ferrara, Italy
关键词
digital subtraction angiography; dual energy angiography; tungsten anode x-ray tube; dichromatic x-ray beams; Bragg diffraction; mosaic crystal;
D O I
10.1117/12.465572
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
In angiography practice an iodate contrast medium is injected in patient vessels with catheters. The absorption of x-rays raises immediately after the iodine K-edge energy (33.17 keV). In digital subtraction angiography, two images are used, acquired before and after the injection of the contrast medium, respectively. The vessels morphology results from the difference of images so obtained. This technique involves a non-negligible risk of morbidity or mortality, due to high concentration of injected contrast agent. We are investigating a new source which produces two thin parallel quasi-monochromatic beams - having peak energies centered before and after the iodine K-edge energy, respectively - by using a conventional x-ray tube and a highly oriented pyrolytic graphite mosaic crystal. The polychromatic x-rays incident on the crystal are monochromatized by Bragg diffraction and splitted in two thin parallel beams, by means of a collimating system. These two beams impinge on the phantom simulating patient vessels and are detected with solid-state array detectors. The image results as difference between the remaining intensities of two beams. We report a preliminary study of the new technique performed both with theoretical simulations and experimental measurements. Results of computer simulation give information about characteristics as size and quality of the quasi-monochromatic beams, that should be considered in detail to design a system dedicated to the clinical practice. Experimental measurements have been performed on a small-field detector in order to shows the enhancement of image contrast obtained with the application of the new technique.
引用
收藏
页码:311 / 319
页数:5
相关论文
共 50 条
  • [31] DUAL-ENERGY X-RAY ABSORPTIOMETRY OF THE CALCANEUS - COMPARISON WITH VERTEBRAL DUAL-ENERGY X-RAY ABSORPTIOMETRY AND QUANTITATIVE COMPUTED-TOMOGRAPHY
    LAVALJEANTET, AM
    BERGOT, C
    WILLIAMS, M
    DAVIDSON, K
    LAVALJEANTET, M
    CALCIFIED TISSUE INTERNATIONAL, 1995, 56 (01) : 14 - 18
  • [32] Resolution enhancement in dual-energy x-ray imaging
    Gravel, P
    Després, P
    Beaudoin, G
    de Guise, JA
    Medical Imaging 2005: Image Processing, Pt 1-3, 2005, 5747 : 614 - 624
  • [33] Dual-energy X-ray absorptiometry scan: Guidelines
    Haymann, Jean-Philippe
    PROGRES EN UROLOGIE, 2010, 20 (04): : F119 - F122
  • [34] DUAL-ENERGY X-RAY ABSORPTIOMETRY IN NEUROMUSCULAR DISEASES
    KANDA, F
    FUJII, Y
    TAKAHASHI, K
    FUJITA, T
    MUSCLE & NERVE, 1994, 17 (04) : 431 - 435
  • [35] DUAL-ENERGY X-RAY ABSORPTIOMETRY - INTEREST IN ORTHOPEDICS
    MAUGARS, Y
    BERTHELOT, JM
    DELECRIN, J
    TRECANT, M
    PASSUTI, N
    DACULSI, G
    PROST, A
    REVUE DE CHIRURGIE ORTHOPEDIQUE ET REPARATRICE DE L APPAREIL MOTEUR, 1995, 81 (04): : 326 - 332
  • [36] Technical standardization for dual-energy X-ray absorptiometry
    Leib, E
    JOURNAL OF CLINICAL DENSITOMETRY, 2004, 7 (01) : 27 - 36
  • [37] Dual-energy X-ray imaging: benefits and limits
    Rebuffel, V.
    Dinten, J-M
    INSIGHT, 2007, 49 (10) : 589 - 594
  • [38] SIMULATION STUDIES OF DUAL-ENERGY X-RAY ABSORPTIOMETRY
    SORENSON, JA
    DUKE, PR
    SMITH, SW
    MEDICAL PHYSICS, 1989, 16 (01) : 75 - 80
  • [39] Dual-energy X-ray absorptiometry in early life
    Lapillonne, A
    Braillon, PM
    Delmas, PD
    Salle, BL
    HORMONE RESEARCH, 1997, 48 : 43 - 49
  • [40] Multiple view dual-energy x-ray imaging
    Evans, P
    Robinson, M
    Hon, HW
    35TH ANNUAL 2001 INTERNATIONAL CARNAHAN CONFERENCE ON SECURITY TECHNOLOGY, PROCEEDINGS, 2001, : 103 - 107