Performance of electronic portal imaging devices (EPIDs) used in radiotherapy:: Image quality and dose measurements

被引:47
|
作者
Cremers, F [1 ]
Frenzel, T [1 ]
Kausch, C [1 ]
Albers, D [1 ]
Schönborn, T [1 ]
Schmidt, R [1 ]
机构
[1] Univ Hamburg Hosp, Clin Radiotherapy & Radiooncol, D-20246 Hamburg, Germany
关键词
portal imaging; image quality; amorphous silicon array; fluoroscopic portal imaging system; response; dosimetry; radiation therapy;
D O I
10.1118/1.1688212
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
The aim of our Study was to compare the image and dosimetric quality of two different imaging systems. The first one is a fluoroscopic electronic portal imaging device (first generation), while the second is based on an amorphous silicon flat-panel array (second generation). The parameters describing image quality include spatial resolution [modulation transfer function (MTF)], noise [noise power spectrum (NPS)], and signal-to-noise transfer [detective quantum efficiency (DQE)]. The dosimetric measurements were compared with ionization chamber as well as with film measurements. The response of the flat-panel imager and the fluoroscopic-optical device was determined performing a two-step Monte Carlo simulation. All measurements were performed in a 6 MV linear accelerator photon beam. The resolution (MTF) of the fluoroscopic device (f(1/2) = 0.3 mm(-1)) is larger than of the amorphous silicon based system (f(1/2) = 0.21 mm(-1)), which is due to the missing backscattered photons and the smaller pixel size. The noise measurements (NPS) show the correlation of neighboring pixels of the amorphous silicon electronic portal imaging device, whereas the NPS of the fluoroscopic system is frequency independent. At zero spatial frequency the DQE of the flat-panel imager has a value of 0.008 (0.8%). Due to the minor frequency dependency this device may be almost x-ray quantum limited. Monte Carlo simulations verified these characteristics. For the fluoroscopic imaging system the DQE at low frequencies is about 0.0008 (0.08%) and degrades with higher frequencies. Dose measurements with the flat-panel imager revealed that images can only be directly converted to portal dose images, if scatter can be neglected. Thus objects distant to the detector (e.g., inhomogeneous dose distribution generated by a modificator) can be verified dosimetrically, while objects close to a detector (e.g., a patient) cannot be verified directly and must be scatter corrected prior to verification. This is justified by the response of the flat-panel imaging device revealing a strong dependency at low energies. (C) 2004 American Association of PhYsicists in Medicine.
引用
收藏
页码:985 / 996
页数:12
相关论文
共 50 条
  • [1] Optimization of image quality and dose for Varian aS500 electronic portal imaging devices (EPIDs)
    McGarry, C. K.
    Grattan, M. W. D.
    Cosgrove, V. P.
    PHYSICS IN MEDICINE AND BIOLOGY, 2007, 52 (23): : 6865 - 6877
  • [2] A REVIEW OF ELECTRONIC PORTAL IMAGING DEVICES (EPIDS)
    BOYER, AL
    ANTONUK, L
    FENSTER, A
    VANHERK, M
    MEERTENS, H
    MUNRO, P
    REINSTEIN, LE
    WONG, J
    MEDICAL PHYSICS, 1992, 19 (01) : 1 - 16
  • [3] Optimization of Electronic Portal Imaging Devices (EPIDs) for Small VMAT Fields
    Koo, M.
    Darko, J.
    Osei, E.
    MEDICAL PHYSICS, 2020, 47 (06) : E822 - E823
  • [4] Image Quality and Information Parameters of Electronic Portal Imaging Devices
    Tzomakas, Marios K.
    Peppa, Vasiliki
    Alexiou, Antigoni
    Karakatsanis, Georgios
    Episkopakis, Anastasios
    Michail, Christos
    Valais, Ioannis
    Fountos, George
    Kalyvas, Nektarios
    Kandarakis, Ioannis S.
    APPLIED SCIENCES-BASEL, 2024, 14 (22):
  • [5] Effect of flat-fielding on intensity-based registrations using electronic portal imaging devices (EPIDs) in radiotherapy
    Wu, J
    Samant, S
    MEDICAL PHYSICS, 2004, 31 (06) : 1729 - 1729
  • [6] 2-D radiation therapy dosimetry using EPIDs: Dose response variation between 3 siemens electronic portal imaging devices (EPIDs)
    Deshpande, Shrikant
    Vial, Philip
    Holloway, Lois
    RADIATION MEASUREMENTS, 2011, 46 (12) : 1916 - 1919
  • [7] Evaluation of an edge method for computed radiography and an electronic portal imaging device in radiotherapy: Image quality measurements
    Soon-Yong Son
    Bo-Young Choe
    Jeong-Woo Lee
    Jung-Min Kim
    Hoi-Woun Jeong
    Ham-Gyum Kim
    Wha-Sun Kim
    Kwang-Yeul Lyu
    Jung-Whan Min
    Ki-Won Kim
    Journal of the Korean Physical Society, 2014, 65 : 1976 - 1984
  • [8] Evaluation of an Edge Method for Computed Radiography and an Electronic Portal Imaging Device in Radiotherapy: Image Quality Measurements
    Son, Soon-Yong
    Choe, Bo-Young
    Lee, Jeong-Woo
    Kim, Jung-Min
    Jeong, Hoi-Woun
    Kim, Ham-Gyum
    Kim, Wha-Sun
    Lyu, Kwang-Yeul
    Min, Jung-Whan
    Kim, Ki-Won
    JOURNAL OF THE KOREAN PHYSICAL SOCIETY, 2014, 65 (11) : 1976 - 1984
  • [9] A comparison of the dosimetric properties of The Electronic Portal Imaging Devices (EPIDs) LC250 and aS500
    Wozniak, Bozena
    Ganowicz, Malgorzata
    Bekman, Adam
    Maniakowski, Zbigniew
    REPORTS OF PRACTICAL ONCOLOGY AND RADIOTHERAPY, 2005, 10 (05) : 249 - 254
  • [10] Routine in-vivo dosimetry using electronic portal imaging devices.(EPIDS): the Maastro experience
    Nijsten, SMJJG
    Mijnheer, BJ
    Lambin, P
    Minken, AWH
    RADIOTHERAPY AND ONCOLOGY, 2004, 73 : S153 - S153