2-D radiation therapy dosimetry using EPIDs: Dose response variation between 3 siemens electronic portal imaging devices (EPIDs)

被引:4
|
作者
Deshpande, Shrikant [1 ,2 ,4 ]
Vial, Philip [1 ,2 ,3 ]
Holloway, Lois [1 ,2 ,3 ,4 ]
机构
[1] Liverpool Canc Therapy Ctr, Sydney, NSW, Australia
[2] Macarthur Canc Therapy Ctr, Sydney, NSW, Australia
[3] Univ Sydney, Sch Phys, Inst Med Phys, Sydney, NSW 2006, Australia
[4] Univ Wollongong, Ctr Med Radiat Phys, Wollongong, NSW 2522, Australia
关键词
VERIFICATION;
D O I
10.1016/j.radmeas.2011.08.006
中图分类号
TL [原子能技术]; O571 [原子核物理学];
学科分类号
0827 ; 082701 ;
摘要
Prerequisite for any new clinical dosimeter is a detailed understanding of the detector's dose response behavior. The purpose of the present study was to investigate the level of variability in dose response characteristics of 3 siemens EPIDs for both 6 and 18 MV photons. Dose response tests for linearity, field size, short- and long-term reproducibilities, ghosting effects and dose rate dependence were undertaken for three different model Siemens EPIDs on three dosimetrically matched linear accelerators. All three EPIDs showed a linear dose response above 20 monitor units. The pixel sensitivity for EPID2 and EPID3 agreed to within 0.5%. EPID1 was 6.3% (6 MV) and 2.3% (18 MV) greater than for EPID2 and EPID3. The field size response and dose rate response agreed within 1% for all three EPIDs. The short-term and long-term reproducibilities for all EPIDs were within 0.5% and 1% respectively. The maximum increase in relative response due to the ghosting effect was 0.5%. The off-axis profiles from uncorrected gain files agreed to within 2% for EPID2 and EPID3 at 6 MV and 18 MV respectively. The off-axis profiles for EPID1 had more pronounced horns. The different dose response behavior of EPID1 is due to a thicker phosphor scintillator compared to EPID2 and EPID3. EPID1 was significantly more sensitive to dose and energy variations and would require separate calibration data for EPID dosimetry. EPIDs with the same phosphor had no significant difference in dose response. Differences in EPID/phosphor design must be considered when commissioning EPIDs for clinical dosimetry. Crown Copyright (C) 2011 Published by Elsevier Ltd. All rights reserved.
引用
收藏
页码:1916 / 1919
页数:4
相关论文
共 27 条
  • [1] Routine in-vivo dosimetry using electronic portal imaging devices.(EPIDS): the Maastro experience
    Nijsten, SMJJG
    Mijnheer, BJ
    Lambin, P
    Minken, AWH
    RADIOTHERAPY AND ONCOLOGY, 2004, 73 : S153 - S153
  • [2] Performance of electronic portal imaging devices (EPIDs) used in radiotherapy:: Image quality and dose measurements
    Cremers, F
    Frenzel, T
    Kausch, C
    Albers, D
    Schönborn, T
    Schmidt, R
    MEDICAL PHYSICS, 2004, 31 (05) : 985 - 996
  • [3] Optimization of image quality and dose for Varian aS500 electronic portal imaging devices (EPIDs)
    McGarry, C. K.
    Grattan, M. W. D.
    Cosgrove, V. P.
    PHYSICS IN MEDICINE AND BIOLOGY, 2007, 52 (23): : 6865 - 6877
  • [4] Effect of flat-fielding on intensity-based registrations using electronic portal imaging devices (EPIDs) in radiotherapy
    Wu, J
    Samant, S
    MEDICAL PHYSICS, 2004, 31 (06) : 1729 - 1729
  • [5] A generalized calibration procedure for in vivo transit dosimetry using siemens electronic portal imaging devices
    Andrea Fidanzio
    Francesca Greco
    Laura Gargiulo
    Savino Cilla
    Domenico Sabatino
    Massimo Cappiello
    Cinzia Di Felice
    Elisabetta Di Castro
    Luigi Azario
    Mariateresa Russo
    Luciano Pompei
    Guido D’Onofrio
    Angelo Piermattei
    Medical & Biological Engineering & Computing, 2011, 49 : 373 - 383
  • [6] A generalized calibration procedure for in vivo transit dosimetry using siemens electronic portal imaging devices
    Fidanzio, Andrea
    Greco, Francesca
    Gargiulo, Laura
    Cilla, Savino
    Sabatino, Domenico
    Cappiello, Massimo
    Di Felice, Cinzia
    Di Castro, Elisabetta
    Azario, Luigi
    Russo, Mariateresa
    Pompei, Luciano
    D'Onofrio, Guido
    Piermattei, Angelo
    MEDICAL & BIOLOGICAL ENGINEERING & COMPUTING, 2011, 49 (03) : 373 - 383
  • [7] Dosimetry in radiotherapy using a-Si EPIDs: Systems, methods, and applications focusing on 3D patient dose estimation
    McCurdy, B. M. C.
    7TH INTERNATIONAL CONFERENCE ON 3D RADIATION DOSIMETRY (IC3DDOSE), 2013, 444
  • [8] The feasibility of the pretreatment verification of 2D dose distributions in radiation therapy with small fields using the electronic portal imaging device
    Tuncel, Nina
    Mahdavi, Seied Rabi
    Haghparast, Mohammad
    Karakus, Ismail
    Haghparast, Abbas
    Nazari, Vahideh
    JOURNAL OF CANCER RESEARCH AND THERAPEUTICS, 2023, 19 : S815 - S820
  • [9] In Vivo 3D Dose Verification for IMRT Using Electronic Portal Imaging Device (EPID)
    Chang, C.
    Anderson, N. G.
    Zhu, T. C.
    MEDICAL PHYSICS, 2011, 38 (06)
  • [10] Pretreatment evaluation and correction of setup variation for external beam radiation therapy patients using an electronic portal imaging device
    Wittmer, M
    Pisansky, TM
    Kruse, JJ
    Herman, MG
    RADIOLOGY, 2001, 221 : 222 - 223